Verification of the active pharmaceutical ingredient in tablets using a low-cost near-infrared spectrometer

被引:0
|
作者
Gabel, Julia [1 ]
Gnegel, Gesa [1 ]
Kessler, Waltraud [2 ]
Sacre, Pierre-Yves [3 ]
Heide, Lutz [1 ]
机构
[1] Eberhard Karls Univ Tubingen, Pharmaceut Inst, D-72070 Tubingen, Germany
[2] Reutlingen Univ, Sch Life Sci, D-72762 Reutlingen, Germany
[3] Univ Liege ULiege, Dept Pharm, Res Support Unit Chemometr, CIRM, B-4000 Liege, Belgium
来源
TALANTA OPEN | 2023年 / 8卷
关键词
Near -infrared spectroscopy; Pharmaceuticals; Medicine quality; Substandard and falsified medicines; DD-SIMCA; Authentication; HAND-HELD NIR; SPECTROSCOPY; STANDARDIZATION; MEDICINES;
D O I
10.1016/j.talo.2023.100270
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The present study investigated the possibilities and limitations of using a low-cost NIR spectrometer for the verification of the presence of the declared active pharmaceutical ingredients (APIs) in tablet formulations, especially for medicine screening studies in low-resource settings. Spectra from 950 to 1650 nm were recorded for 170 pharmaceutical products representing 41 different APIs, API combinations or placebos. Most of the products, including 20 falsified medicines, had been collected in medicine quality studies in African countries. After exploratory principal component analysis, models were built using data-driven soft independent modelling of class analogy (DD-SIMCA), a one-class classifier algorithm, for tablet products of penicillin V, sulfamethox-azole/trimethoprim, ciprofloxacin, furosemide, metronidazole, metformin, hydrochlorothiazide, and doxycy-cline. Spectra of amoxicillin and amoxicillin/clavulanic acid tablets were combined into a single model. Models were tested using Procrustes cross-validation and by projection of spectra of tablets containing the same or different APIs. Tablets containing no or different APIs could be identified with 100 % specificity in all models. A separation of the spectra of amoxicillin and amoxicillin/clavulanic acid tablets was achieved by partial least squares discriminant analysis. 15 out of 19 external validation products (79 %) representing different brands of the same APIs were correctly identified as members of the target class; three of the four rejected samples showed an API mass percentage of the total tablet weight that was out of the range covered in the respective calibration set. Therefore, in future investigations larger and more representative spectral libraries are required for model building. Falsified medicines containing no API, incorrect APIs, or grossly incorrect amounts of the declared APIs could be readily identified. Variation between different NIR-S-G1 spectroscopic devices led to a loss of accuracy if spectra recorded with different devices were pooled. Therefore, piecewise direct standardization was applied for calibration transfer. The investigated method is a promising tool for medicine screening studies in low-resource settings.
引用
收藏
页数:16
相关论文
共 50 条
  • [41] Smart Detection of Faults in Beers Using Near-Infrared Spectroscopy, a Low-Cost Electronic Nose and Artificial Intelligence
    Gonzalez Viejo, Claudia
    Fuentes, Sigfredo
    Hernandez-Brenes, Carmen
    [J]. FERMENTATION-BASEL, 2021, 7 (03):
  • [42] Low-Cost Recognition of Plastic Waste Using Deep Learning and a Multi-Spectral Near-Infrared Sensor
    Martinez-Hernandez, Uriel
    West, Gregory
    Assaf, Tareq
    [J]. SENSORS, 2024, 24 (09)
  • [43] Imaging the invisible using modified digital still cameras for straightforward and low-cost archaeological near-infrared photography
    Verhoeven, Geert
    [J]. JOURNAL OF ARCHAEOLOGICAL SCIENCE, 2008, 35 (12) : 3087 - 3100
  • [44] SIR - a flexible, compact, low mass, near-infrared spectrometer
    Mall, U
    Nathues, A
    Keller, HU
    [J]. SENSORS, SYSTEMS AND NEXT-GENERATION SATELLITES VII, 2004, 5234 : 314 - 322
  • [45] Real-time simultaneous detection of microbial contamination and determination of an ultra low content active pharmaceutical ingredient in tazarotene gel by near-infrared spectroscopy
    Xie, Qian
    Wu, Ruanqi
    Zhong, Xiaoxiao
    Dong, Yanhong
    Fan, Qi
    [J]. RSC ADVANCES, 2018, 8 (48) : 27037 - 27044
  • [46] Active content determination of pharmaceutical tablets using near infrared spectroscopy as Process Analytical Technology tool
    Chavez, Pierre-Francois
    Sacre, Pierre-Yves
    De Bleye, Charlotte
    Netchacovitch, Lauranne
    Mantanus, Jerome
    Motte, Henri
    Schubert, Martin
    Hubert, Philippe
    Ziemons, Eric
    [J]. TALANTA, 2015, 144 : 1352 - 1359
  • [47] ASSAY OF THE ACTIVE INGREDIENT OF AN MRI CONTRAST AGENT USING MID-INFRARED AND NEAR-INFRARED SPECTROSCOPY AND MULTIVARIATE CALIBRATION
    HARBORG, E
    KLAEBOE, P
    NIELSEN, CJ
    SJOGREN, CE
    [J]. JOURNAL OF MOLECULAR STRUCTURE, 1995, 348 : 139 - 142
  • [48] Measurement of Moisture Content in Pharmaceutical Tablets by Handheld Near-Infrared Spectrometer: Adopting Quality by Design Approach to Analytical Method Lifecycle Management
    Patel, Amrish
    Jin, Chunguang
    Handzo, Brittany
    Kalyanaraman, Ravi
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2023, 229
  • [49] Monitoring the quality of ethanol-based hand sanitizers by low-cost near-infrared spectroscopy
    Pasquini, Celio
    Hespanhol, Maria C.
    Cruz, Kaique A. M. L.
    Pereira, Alexandre F.
    [J]. MICROCHEMICAL JOURNAL, 2020, 159
  • [50] An in-depth analysis of Raman and near-infrared chemical images of common pharmaceutical tablets
    Sasic, Slobodan
    [J]. APPLIED SPECTROSCOPY, 2007, 61 (03) : 239 - 250