Randomized multicenter trial to assess posterior capsule opacification and glistenings in two hydrophobic acrylic intraocular lenses

被引:7
|
作者
Auffarth, Gerd U. [1 ]
Brezin, Antoine [2 ]
Lignereux, Francois [3 ]
Khoramnia, Ramin [1 ]
Yildirim, Timur M. [1 ]
Kohnen, Thomas [4 ]
Bianco, Jessica [5 ]
机构
[1] Heidelberg Univ, Heidelberg, Germany
[2] Cochin Hosp, Paris, France
[3] Inst Ophtalmol Sourdille Atlantique, Saint-herblain, France
[4] Goethe Univ, Frankfurt, Germany
[5] HOYA Surg Opt GmbH, Frankfurt, Germany
关键词
SURFACE LIGHT-SCATTERING; VISUAL FUNCTION; STRAYLIGHT; DESIGN;
D O I
10.1038/s41598-023-29855-8
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
To evaluate the long-term posterior capsule opacification (PCO) formation, and glistening rate of the HOYA Vivinex (XY1) IOL compared to Alcon AcrySof (SN60WF). In this prospective, multicentric, randomized, paired-eye, open-label study, we included 87 subjects that underwent cataract surgery with IOL implantation, with 67 patients completing the 3-year follow-up. The completer population consisted of 32 subjects implanted with XY1 and 35 implanted with SN60WF. Primary endpoints consisted of the evaluation of glistenings and measurement of PCO. Secondary outcomes included Best Corrected Distance Visual Acuity (BCVA), Contrast Acuity (CA), uncorrected visual acuities, subjective refraction, medical and lens complication rates, adverse events, and optical/visual symptoms. Follow-up visits occurred at 6-months, 1-, 2- and 3-years. At 3-years follow-up, mean PCO score was 0.121 +/- 0.193 for eyes implanted with Vivinex versus 0.239 +/- 0.463 for AcrySof (p = 0.026). The Vivinex IOL showed statistically significantly lower glistening occurrence through 3-years postoperatively (0.14 +/- 0.26) compared to AcrySof (1.79 +/- 1.43; p < 0.0001). Postoperative visual acuities improved from baseline in both IOL groups (p < 0.0001), and remained stable through the 3-year follow-up period. Eyes implanted with a HOYA Vivinex IOL exhibited significantly lower occurrence of glistening at 3-years versus Alcon AcrySof (p < 0.0001). Incidence of PCO was very low and comparable in both Vivinex and AcrySof eyes.
引用
收藏
页数:9
相关论文
共 50 条
  • [11] Findings of an experimental study in a rabbit model on posterior capsule opacification after implantation of hydrophobic acrylic and hydrophilic acrylic intraocular lenses
    Trakos, Nikolaos
    Ioachim, Elli
    Tsanou, Elena
    Aspiotis, Miltiadis
    Psilas, Konstantinos
    Kalogeropoulos, Chris
    CLINICAL OPHTHALMOLOGY, 2008, 2 (04): : 997 - 1005
  • [12] Glistenings in the single-piece, hydrophobic, acrylic intraocular lenses
    Waite, Aaron
    Faulkner, Nathan
    Olson, Randall J.
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (01) : 143 - 144
  • [13] Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance
    Lei Xi
    Yi Liu
    Feng Zhao
    Chuan Chen
    Bing Cheng
    International Journal of Ophthalmology, 2014, (03) : 446 - 451
  • [14] Analysis of glistenings in hydrophobic acrylic intraocular lenses on visual performance
    Xi, Lei
    Liu, Yi
    Zhao, Feng
    Chen, Chuan
    Cheng, Bing
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2014, 7 (03) : 446 - 451
  • [15] Effect of round-edged acrylic intraocular lenses on preventing posterior capsule opacification
    Nishi, O
    Nishi, K
    Akura, J
    Nagata, T
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (04): : 608 - 613
  • [16] Glistenings in hydrophobic acrylic intraocular lenses do affect visual function
    Beiko, George H. H.
    Grzybowski, Andrzej
    CLINICAL OPHTHALMOLOGY, 2013, 7 : 2271 - +
  • [17] Posterior Capsule Opacification in Silicone and Hydrophobic Acrylic Intraocular Lenses with Sharp-edge Optics Six Years After Surgery
    Vock, Lorenz
    Crnej, Alia
    Findl, Oliver
    Neumayer, Thomas
    Buehl, Wolf
    Sacu, Stefan
    Rainer, Georg
    Menapace, Rupert
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2009, 147 (04) : 683 - 690
  • [18] Posterior capsule opacification in rabbit eyes implanted with 1-piece and 3-piece hydrophobic acrylic intraocular lenses
    Werner, L
    Mamalis, N
    Izak, AM
    Pandey, SK
    Davis, BL
    Nilson, CD
    Weight, C
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (04): : 805 - 811
  • [19] Effect of types of hydrophilic acrylic monomers in reducing glistenings of hydrophobic acrylic intraocular lenses
    Kim, Tae Hyoung
    Song, Ki Chang
    OPTICAL MATERIALS, 2021, 119
  • [20] Relationship of AcrySof acrylic and PhacoFlex silicone intraocular lenses to visual acuity and posterior capsule opacification
    Halpern, MT
    Covert, D
    Battista, C
    Weinstein, AJ
    Levinson, RD
    Yan, L
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2002, 28 (04): : 662 - 669