Effects of Ion Concentration on Ion Current Rectification Using a Glass Nanocapillary

被引:0
|
作者
Ito, Chisato [1 ]
Ino, Kosuke [2 ]
Hiramoto, Kaoru [3 ]
Nashimoto, Yuji [4 ]
Shiku, Hitoshi [1 ,2 ]
机构
[1] Tohoku Univ, Grad Sch Environm Studies, 6-6-11 Aoba,Aoba Ku, Sendai, Miyagi 9808579, Japan
[2] Tohoku Univ, Grad Sch Engn, 6-6-11 Aoba,Aoba Ku, Sendai, Miyagi 9808579, Japan
[3] Tohoku Univ, Frontier Res Inst Interdisciplinary Sci, 2-1-1 Katahira,Aoba Ku, Sendai, Miyagi 9808577, Japan
[4] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, 2-3-10 Kanda Surugadai,Chiyoda Ku, Tokyo 1010062, Japan
关键词
ion current rectification; glass nanocapillary; nanopore;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
When a glass capillary with a nanopore at its tip is filled with a solution of low ion concentration, the ion concentration at the tip is higher than expected. This is because the ions form an electrical double layer inside the wall. In this case, the ion currents conducted via the nanopore differ dramatically depending on the polarity of the applied potential. This phenomenon is known as ion current rectification (ICR). ICR shows great potential for applications in bioassay methodology, as it is heavily influenced by the amount of surface charges on the inside wall of the nanopore, allowing it to be seamlessly applied in biosensing procedures. However, ICR can only be induced under low ion concentrations, preventing it from being applied for biosensing under physiological conditions. To address this issue, the effects of the ion concentration of phosphate buffer solutions (PBS) in a glass nanocapillary and bulk outside of the capillary on ICR were investigated in this study. Three ion concentrations were introduced into the capillary and the bulk, and ion current via the nanopore was measured. The results show that ICR was successfully monitored even though PBS of a physiological condition was used in either the glass nanocapillary or bulk. In addition, cell spheroids were monitored inside the capillary as a preliminary study for biosensing. This finding can be utilized for ICR-based biosensing applications under a physiological condition.
引用
收藏
页码:117 / 123
页数:7
相关论文
共 50 条
  • [1] Ion current rectification of glass conical nanopore
    College of Engineering, Nanjing Agricultural University, Nanjing, China
    不详
    J. Comput. Theor. Nanosci., 10 (3994-3998):
  • [2] Ion current rectification at nanopores in glass membranes
    White, Henry S.
    Bund, Andreas
    LANGMUIR, 2008, 24 (05) : 2212 - 2218
  • [3] Effects of salt concentration and pH value on ion rectification of conical glass nanopore
    Chen, Y. (yunfeichen@seu.edu.cn), 1600, Southeast University (44):
  • [4] Ion current rectification in combination with ion current saturation
    Liu, Guo-Chang
    Song, Lai-Bo
    Wang, Xiao-Hong
    Li, Chao-Qing
    Liu, Bo
    Zhao, Yuan-Di
    Chen, Wei
    ANALYTICA CHIMICA ACTA, 2020, 1117 : 35 - 40
  • [5] Ion Current Rectification Behavior at Novel Borosilicate Glass Capillaries
    Silver, Barry
    Holub, Karel
    Marecek, Vladimir
    XXXII. MODERNI ELEKTROCHEMICKE METODY, 2012, : 120 - 124
  • [6] Effect of Electrolyte Concentration and Pore Size on Ion Current Rectification Inversion
    Duleba, Dominik
    Dutta, Pallavi
    Denuga, Shekemi
    Johnson, Robert P.
    ACS MEASUREMENT SCIENCE AU, 2022, 2 (03): : 271 - 277
  • [7] Modulating ion current rectification generating high energy output in a single glass conical nanopore channel by concentration gradient
    Zhang, Li-Xiang
    Zheng, Yu-Bin
    Cai, Sheng-Lin
    Cao, Xiao-Hong
    Li, Yao-Qun
    CHINESE CHEMICAL LETTERS, 2015, 26 (01) : 43 - 46
  • [8] Modulating ion current rectification generating high energy output in a single glass conical nanopore channel by concentration gradient
    Li-Xiang Zhang
    Yu-Bin Zheng
    Sheng-Lin Cai
    Xiao-Hong Cao
    Yao-Qun Li
    ChineseChemicalLetters, 2015, 26 (01) : 43 - 46
  • [9] Marked ion current rectification in microchannels
    Xing-Hua Xia
    Science China Chemistry, 2017, 60 : 685 - 686
  • [10] Marked ion current rectification in microchannels
    Xing-Hua Xia
    Science China(Chemistry), 2017, 60 (06) : 685 - 686