Home-Use Hyaluronic Acid Jet Injectors: Unreliable and Unsafe

被引:0
|
作者
Juch, Rosalie N. S. [1 ]
Bik, Liora [2 ]
Boeijink, Neill [1 ]
de Vos, Judith [3 ]
Dobbe, Iwan J. G. G. [3 ,4 ,5 ]
Bloemen, Paul R. [3 ]
van Doorn, Martijn B. A. [2 ]
Velthuis, Peter [2 ]
Aalders, Maurice C. G. [3 ]
Wolkerstorfer, Albert [1 ]
机构
[1] Univ Amsterdam, Dermatol, Amsterdam UMC Locat, Amsterdam, Netherlands
[2] Erasmus MC Univ, Med Ctr, Dermatol, Rotterdam, Netherlands
[3] Univ Amsterdam, Biomed Engn & Phys, Amsterdam UMC Locat, Amsterdam, Netherlands
[4] Musculoskeletal Hlth Restorat & Dev, Amsterdam Movement Sci, Amsterdam, Netherlands
[5] Amsterdam Cardiovasc Sci, Microcirculat, Amsterdam, Netherlands
关键词
D O I
10.1097/DSS.0000000000003971
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
BACKGROUNDNeedle-free hyaluronic acid (HA) jet injectors are gaining popularity for rejuvenation treatment. The devices are widely available online and are used for self-injection or in beauty salons by nonphysicians. However, little is known about their performance and safety.OBJECTIVETo explore the injection efficiency and cutaneous biodistribution patterns administered with home-use compared with medical jet injectors and to assess safety aspects.MATERIALS AND METHODSThe authors injected HA into ex vivo human skin with 4 home-use and 2 medical injectors. The intracutaneous dose of HA was calculated, and the cutaneous biodistribution of HA was assessed using a 3-dimensional Fluorescent Imaging Cryomicrotome System (3D-FICS). Safety aspects were evaluated based on the presence of a manual, CE (conformite europeenne) mark, and sterility.RESULTSThe intracutaneous dose delivered by the home-use injectors was markedly lower compared with the medical injectors. 3D imaging for home-use injectors showed superficial epidermal distribution with low distribution volumes. For medical injectors, volumes were substantially larger and mainly middermal. All evaluated safety aspects were lacking.CONCLUSIONResults of this study suggest that the specific combinations of home-use injectors and HA used in this study are unreliable and unsafe, which casts doubts on the performance of these treatments in general.
引用
收藏
页码:62 / 68
页数:7
相关论文
共 50 条
  • [11] Home-use preventive and therapeutic oral products
    West, Nicola X.
    Moran, John M.
    [J]. PERIODONTOLOGY 2000, 2008, 48 : 7 - 9
  • [12] Central Location Test and Home-use Test
    Nakano, Yuko
    [J]. JOURNAL OF THE JAPANESE SOCIETY FOR FOOD SCIENCE AND TECHNOLOGY-NIPPON SHOKUHIN KAGAKU KOGAKU KAISHI, 2021, 68 (02): : 92 - 93
  • [13] Home-use oral hygiene products: mouthrinses
    Moran, John M.
    [J]. PERIODONTOLOGY 2000, 2008, 48 : 42 - 53
  • [14] Home-use HIV test kits in Turkey
    Doganci, L
    [J]. LANCET, 1997, 349 (9062): : 1401 - 1401
  • [15] Home-Use Intense Pulsed Light Device
    不详
    [J]. JOURNAL OF DRUGS IN DERMATOLOGY, 2009, 8 (06) : 595 - 595
  • [16] Acceptability of home-use of misoprostol in medical abortion
    Fiala, C
    Winikoff, B
    Helström, L
    Hellborg, M
    Gemzell-Danielsson, K
    [J]. CONTRACEPTION, 2004, 70 (05) : 387 - 392
  • [17] Survey of Home-Use UV Disinfection Products†
    Khazova, Marina
    Johnstone, Liam
    Naldzhiev, Dzhordzhio
    O'Hagan, John B.
    [J]. PHOTOCHEMISTRY AND PHOTOBIOLOGY, 2021, 97 (03) : 560 - 565
  • [18] Home-use cancer detecting band aid
    Zalevsky, Zeev
    Rudnitsky, Arkady
    Sheinman, Victor
    Tzoy, Andrey
    Toktosunov, Aitmamat
    Adashov, Arkady
    [J]. OPTICAL METHODS FOR TUMOR TREATMENT AND DETECTION: MECHANISMS AND TECHNIQUES IN PHOTODYNAMIC THERAPY XXV, 2016, 9694
  • [19] USE OF JET-INJECTORS IN BCG VACCINATION
    DAM, HGT
    FILLASTR.C
    CONGE, G
    ORSSAUD, E
    GATEFF, C
    TANAKA, A
    ORTEGARA.O
    COLLAS, R
    WRIGHT, J
    CHAMBON, L
    BARME, M
    TOMMASI, UB
    SARRAT, H
    BRES, P
    DIALLO, L
    GAUTHIER, M
    PIOT, M
    GULD, J
    [J]. BULLETIN OF THE WORLD HEALTH ORGANIZATION, 1970, 43 (05) : 707 - &
  • [20] Home-use pulse signal instrument system design
    Liu, Q.
    Gao, L.
    Yang, G.
    [J]. Wuhan Gongye Daxue Xuebao/Journal of Wuhan University of Technology, 2001, 23 (04): : 48 - 50