Harnessing dual-energy CT for glycogen quantification: a phantom analysis

被引:0
|
作者
Li, Meiqin [1 ]
Li, Zhoulei [1 ]
Wei, Luyong [1 ]
Li, Lujie [1 ]
Wang, Meng [1 ]
He, Shaofu [1 ]
Peng, Zhenpeng [1 ]
Feng, Shi-Ting [1 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 1, Dept Radiol, 58 Zhongshan Second Rd, Guangzhou 510030, Peoples R China
基金
中国国家自然科学基金;
关键词
Glycogen; quantification; dual-energy computed tomography; LIVER; RADIOTRACER;
D O I
10.21037/qims-22-1234
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Non-invasive glycogen quantification in vivo could provide crucial information on biological processes for glycogen storage disorder. Using dual-energy computed tomography (DECT), this study aimed to assess the viability of quantifying glycogen content in vitro.Methods: A fast kilovolt-peak switching DECT was used to scan a phantom containing 33 cylinders with different proportions of glycogen and iodine mixture at varying doses. The virtual glycogen concentration (VGC) was then measured using material composition images. Additionally, the correlations between VGC and nominal glycogen concentration (NGC) were evaluated using least-square linear regression, then the calibration curve was constructed. Quantitative estimation was performed by calculating the linearity, conversion factor (inverse of curve slope), stability, sensitivity (limit of detection/limit of quantification), repeatability (inter-class correlation coefficient), and variability (coefficient of variation).Results: In all conditions, excellent linear relationship between VGC and NGC were observed (P<0.001, coefficient of determination: 0.989-0.997; residual root-mean-square error of glycogen: 1.862-3.267 mg/mL). The estimated conversion factor from VGC to NGC was 3.068-3.222. In addition, no significant differences in curve slope were observed among different dose levels and iodine densities. The limit of detection and limit of quantification had respective ranges of 6.421-15.315 and 10.95-16.46 mg/mL. The data demonstrated excellent scan-repeat scan agreement (inter-class correlation coefficient, 0.977-0.991) and small variation (coefficient of variation, 0.1-0.2%).Conclusions: The pilot phantom analysis demonstrated the feasibility and efficacy of detecting and quantifying glycogen using DECT and provided good quantitative performance with significant stability and reproducibility/variability. Thus, in the future, DECT could be used as a convenient method for glycogen quantification to provide more reliable information for clinical decision-making.
引用
收藏
页码:4933 / 4942
页数:10
相关论文
共 50 条
  • [31] Characterization of Vulnerable Plaque with Dual-Energy during CT Coronary Angiography: A Phantom Study
    Ursani, A.
    Hoy, C.
    Moghe, Sachin
    Paul, N. S.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, 2015, VOLS 1 AND 2, 2015, 51 : 91 - 94
  • [32] Enhancing Soft Tissue Differentiation with Different Dual-Energy CT Systems: A Phantom Study
    Gallo, Pasqualina
    D'Alessio, Andrea
    Pascuzzo, Riccardo
    Gallo, Salvatore
    Fumagalli, Maria Luisa
    Ortenzia, Ornella
    Tenconi, Chiara
    Cavatorta, Claudia
    Pignoli, Emanuele
    Ghetti, Caterina
    Bruzzone, Maria Grazia
    De Martin, Elena
    APPLIED SCIENCES-BASEL, 2024, 14 (05):
  • [33] Dual-energy electronic cleansing for non-cathartic CT Colonography: a phantom study
    Cai, Wenli
    Liu, Bob
    Yoshida, Hiroyuki
    MEDICAL IMAGING 2010: COMPUTER - AIDED DIAGNOSIS, 2010, 7624
  • [34] Accuracy of dual-energy computed tomography for the quantification of iodine in a soft tissue-mimicking phantom
    Li, Jung-Hui
    Du, Yeh-Ming
    Huang, Hsuan-Ming
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (05): : 418 - 426
  • [35] Investigating new CT contrast agents: a phantom study exploring quantification and differentiation methods for high-Z elements using dual-energy CT
    Soesbe, Todd C.
    Xi, Yin
    Nasr, Khaled
    Leyendecker, John R.
    Lenkinski, Robert E.
    Lewis, Matthew A.
    EUROPEAN RADIOLOGY, 2021, 31 (10) : 8060 - 8067
  • [36] Investigating new CT contrast agents: a phantom study exploring quantification and differentiation methods for high-Z elements using dual-energy CT
    Todd C. Soesbe
    Yin Xi
    Khaled Nasr
    John R. Leyendecker
    Robert E. Lenkinski
    Matthew A. Lewis
    European Radiology, 2021, 31 : 8060 - 8067
  • [37] Dual-Energy CT Pulmonary Angiography: Quantification of Disease Burden and Impact on Management
    Martin S.S.
    van Assen M.
    Griffith L.P.
    De Cecco C.N.
    Varga-Szemes A.
    Bauer M.J.
    Wichmann J.L.
    Vogl T.J.
    Schoepf U.J.
    Current Radiology Reports, 6 (10)
  • [38] Gastrointestinal Applications of Iodine Quantification Using Dual-Energy CT: A Systematic Review
    Xu, Jack Junchi
    Taudorf, Mikkel
    Ulriksen, Peter Sommer
    Achiam, Michael Patrick
    Resch, Timothy Andrew
    Nielsen, Michael Bachmann
    Lonn, Lars Birger
    Hansen, Kristoffer Lindskov
    DIAGNOSTICS, 2020, 10 (10)
  • [39] Simple Approach for Quantification of Metal Artifact Reduction Capabalities of Dual-Energy CT
    Lamichhane, N.
    Padgett, K.
    Li, X.
    Mihaylov, I.
    MEDICAL PHYSICS, 2016, 43 (06) : 3423 - 3423
  • [40] Quantification of liver iron content with CT-added value of dual-energy
    Fischer, Michael A.
    Reiner, Caecilia S.
    Raptis, Dimitri
    Donati, Olivio
    Goetti, Robert
    Clavien, Pierre-Alain
    Alkadhi, Hatem
    EUROPEAN RADIOLOGY, 2011, 21 (08) : 1727 - 1732