3D Printed Lung Phantom for Individual Monitoring

被引:1
|
作者
Capello, Kevin [1 ]
Tremblay, Marilyn [1 ]
Schiebelbein, Aryanna [1 ]
Janzen, Noah [1 ]
机构
[1] Hlth Canada, Radiat Protect Bur, 775 Brookfield Rd, Ottawa, ON K1A 1C1, Canada
来源
HEALTH PHYSICS | 2024年 / 126卷 / 05期
关键词
Radiological monitoring; individual monitoring; internal contamination; lung counting; proficiency test program; 3D printing; SETS;
D O I
10.1097/HP.0000000000001777
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The Human Monitoring Laboratory, Health Canada (HML), has used a 3D printer to re-engineer its Lawrence Livermore National Laboratory (LLNL) foam lung sets (manufactured by Radiology Support Devices, Inc., Long Beach, CA). The foam sets are currently the HML standard for calibrating and performance testing lung-counting systems in Canada. This paper describes the process of creating and validating new 3D-printed lung sets modeled from one of the HML's existing RSD foam sets. The existing sets were custom made, making them costly and difficult to obtain or replace. Also, after many years of use, the HML has found that they are prone to wear and tear. When used with planar inserts containing various isotopes, the blank sets can become contaminated and are difficult to clean. Using 3D printing, the HML has created new blank lung sets that are nearly identical copies of the originals and are inexpensive and easily manufactured. Measurements using natural uranium (Nat U), 241Am, and 152Eu planar lung inserts were performed to compare obtained efficiencies at a wide range of energies using the original RSD foam sets and the 3D-printed ones. Both the foam and the 3D-printed lung sets were counted using the LLNL chest phantom positioned in the same counting geometry in the lung counting system. Biases, all below 15%, were obtained between the foam and the 3D-printed sets for energies above 40 KeV. Based on these results, as well as cost benefits and ease of use, the HML has decided to replace its original RSD foam lung set with the 3D-printed version for its lung performance testing program.
引用
收藏
页码:292 / 295
页数:4
相关论文
共 50 条
  • [21] A Multi-Frequency 3D Printed Hand Phantom for Electromagnetic Measurements
    Beard B.B.
    Iacono M.I.
    Guag J.W.
    Liu Y.
    IEEE Electromagnetic Compatibility Magazine, 2022, 11 (03) : 49 - 54
  • [22] Fast and Simple Automatic 3D Ultrasound Probe Calibration Based on 3D Printed Phantom and an Untracked Marker
    Shen, Jun
    Zemiti, Nabil
    Dillenseger, Jean-Louis
    Poignet, Philippe
    2018 40TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2018, : 878 - 882
  • [23] Development of a 3D Printed Motion Mechanism for a 4D Respiratory Motion Phantom
    Ahmad, Muhammad Akmal
    Mustaza, Seri Mastura
    Mokri, Siti Salasiah
    Azmi, Nor Aniza
    Ahmad, Rozilawati
    Ramli, Rizauddin
    Rahman, Wan Nordiana Wan Abdul
    Abd Rahni, Ashrani Aizzuddin
    2020 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES 2020): LEADING MODERN HEALTHCARE TECHNOLOGY ENHANCING WELLNESS, 2021, : 132 - 135
  • [24] Biofabrication and Monitoring of a 3D Printed Skin Model for Melanoma
    Vazquez-Aristizabal, Paula
    Henriksen-Lacey, Malou
    Garcia-Astrain, Clara
    de Aberasturi, Dorleta Jimenez
    Langer, Judith
    Epelde, Claudia
    Litti, Lucio
    Liz-Marzan, Luis M.
    Izeta, Ander
    ADVANCED HEALTHCARE MATERIALS, 2024, 13 (27)
  • [25] 3D-printed biological cell phantom for testing 3D quantitative phase imaging systems
    Ziemczonok, Michal
    Kus, Arkadiusz
    Wasylczyk, Piotr
    Kujawinska, Malgorzata
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [26] 3D-printed biological cell phantom for testing 3D quantitative phase imaging systems
    Michał Ziemczonok
    Arkadiusz Kuś
    Piotr Wasylczyk
    Małgorzata Kujawińska
    Scientific Reports, 9
  • [27] 3D Printed
    Good, Andrew
    MATERIALS EVALUATION, 2016, 74 (07) : 984 - 989
  • [28] Preclinical dose verification using a 3D printed mouse phantom for radiobiology experiments
    Esplen, Nolan
    Therriault-Proulx, Francois
    Beaulieu, Luc
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2019, 46 (11) : 5294 - 5303
  • [29] Assessment of structural similarity in CT using filtered backprojection and iterative reconstruction: a phantom study with 3D printed lung vessels
    Joemai, Raoul M. S.
    Geleijns, Jacob
    BRITISH JOURNAL OF RADIOLOGY, 2017, 90 (1079):
  • [30] RSC: A 3D printed eyeball phantom for Sr-90 dosimetry measurements
    Maxwell, S. K.
    Crowel, S. B.
    Simpson-Pagel, E. M.
    Cassini, N.
    Kairn, T.
    11TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, IC3DDOSE, 2022, 2167