Development and clinical evaluation of an online automated quality control system for improving laboratory quality management

被引:2
|
作者
Shang, Xuesong [1 ,2 ,3 ]
Zhao, Fang [1 ,2 ,3 ]
Du, Juan [1 ,2 ,3 ]
Li, Rui [1 ,2 ,3 ]
Xia, Liangyu [1 ,2 ,3 ]
Hou, Lian [1 ,2 ,3 ]
Cheng, Xinqi [1 ,2 ,3 ]
Ding, Yingying [1 ,2 ,3 ]
Li, Peng [1 ,2 ,3 ]
Liu, Meizhi [1 ,2 ,3 ]
Ma, Chaochao [1 ,2 ,3 ]
Wang, Zhe [1 ,2 ,3 ]
Ding, Jinwen [1 ,2 ,3 ]
Ge, Zengmei [1 ,2 ,3 ]
Wang, Guanhua [1 ,2 ,3 ]
Guo, Tai [1 ,2 ,3 ]
Fu, Yu [4 ]
Xiao, Meng [1 ,2 ,3 ]
Yang, Qiwen [1 ,2 ,3 ]
Qiu, Ling [1 ,2 ,3 ]
Xu, Yingchun [1 ,2 ,3 ]
机构
[1] Peking Union Med Coll & Chinese Acad Med Sci, Peking Union Med Coll Hosp, Dept Lab Med, Beijing 100730, Peoples R China
[2] Chinese Acad Med Sci, Peking Union Med Coll Hosp, Dept State Key Lab Complex Severe & Rare Dis, Beijing, Peoples R China
[3] Peking Union Med Coll, Beijing, Peoples R China
[4] Sightnovo Med Technol Co Ltd, Beijing, Peoples R China
关键词
Automated quality control; Internal quality control; Quality management; Six Sigma; TAT;
D O I
10.1016/j.cca.2023.117240
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: We developed an efficient online automated quality control (AUTO QC) system and tested its feasibility on automatic laboratory assembly lines.Methods: AUTO QC is based on developed quality control software (Smart QC) and designed adaptable con-sumables before. We applied the system to two assembly lines in our laboratory. Using third-party quality control samples, we evaluated the impact of the online AUTO QC system on out-of-control rate, biosecurity risk, turn-around time (TAT) and cost.Results: AUTO QC significantly decreased the occurrence rate of the Westgard quality control rules 13S/22S/R4s and 12S, representing out-of-control and warning, respectively. The out-of-control rates were reduced by 58%, and the potential biosecurity risk of the samples decreased by 90%. The AUTO QC implementation also reduced the median TAT (by 7 min), the number of full-time employees and the cost of the quality control samples (by 45%). Conclusions: The total laboratory AUTO QC system can improve the quality and stability of QC testing and reduce cost.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Development of an automated quality control system of confectionery using a vision system
    Parkot, Katarzyna
    Sioma, Andrzej
    PHOTONICS APPLICATIONS IN ASTRONOMY, COMMUNICATIONS, INDUSTRY, AND HIGH-ENERGY PHYSICS EXPERIMENTS 2018, 2018, 10808
  • [22] The Impact of Knowledge Management System on the Quality of Test Result in Quality Control and Calibration Laboratory
    Hadian, Amir
    Vaghasloo, Younes Alizadeh
    Doroud, Reza
    2014 IEEE INTERNATIONAL TECHNOLOGY MANAGEMENT CONFERENCE (ITMC), 2014,
  • [23] QUALIDAT - A SYSTEM FOR QUALITY-CONTROL AND LABORATORY INFORMATION MANAGEMENT
    SASKO, P
    WELSCH, N
    FETT WISSENSCHAFT TECHNOLOGIE-FAT SCIENCE TECHNOLOGY, 1989, 91 (13): : 564 - 568
  • [24] LABORATORY DATA MANAGEMENT-SYSTEM WITH QUALITY-CONTROL
    GILLESPIE, WT
    AMERICAN LABORATORY, 1982, 14 (02) : 106 - &
  • [25] Development of an Automated Remote Asphalt Paving Quality Control System
    Zhu, Sicong
    Li, Xiangdi
    Wang, Haoyang
    Yu, Dongxiao
    TRANSPORTATION RESEARCH RECORD, 2018, 2672 (26) : 28 - 39
  • [26] Improving Construction Quality Control and Management Using Automated Data Collection Technologies
    Hedayatnasab, Arastoo
    Majrouhi Sardroud, Javad
    Limbachiya, Mukesh
    IMCIC'11: THE 2ND INTERNATIONAL MULTI-CONFERENCE ON COMPLEXITY, INFORMATICS AND CYBERNETICS, VOL I, 2011, : 24 - 29
  • [27] Quality management and the clinical microbiology laboratory
    Baron, EJ
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 1995, 23 (1-2) : 23 - 34
  • [28] QUALITY-CONTROL IN A CLINICAL CHEMICAL LABORATORY WITH ONLINE CONNECTION TO A PROCESS COMPUTER
    SCHMULLING, RM
    GRASER, W
    EGGSTEIN, M
    ANNALES DE BIOLOGIE CLINIQUE, 1975, 33 (03) : 200 - 200
  • [29] Development of an Online Quality Control System for Injection Molding Process
    Tsai, Ming-Hong
    Fan-Jiang, Jia-Chen
    Liou, Guan-Yan
    Cheng, Feng-Jung
    Hwang, Sheng-Jye
    Peng, Hsin-Shu
    Chu, Hsiao-Yeh
    POLYMERS, 2022, 14 (08)
  • [30] EVALUATION OF A MICRO COMPUTER-SYSTEM FOR HANDLING QUALITY-CONTROL DATA IN THE CLINICAL LABORATORY
    MACFARLANE, RE
    AGGARWAL, V
    CLINICAL CHEMISTRY, 1980, 26 (07) : 998 - 999