Harnessing dual-energy CT for glycogen quantification: a phantom analysis

被引:0
|
作者
Li, Meiqin [1 ]
Li, Zhoulei [1 ]
Wei, Luyong [1 ]
Li, Lujie [1 ]
Wang, Meng [1 ]
He, Shaofu [1 ]
Peng, Zhenpeng [1 ]
Feng, Shi-Ting [1 ]
机构
[1] Sun Yat Sen Univ, Affiliated Hosp 1, Dept Radiol, 58 Zhongshan Second Rd, Guangzhou 510030, Peoples R China
基金
中国国家自然科学基金;
关键词
Glycogen; quantification; dual-energy computed tomography; LIVER; RADIOTRACER;
D O I
10.21037/qims-22-1234
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Non-invasive glycogen quantification in vivo could provide crucial information on biological processes for glycogen storage disorder. Using dual-energy computed tomography (DECT), this study aimed to assess the viability of quantifying glycogen content in vitro.Methods: A fast kilovolt-peak switching DECT was used to scan a phantom containing 33 cylinders with different proportions of glycogen and iodine mixture at varying doses. The virtual glycogen concentration (VGC) was then measured using material composition images. Additionally, the correlations between VGC and nominal glycogen concentration (NGC) were evaluated using least-square linear regression, then the calibration curve was constructed. Quantitative estimation was performed by calculating the linearity, conversion factor (inverse of curve slope), stability, sensitivity (limit of detection/limit of quantification), repeatability (inter-class correlation coefficient), and variability (coefficient of variation).Results: In all conditions, excellent linear relationship between VGC and NGC were observed (P<0.001, coefficient of determination: 0.989-0.997; residual root-mean-square error of glycogen: 1.862-3.267 mg/mL). The estimated conversion factor from VGC to NGC was 3.068-3.222. In addition, no significant differences in curve slope were observed among different dose levels and iodine densities. The limit of detection and limit of quantification had respective ranges of 6.421-15.315 and 10.95-16.46 mg/mL. The data demonstrated excellent scan-repeat scan agreement (inter-class correlation coefficient, 0.977-0.991) and small variation (coefficient of variation, 0.1-0.2%).Conclusions: The pilot phantom analysis demonstrated the feasibility and efficacy of detecting and quantifying glycogen using DECT and provided good quantitative performance with significant stability and reproducibility/variability. Thus, in the future, DECT could be used as a convenient method for glycogen quantification to provide more reliable information for clinical decision-making.
引用
收藏
页码:4933 / 4942
页数:10
相关论文
共 50 条
  • [1] Quantification of Metal Artifact Reduction of Dual-Energy CT: A Phantom Study
    Li, K.
    Leung, W.
    MEDICAL PHYSICS, 2017, 44 (06) : 3036 - 3036
  • [2] Lower limit of iron quantification using dual-energy CT - a phantom study
    Jiang, Xia
    Hintenlang, David E.
    White, Richard D.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (01): : 299 - 307
  • [3] Intermanufacturer Comparison of Dual-Energy CT Iodine Quantification and Monochromatic Attenuation: A Phantom Study
    Jacobsen, Megan C.
    Schellingerhout, Dawid
    Wood, Cayla A.
    Tamm, Eric P.
    Godoy, Myrna C.
    Sun, Jia
    Cody, Dianna D.
    RADIOLOGY, 2018, 287 (01) : 224 - 234
  • [4] Iodine Quantification With Dual-Energy CT: Phantom Study and Preliminary Experience With Renal Masses
    Chandarana, Hersh
    Megibow, Alec J.
    Cohen, Benjamin A.
    Srinivasan, Ramya
    Kim, Danny
    Leidecker, Christianne
    Macari, Michael
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (06) : W693 - W700
  • [5] Dual-Energy CT for Quantification of Urinary Stone Composition in Mixed Stones: A Phantom Study
    Leng, Shuai
    Huang, Alice
    Cardona, Juan Montoya
    Duan, Xinhui
    Williams, James C.
    McCollough, Cynthia H.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 207 (02) : 321 - 329
  • [6] Multivendor comparison of quantification accuracy of effective atomic number by Dual-Energy CT: A phantom study
    Qin, Tian
    Wang, Mengting
    Fan, Yihan
    Wang, Jing
    Gao, Zhizhen
    Wang, Fan
    Li, Ruomei
    Li, Kui
    Ruan, Chengcheng
    Liang, Baohui
    EUROPEAN JOURNAL OF RADIOLOGY, 2024, 180
  • [7] Evaluation of Intra-Manufacturer Dual-Energy CT Iodine Quantification Accuracy: Phantom Studies
    Zhao, A.
    Borhani, A.
    Catania, R.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [8] Multivendor Comparison of Quantification Accuracy of Iodine Concentration and Attenuation Measurements by Dual-Energy CT: A Phantom Study
    Chen, Yong
    Zhong, Jingyu
    Wang, Lan
    Shi, Xiaomeng
    Chang, Rui
    Fan, Jing
    Jiang, Jiang
    Xia, Yihan
    Yan, Fuhua
    Yao, Weiwu
    Zhang, Huan
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2022, 219 (05) : 827 - 839
  • [9] Quantification of Nanoparticle Radiosensitizers Using Dual-Energy CT
    Harris, T.
    Jacobson, M.
    Bennett, S.
    Myronakis, M.
    Ozoemelam, I.
    Hu, Y.
    Sudhyadhom, A.
    Lux, F.
    Tillement, O.
    Berbeco, R.
    MEDICAL PHYSICS, 2022, 49 (06) : E270 - E270
  • [10] Dual-energy CT: a phantom comparison of different platforms for abdominal imaging
    Thorsten Sellerer
    Peter B. Noël
    Manuel Patino
    Anushri Parakh
    Sebastian Ehn
    Sascha Zeiter
    Jasmin A. Holz
    Johannes Hammel
    Alexander A. Fingerle
    Franz Pfeiffer
    David Maintz
    Ernst J. Rummeny
    Daniela Muenzel
    Dushyant V. Sahani
    European Radiology, 2018, 28 : 2745 - 2755