Diagnostic reference level quantities for adult chest and abdomen-pelvis CT examinations: correlation with organ doses

被引:2
|
作者
Costa, Paulo Roberto [1 ]
Tomal, Alessandra [2 ]
de Oliveira Castro, Jullianna Cristina [3 ]
Nunes, Isabella Paziam Fernandes [3 ]
Nersissian, Denise Yanikian [1 ]
Sawamura, Marcio Valente Yamada [4 ]
Leao Filho, Hilton [4 ]
Lee, Choonsik [5 ]
机构
[1] Univ Sao Paulo, Inst Phys, R Matao 1371, BR-05508090 Sao Paulo, SP, Brazil
[2] Univ Estadual Campinas, Inst Phys Gleb Watagin, Campinas, Brazil
[3] Univ Sao Paulo, Med Sch, Sao Paulo, Brazil
[4] Univ Sao Paulo, Med Sch, Div Radiol, Sao Paulo, Brazil
[5] NCI, Div Canc Epidemiol & Genet, NIH, Bethesda, MD USA
基金
瑞典研究理事会; 巴西圣保罗研究基金会;
关键词
Organ doses; Computed tomography; Chest; Abdomen-pelvis; Statistics; SIZE; PHANTOMS; FAMILY; SSDE;
D O I
10.1186/s13244-023-01403-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesTo evaluate correlations between DRL quantities (DRLq) stratified into patient size groups for non-contrast chest and abdomen-pelvis CT examinations in adult patients and the corresponding organ doses.MethodsThis study presents correlations between DRLq (CTDIvol, DLP and SSDE) stratified into patient size ranges and corresponding organ doses shared in four groups: inside, peripheral, distributed and outside. The demographic, technical and dosimetric parameters were used to identify the influence of these quantities in organ doses. A robust statistical method was implemented in order to establish these correlations and its statistical significance.ResultsMedian values of the grouped organ doses are presented according to the effective diameter ranges. Organ doses in the regions inside the imaged area are higher than the organ doses in peripheral, distributed and outside regions, excepted to the peripheral doses associated with chest examinations. Different levels of statistical significance between organ doses and the DRLq were presented.ConclusionsCorrelations between DRLq and target-organ doses associated with clinical practice can support guidance's to the establishment of optimization criteria. SSDE demonstrated to be significant in the evaluation of organ doses is also highlighted. The proposed model allows the design of optimization actions with specific risk-reduction results.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Diagnostic reference level quantities for adult chest and abdomen-pelvis CT examinations: correlation with organ doses
    Paulo Roberto Costa
    Alessandra Tomal
    Jullianna Cristina de Oliveira Castro
    Isabella Paziam Fernandes Nunes
    Denise Yanikian Nersissian
    Márcio Valente Yamada Sawamura
    Hilton Leão Filho
    Choonsik Lee
    [J]. Insights into Imaging, 14
  • [2] Typical values statistical analysis for adult chest and abdomen-pelvis CT examinations
    Costa, Paulo Roberto
    Castro, Jullianna Cristina de Oliveira
    Nunes, Isabella Paziam Fernandes
    Nersissian, Denise Yanikian
    Sawamura, Marcio Yamada
    Leao Filho, Hilton
    Tomal, Alessandra
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2024, 221
  • [3] Pediatric chest-abdomen-pelvis and abdomen-pelvis CT images with expert organ contours
    Jordan, Petr
    Adamson, Philip M.
    Bhattbhatt, Vrunda
    Beriwal, Surabhi
    Shen, Sangyu
    Radermecker, Oskar
    Bose, Supratik
    Strain, Linda S.
    Offe, Michael
    Fraley, David
    Principi, Sara
    Ye, Dong Hye
    Wang, Adam S.
    van Heteren, John
    Vo, Nghia-Jack
    Schmidt, Taly Gilat
    [J]. MEDICAL PHYSICS, 2022, 49 (05) : 3523 - 3528
  • [4] Correlation analysis of organ doses determined by Monte Carlo simulation with dose metrics for patients undergoing chest-abdomen-pelvis CT examinations
    Fujii, Keisuke
    Nomura, Keiichi
    Muramatsu, Yoshihisa
    Goto, Takahiro
    Obara, Satoshi
    Ota, Hiroyuki
    Tsukagoshi, Shinsuke
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 1 - 9
  • [5] RADIATION DOSES IN CHEST, ABDOMEN AND PELVIS CT PROCEDURES
    Manssor, E.
    Abuderman, A.
    Osman, S.
    Alenezi, S. B.
    Almehemeid, S.
    Babikir, E.
    Alkhorayef, M.
    Sulieman, A.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 194 - 198
  • [6] ROLE OF FIT AND CT ABDOMEN-PELVIS IN PRIORITISING DIAGNOSTIC COLONOSCOPY FOR COLORECTAL CANCER
    Asmat, Huma
    Javaid, Babur
    Yanagisawa, Yuuki
    Haddad, Aiman
    Ashour, Abd-El-Aziz
    Javaid, Zaynah
    [J]. GUT, 2022, 71 : A112 - A112
  • [7] US Diagnostic Reference Levels and Achievable Doses for 10 Adult CT Examinations
    Kanal, Kalpana M.
    Butler, Priscilla F.
    Sengupta, Debapriya
    Bhargavan-Chatfield, Mythreyi
    Coombs, Laura P.
    Morin, Richard L.
    [J]. RADIOLOGY, 2017, 284 (01) : 120 - 133
  • [8] Quality Improvement: Diagnostic Reference Levels for Interior Health CT exams -L-Spine, Chest/Abdomen/pelvis, Abdomen/Pelvis, Head
    Bjarnason, Thorarin
    [J]. MEDICAL PHYSICS, 2016, 43 (08) : 4951 - 4951
  • [9] Organ doses to adult patients for chest CT
    Huda, Walter
    Sterzik, Alexander
    Tipnis, Sameer
    Schoepf, U. Joseph
    [J]. MEDICAL PHYSICS, 2010, 37 (02) : 842 - 847
  • [10] Development of a Web-Based Tool for the Estimation of Organ Doses in Pediatric and Adult Chest CT Examinations
    Damilakis, J.
    Stratakis, J.
    Myronakis, M.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E647 - E647