Advances in High-Energy-Resolution CdZnTe Linear Array Pixel Detectors with Fast and Low Noise Readout Electronics

被引:12
|
作者
Mele, Filippo [1 ,2 ]
Quercia, Jacopo [1 ]
Abbene, Leonardo [3 ]
Benassi, Giacomo [4 ]
Bettelli, Manuele [5 ]
Buttacavoli, Antonino [3 ]
Principato, Fabio [3 ]
Zappettini, Andrea [5 ]
Bertuccio, Giuseppe [1 ,2 ]
机构
[1] Politecn Milan, Dept Elect Informat & Bioengn DEIB, Via Anzani 42, I-22100 Como, Italy
[2] Natl Inst Nucl Phys INFN, Sez Milano, Via Celoria 16, I-22133 Milan, Italy
[3] Univ Palermo, Dept Phys & Chem DiFC Emilio Segre, Viale Sci,Edificio 18, I-90128 Palermo, Italy
[4] Due2lab srl, Via Paolo Borsellino 2, I-42019 Scandiano, Italy
[5] CNR, IMEM, Parco Area Sci 37-A, I-43100 Parma, Italy
关键词
CZT; CdZnTe; X-ray spectroscopy; Gamma-ray spectroscopy; semiconductor radiation detectors; nuclear microelectronics; GAMMA-RAY SPECTROSCOPY; X-RAY; CHARGE AMPLIFIERS; CDTE; PREAMPLIFIER; OPTIMIZATION; PERFORMANCE; SIRIO;
D O I
10.3390/s23042167
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Radiation detectors based on Cadmium Zinc Telluride (CZT) compounds are becoming popular solutions thanks to their high detection efficiency, room temperature operation, and to their reliability in compact detection systems for medical, astrophysical, or industrial applications. However, despite a huge effort to improve the technological process, CZT detectors' full potential has not been completely exploited when both high spatial and energy resolution are required by the application, especially at low energies (<10 keV), limiting their application in energy-resolved photon counting (ERPC) systems. This gap can also be attributed to the lack of dedicated front-end electronics which can bring out the best in terms of detector spectroscopic performances. In this work, we present the latest results achieved in terms of energy resolution using SIRIO, a fast low-noise charge sensitive amplifier, and a linear-array pixel detector, based on boron oxide encapsulated vertical Bridgman-grown B-VB CZT crystals. The detector features a 0.25-mm pitch, a 1-mm thickness and is operated at a -700-V bias voltage. An equivalent noise charge of 39.2 el. r.m.s. (corresponding to 412 eV FWHM) was measured on the test pulser at 32 ns peaking time, leading to a raw resolution of 1.3% (782 eV FWHM) on the 59 keV line at room temperature (+20 degrees C) using an uncollimated Am-241, largely improving the current state of the art for CZT-based detection systems at such short peaking times, and achieving an optimum resolution of 0.97% (576 eV FWHM) at 1 mu s peaking time. The measured energy resolution at the 122 keV line and with 1 mu s peaking time of a Co-57 raw uncollimated spectrum is 0.96% (1.17 keV). These activities are in the framework of an Italian collaboration on the development of energy-resolved X-ray scanners for material recycling, medical applications, and non-destructive testing in the food industry.
引用
收藏
页数:12
相关论文
共 20 条
  • [1] High resolution CdZnTe pixel detectors with VLSI readout
    Cook, WR
    Boggs, SE
    Bolotnikov, AE
    Burnham, JA
    Fitzsimmons, MJ
    Harrison, FA
    Kecman, B
    Matthews, B
    Schindler, SM
    1999 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1999, : 529 - 532
  • [2] High resolution CdZnTe pixel detectors with VLSI readout
    Cook, WR
    Boggs, SE
    Bolotnikov, AE
    Burnham, JA
    Fitzsimmons, MJ
    Harrison, FA
    Kecman, B
    Matthews, B
    Schindler, SM
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2000, 47 (04) : 1454 - 1457
  • [3] High Energy Resolution Pixel Detectors Based on Boron Oxide Vertical Bridgman Grown CdZnTe Crystals
    Zappettini, Andrea
    Macera, Daniele
    Benassi, Giacomo
    Zambelli, Nicola
    Calestani, Davide
    Ahangarianabhari, M.
    Shi, Yongbiao
    Rotondo, Giuseppe
    Garavelli, Bruno
    Pozzi, Pietro
    Bertuccio, Giuseppe
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [4] Fast, Low-Noise, Low-Power Electronics for the Analog Readout of Non-Linear DEPFET Pixels
    Facchinetti, Stefano
    Bombelli, Luca
    Castoldi, Andrea
    Fiorini, Carlo
    Guazzoni, Chiara
    Mezza, Davide
    Porro, Matteo
    De Vita, Giulio
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 1846 - 1851
  • [5] Development of a low-noise readout ASIC for Silicon Drift Detectors in high energy resolution X-ray spectrometry
    Atkin, E.
    Levin, V.
    Malankin, E.
    Shumikhin, V.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [6] Low-noise electronics for high-resolution CdZnTe and CdTe X-ray detection systems
    Niemela, A
    JOURNAL DE PHYSIQUE IV, 1996, 6 (C4): : 721 - 731
  • [7] Linear-mode linear arrays 16 pixel silicon avalanche photodiodes with high gain and low noise readout
    Chen, Jian
    Yang, Yunxiu
    Guo, Changdong
    Kong, Fanlin
    Kou, Xianguo
    Wang, Ou
    Yu, Libo
    Song, Haizhi
    AOPC 2017: OPTICAL SENSING AND IMAGING TECHNOLOGY AND APPLICATIONS, 2017, 10462
  • [8] Optimization of low-noise read-out electronics for high energy resolution X-ray strip detectors
    Zubrzycka, W.
    Grybos, P.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (01):
  • [9] Preliminary test results of a new high energy resolution silicon & CdZnTe pixel detectors for application to X-ray astronomy
    Sushkov, VV
    Hamilton, WJ
    Hurley, K
    Maeding, DG
    Ogelman, H
    Paulos, RJ
    Puetter, RC
    Tümer, TO
    Zweerink, J
    EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY X, 1999, 3765 : 379 - 387
  • [10] Low-power optimisation of a pixel array architecture for next generation High Energy Physics detectors
    Marconi, Sara
    Hemperek, Tomasz
    Placidi, Pisana
    Scorzoni, Andrea
    Conti, Elia
    Christiansen, Jorgen
    2017 13TH CONFERENCE ON PH.D. RESEARCH IN MICROELECTRONICS AND ELECTRONICS (PRIME), 2017, : 201 - 204