A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP

被引:1
|
作者
Coppadoro, Andrea [1 ]
Bellani, Giacomo [2 ,3 ]
Foti, Giuseppe [1 ,4 ]
机构
[1] San Gerardo Hosp, Fdn IRCCS San Gerardo Tintori, Dept Anesthesia & Intens care, Monza, Italy
[2] Univ Trento, Ctr Med Sci CISMed, Trento, Italy
[3] St Chiara Reg Hosp, Dept Anesthesia & Intens Care, APSS Trento, Trento, Italy
[4] Univ Milano Bicocca, Dept Med & Surg, Via Cadore 48, Monza, MB, Italy
关键词
Noninvasive respiratory support; Continuous flow Helmet CPAP; Acute hypoxemic respiratory failure; Tidal volume respiratory monitoring; Spontaneous breathing; VENTILATION;
D O I
10.1007/s10877-023-01034-0
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
PurposeThe coronavirus disease 2019 (COVID-19) pandemic has promoted the use of helmet continuous positive airway pressure (CPAP) for noninvasive respiratory support in hypoxic respiratory failure patients, despite the lack of tidal volume monitoring. We evaluated a novel technique designed to measure tidal volume during noninvasive continuous-flow helmet CPAP.MethodsA bench model of spontaneously breathing patients undergoing helmet CPAP therapy (three positive end-expiratory pressure [PEEP] levels) at different levels of respiratory distress was used to compare measured and reference tidal volumes. Tidal volume measurement by the novel technique was based on helmet outflow-trace analysis. Helmet inflow was increased from 60 to 75 and 90 L/min to match the patient's peak inspiratory flow; an additional subset of tests was conducted under the condition of purposely insufficient inflow (i.e., high respiratory distress and 60 L/min inflow).ResultsThe tidal volumes examined herein ranged from 250 to 910 mL. The Bland-Altman analysis showed a bias of -3.2 +/- 29.3 mL for measured tidal volumes compared to the reference, corresponding to an average relative error of -1 +/- 4.4%. Tidal volume underestimation correlated with respiratory rate (rho = .411, p = .004) but not with peak inspiratory flow, distress, or PEEP. When the helmet inflow was maintained purposely low, tidal volume underestimation occurred (bias - 93.3 +/- 83.9 mL), corresponding to an error of -14.8 +/- 6.3%.ConclusionTidal volume measurement is feasible and accurate during bench continuous-flow helmet CPAP therapy by the analysis of the outflow signal, provided that helmet inflow is adequate to match the patient's inspiratory efforts. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm these findings.
引用
收藏
页码:1473 / 1479
页数:7
相关论文
共 50 条
  • [21] Real time noninvasive estimation of work of breathing using facemask leak-corrected tidal volume during noninvasive pressure support: validation study
    Banner, Michael J.
    Tams, Carl G.
    Euliano, Neil R.
    Stephan, Paul J.
    Leavitt, Trevor J.
    Martin, A. Daniel
    Al-Rawas, Nawar
    Gabrielli, Andrea
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2016, 30 (03) : 285 - 294
  • [22] Measurement of Blood Pressure during Support with a Continuous-Flow Left Ventricular Assist Device in the Outpatient Setting
    Coyle, L. A.
    Gallagher, C.
    Maier, J.
    Bhat, G.
    Pappas, P.
    Tatooles, A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (04): : S235 - S235
  • [23] Predictors of Out-of-Therapeutic-Range INR during Support with Continuous-Flow Left Ventricular Assist Device
    Yin, M.
    Schultz, W.
    Ko, Y.
    Cole, R. T.
    Gupta, D.
    Laskar, S.
    Smith, A.
    Vega, D.
    Nguyen, D.
    Pekarek, A.
    Wittersheim, K.
    Morris, A.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (04): : S84 - S85
  • [24] Interventricular dyssynchrony during continuous-flow left ventricular assist device support: observation using the conductance method
    Shimamura, Junichi
    Nishimura, Takashi
    Mizuno, Toshihide
    Takewa, Yoshiaki
    Tsukiya, Tomonori
    Inatomi, Ayako
    Katagiri, Nobumasa
    Ando, Masahiko
    Umeki, Akihide
    Akiyama, Daichi
    Arakawa, Mamoru
    Ono, Minoru
    Tatsumi, Eisuke
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2019, 22 (04) : 348 - 352
  • [25] B-Blockers Improves Overall Survival during Support with Continuous-Flow Left Ventricular Assist Device
    Yin, Michael
    Ko, Yi-An
    Pekarek, Ann
    Wittersheim, Kris
    Cole, Robert T.
    Gupta, Divya
    Duc Nguyen
    Laskar, Raja
    Smith, Andrew
    Vega, J. David
    Morris, Alanna A.
    [J]. JOURNAL OF CARDIAC FAILURE, 2016, 22 (08) : S21 - S21
  • [26] Matrix-Degrading Enzyme Expression and Aortic Fibrosis During Continuous-Flow Left Ventricular Mechanical Support
    V. Ambardekar, Amrut
    Stratton, Matthew S.
    Dobrinskikh, Evgenia
    Hunter, Kendall S.
    Tatman, Philip D.
    Lemieux, Madeleine E.
    Cleveland, Joseph C.
    Tuder, Rubin M.
    Weiser-Evans, Mary C. M.
    Moulton, Karen S.
    McKinsey, Timothy A.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 78 (18) : 1782 - 1795
  • [27] Aortic regurgitation during continuous-flow left ventricular assist device support: An insufficient understanding of an insufficient lesion
    Cowger, Jennifer
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2016, 35 (08): : 973 - 975
  • [28] INTERMACS Analysis of Stroke During Support With Continuous-Flow Left Ventricular Assist Devices Risk Factors and Outcomes
    Acharya, Deepak
    Loyaga-Rendon, Renzo
    Morgan, Charity J.
    Sands, Kara A.
    Pamboukian, Salpy V.
    Rajapreyar, Indranee
    Holman, William L.
    Kirklin, James K.
    Tallaj, Jose A.
    [J]. JACC-HEART FAILURE, 2017, 5 (10) : 703 - 711
  • [29] Bleeding and Angiogenesis During Continuous-Flow Left Ventricular Assist Device Support Enter the von Willebrand Matrix
    Saeed, Omar
    Patel, Snehal R.
    Jorde, Ulrich P.
    [J]. CIRCULATION-HEART FAILURE, 2018, 11 (09)
  • [30] Interventricular dyssynchrony during continuous-flow left ventricular assist device support: observation using the conductance method
    Junichi Shimamura
    Takashi Nishimura
    Toshihide Mizuno
    Yoshiaki Takewa
    Tomonori Tsukiya
    Ayako Inatomi
    Nobumasa Katagiri
    Masahiko Ando
    Akihide Umeki
    Daichi Akiyama
    Mamoru Arakawa
    Minoru Ono
    Eisuke Tatsumi
    [J]. Journal of Artificial Organs, 2019, 22 : 348 - 352