A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP

被引:1
|
作者
Coppadoro, Andrea [1 ]
Bellani, Giacomo [2 ,3 ]
Foti, Giuseppe [1 ,4 ]
机构
[1] San Gerardo Hosp, Fdn IRCCS San Gerardo Tintori, Dept Anesthesia & Intens care, Monza, Italy
[2] Univ Trento, Ctr Med Sci CISMed, Trento, Italy
[3] St Chiara Reg Hosp, Dept Anesthesia & Intens Care, APSS Trento, Trento, Italy
[4] Univ Milano Bicocca, Dept Med & Surg, Via Cadore 48, Monza, MB, Italy
关键词
Noninvasive respiratory support; Continuous flow Helmet CPAP; Acute hypoxemic respiratory failure; Tidal volume respiratory monitoring; Spontaneous breathing; VENTILATION;
D O I
10.1007/s10877-023-01034-0
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
PurposeThe coronavirus disease 2019 (COVID-19) pandemic has promoted the use of helmet continuous positive airway pressure (CPAP) for noninvasive respiratory support in hypoxic respiratory failure patients, despite the lack of tidal volume monitoring. We evaluated a novel technique designed to measure tidal volume during noninvasive continuous-flow helmet CPAP.MethodsA bench model of spontaneously breathing patients undergoing helmet CPAP therapy (three positive end-expiratory pressure [PEEP] levels) at different levels of respiratory distress was used to compare measured and reference tidal volumes. Tidal volume measurement by the novel technique was based on helmet outflow-trace analysis. Helmet inflow was increased from 60 to 75 and 90 L/min to match the patient's peak inspiratory flow; an additional subset of tests was conducted under the condition of purposely insufficient inflow (i.e., high respiratory distress and 60 L/min inflow).ResultsThe tidal volumes examined herein ranged from 250 to 910 mL. The Bland-Altman analysis showed a bias of -3.2 +/- 29.3 mL for measured tidal volumes compared to the reference, corresponding to an average relative error of -1 +/- 4.4%. Tidal volume underestimation correlated with respiratory rate (rho = .411, p = .004) but not with peak inspiratory flow, distress, or PEEP. When the helmet inflow was maintained purposely low, tidal volume underestimation occurred (bias - 93.3 +/- 83.9 mL), corresponding to an error of -14.8 +/- 6.3%.ConclusionTidal volume measurement is feasible and accurate during bench continuous-flow helmet CPAP therapy by the analysis of the outflow signal, provided that helmet inflow is adequate to match the patient's inspiratory efforts. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm these findings.
引用
收藏
页码:1473 / 1479
页数:7
相关论文
共 50 条
  • [1] A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP
    Andrea Coppadoro
    Giacomo Bellani
    Giuseppe Foti
    [J]. Journal of Clinical Monitoring and Computing, 2023, 37 : 1473 - 1479
  • [2] In Vivo Measurement of Tidal Volume During Non-invasive Respiratory Support by Continuous-Flow Helmet CPAP
    Lomauro, A.
    De Luca, A.
    Scarpazza, P.
    Aliverti, A.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2024, 52 (09) : 2546 - 2555
  • [3] Tidal Volume Estimation during Helmet Noninvasive Ventilation: an Experimental Feasibility Study
    Cortegiani, Andrea
    Navalesi, Paolo
    Accurso, Giuseppe
    Sabella, Ignazio
    Misseri, Giovanni
    Ippolito, Mariachiara
    Bruni, Andrea
    Garofalo, Eugenio
    Palmeri, Cesira
    Gregoretti, Cesare
    [J]. SCIENTIFIC REPORTS, 2019, 9 (1)
  • [4] Tidal Volume Estimation during Helmet Noninvasive Ventilation: an Experimental Feasibility Study
    Andrea Cortegiani
    Paolo Navalesi
    Giuseppe Accurso
    Ignazio Sabella
    Giovanni Misseri
    Mariachiara Ippolito
    Andrea Bruni
    Eugenio Garofalo
    Cesira Palmeri
    Cesare Gregoretti
    [J]. Scientific Reports, 9
  • [5] Nasal High Flow Delivered within the Helmet: A New Noninvasive Respiratory Support
    Mauri, Tommaso
    Spinelli, Elena
    Mariani, Massimiliano
    Guzzardella, Amedeo
    Del Prete, Chiara
    Carlesso, Eleonora
    Tortolani, Donatella
    Tagliabue, Paola
    Pesenti, Antonio
    Grasselli, Giacomo
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2019, 199 (01) : 115 - 117
  • [6] Noninvasive Measures of Pulsatility and Blood Pressure During Continuous-Flow Left Ventricular Assist Device Support
    Rangasamy, Sabarivinoth
    Madan, Shivank
    Saeed, Omar
    Goldstein, Daniel J.
    Jorde, Ulrich P.
    Negassa, Abdissa
    Patel, Snehal R.
    [J]. ASAIO JOURNAL, 2019, 65 (03) : 241 - 246
  • [7] Flow-volume curves as measure of patients' respiratory mechanics during ventilatory support.
    Lourens, MS
    van den Berg, B
    Hoogsteden, HC
    Bogaard, JM
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 159 (03) : A363 - A363
  • [8] Defining pulsatility during continuous-flow ventricular assist device support
    Soucy, Kevin G.
    Koenig, Steven C.
    Giridharan, Guruprasad A.
    Sobieski, Michael. A.
    Slaughter, Mark S.
    [J]. JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2013, 32 (06): : 581 - 587
  • [9] Effects of face mask (FM) and pressure helmet (PH) on work of breathing (WOB) and rebreathing during noninvasive continuous positive airway pressure (CPAP)
    Zito, G
    Banner, MJ
    Blanch, P
    Hardcastle, J
    Gabrielli, A
    [J]. CRITICAL CARE MEDICINE, 2003, 31 (12) : A103 - A103
  • [10] Quantification of interventricular dyssynchrony during continuous-flow left ventricular assist device support
    Shimamura, Junichi
    Nishimura, Takashi
    Mizuno, Toshihide
    Takewa, Yoshiaki
    Tsukiya, Tomonori
    Inatomi, Ayako
    Ando, Masahiko
    Umeki, Akihide
    Naito, Noritsugu
    Ono, Minoru
    Tatsumi, Eisuke
    [J]. JOURNAL OF ARTIFICIAL ORGANS, 2019, 22 (04) : 269 - 275