A Geometrical Calibration Method for C-Arm CT Based on a Nonlinear Registration Model

被引:0
|
作者
Ji, Xu [1 ,2 ]
Lu, Yuchen [3 ]
Zhuo, Xu [3 ]
Zhang, Yikun [3 ]
Zhu, Shiyu [3 ]
Chen, Yang [1 ,2 ]
机构
[1] Southeast Univ, Sch Comp Sci & Engn, Sch Cyber Sci & Engn, Jiangsu Prov Joint Int Res Lab Med Informat Proc,L, Nanjing 210096, Peoples R China
[2] Southeast Univ, Key Lab New Generat Artificial Intelligence Techno, Minist Educ, Nanjing 210096, Peoples R China
[3] Southeast Univ, Sch Comp Sci & Engn, Lab Image Sci & Technol, Nanjing 210096, Peoples R China
基金
中国国家自然科学基金;
关键词
Index Terms- C-arm computed tomography (CT); cone beam CT; error analysis; geometrical calibration; image registration; CONE-BEAM CT; IMAGING-SYSTEMS; PARAMETERS; DETECTOR;
D O I
10.1109/TIM.2023.3307182
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The image quality of C-arm computed tomography (CT) degrades due to the nonideal projection geometry due to mechanical vibrations of the gantry during the CT scan; therefore, geometrical calibration methods need to be applied to derive the view-specific projection matrices, which are incorporated in the reconstruction processes. Previous geometrical calibration methods either require specific calibration phantoms or long computational time due to the 3-D-2-D registration process. This work begins with a theoretical derivation on how the perturbed geometrical parameters impact the projection images. It was found that with small perturbations of the view angle and the altitude angle, the relationship between the projection images under the ideal geometry and those under the perturbed geometry can be related based on a model-based 2-D deformation dependent on the perturbed geometrical parameters. Guided by the theoretical derivations, a new geometrical correction method was proposed, which generates the geometrical parameters based on 2-D-2-D registrations between the ideal projection images and the perturbed projection images. The proposed method was applied to offline and online calibration and was validated by both numerical experiments and physical experiments.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Statistical model registration for a C-arm CT system
    Fleute, M
    Desbat, L
    Martin, R
    Lavallée, S
    Defrise, M
    Liu, X
    Taylor, R
    [J]. 2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1667 - 1670
  • [2] Online geometrical calibration of a mobile C-arm using external sensors
    Mitschke, MM
    Navab, N
    Schütz, O
    [J]. MEDICAL IMAGING 2000: IMAGE DISPLAY AND VISUALIZATION, 2000, 3976 : 580 - 587
  • [3] Visual Model Based C-arm System Calibration and Image Correction
    Lv, Shaojie
    Meng, Cai
    Zhou, Fugen
    Liu, Bo
    Zhou, Xiaojun
    [J]. PROCEEDINGS OF THE 2009 2ND INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS, VOLS 1-4, 2009, : 454 - +
  • [4] C-arm calibration - Is it really necessary?
    Jain, A
    Kon, R
    Zhou, Y
    Fichtinger, G
    [J]. MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 1, 2005, 3749 : 639 - 646
  • [5] C-arm calibration - is it really necessary?
    Jain, Ameet Kumar
    An, Michael
    Chitphakdithai, Nicha
    Chintalapani, Gouthami
    Fichtinger, Gabor
    [J]. MEDICAL IMAGING 2007: VISUALIZATION AND IMAGE-GUIDED PROCEDURES, PTS 1 AND 2, 2007, 6509
  • [6] A model based algorithm for perfusion estimation in interventional C-arm CT systems
    Wagner, Martin
    Deuerling-Zheng, Yu
    Moehlenbruch, Markus
    Bendszus, Martin
    Boese, Jan
    Heiland, Sabine
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [7] An Optimized Spline-Based Registration of a 3D CT to a Set of C-Arm Images
    Jonic, S.
    Thevenaz, P.
    Zheng, G.
    Nolte, L. -P.
    Unser, M.
    [J]. INTERNATIONAL JOURNAL OF BIOMEDICAL IMAGING, 2006, 2006
  • [8] New method for geometric calibration and distortion correction of conventional C-arm
    Meng, Cai
    Zhang, Jun
    Zhou, Fugen
    Wang, Tianmiao
    [J]. COMPUTERS IN BIOLOGY AND MEDICINE, 2014, 52 : 49 - 56
  • [9] Geometric calibration of a mobile C-arm for intraoperative cone-beam CT
    Daly, M. J.
    Siewerdsen, J. H.
    Cho, Y. B.
    Jaffray, D. A.
    Irish, J. C.
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 2124 - 2136
  • [10] On-the-fly geometrical calibration fine-tuning of a mobile C-arm CBCT system
    El Hakimi, Wissam
    Sakas, Georgios
    [J]. TM-TECHNISCHES MESSEN, 2016, 83 (05) : 300 - 308