A computational model to simulate spectral modulation and speech perception experiments of cochlear implant users

被引:3
|
作者
Alvarez, Franklin [1 ,2 ]
Kipping, Daniel [1 ,2 ]
Nogueira, Waldo [1 ,2 ]
机构
[1] Hannover Med Sch, Hannover, Germany
[2] Cluster Excellence Hearing4All, Hannover, Germany
基金
欧洲研究理事会;
关键词
computational model; cochlear implant; neural health; sound coding strategies; speech-in-noise recognition; spectral modulation detection; speech understanding prediction; PULSATILE ELECTRICAL-STIMULATION; AUDITORY-NERVE; SPIRAL GANGLION; INTELLIGIBILITY; DISCRIMINATION; INTERFERENCE; DEGENERATION; RECOGNITION; EXCITATION; RESOLUTION;
D O I
10.3389/fninf.2023.934472
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Speech understanding in cochlear implant (CI) users presents large intersubject variability that may be related to different aspects of the peripheral auditory system, such as the electrode-nerve interface and neural health conditions. This variability makes it more challenging to proof differences in performance between different CI sound coding strategies in regular clinical studies, nevertheless, computational models can be helpful to assess the speech performance of CI users in an environment where all these physiological aspects can be controlled. In this study, differences in performance between three variants of the HiRes Fidelity 120 (F120) sound coding strategy are studied with a computational model. The computational model consists of (i) a processing stage with the sound coding strategy, (ii) a three-dimensional electrode-nerve interface that accounts for auditory nerve fiber (ANF) degeneration, (iii) a population of phenomenological ANF models, and (iv) a feature extractor algorithm to obtain the internal representation (IR) of the neural activity. As the back-end, the simulation framework for auditory discrimination experiments (FADE) was chosen. Two experiments relevant to speech understanding were performed: one related to spectral modulation threshold (SMT), and the other one related to speech reception threshold (SRT). These experiments included three different neural health conditions (healthy ANFs, and moderate and severe ANF degeneration). The F120 was configured to use sequential stimulation (F120-S), and simultaneous stimulation with two (F120-P) and three (F120-T) simultaneously active channels. Simultaneous stimulation causes electric interaction that smears the spectrotemporal information transmitted to the ANFs, and it has been hypothesized to lead to even worse information transmission in poor neural health conditions. In general, worse neural health conditions led to worse predicted performance; nevertheless, the detriment was small compared to clinical data. Results in SRT experiments indicated that performance with simultaneous stimulation, especially F120-T, were more affected by neural degeneration than with sequential stimulation. Results in SMT experiments showed no significant difference in performance. Although the proposed model in its current state is able to perform SMT and SRT experiments, it is not reliable to predict real CI users' performance yet. Nevertheless, improvements related to the ANF model, feature extraction, and predictor algorithm are discussed.
引用
收藏
页数:19
相关论文
共 50 条
  • [21] Bimodal speech perception in infant hearing aid and cochlear implant users
    Barker, BA
    Tomblin, JB
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2004, 130 (05) : 582 - 586
  • [22] Behavioral Measures of Temporal Processing and Speech Perception in Cochlear Implant Users
    Blankenship, Chelsea
    Zhang, Fawen
    Keith, Robert
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2016, 27 (09) : 701 - 713
  • [23] Impact of a Moving Noise Masker on Speech Perception in Cochlear Implant Users
    Weissgerber, Tobias
    Rader, Tobias
    Baumann, Uwe
    PLOS ONE, 2015, 10 (05):
  • [24] Cross-modal reorganization and speech perception in cochlear implant users
    Doucet, M. E.
    Bergeron, F.
    Lassonde, M.
    Ferron, P.
    Lepore, F.
    BRAIN, 2006, 129 : 3376 - 3383
  • [25] Improving speech perception in noise with current focusing in cochlear implant users
    Srinivasan, Arthi G.
    Padilla, Monica
    Shannon, Robert V.
    Landsberger, David M.
    HEARING RESEARCH, 2013, 299 : 29 - 36
  • [26] Ceiling effects for speech perception tests in pediatric cochlear implant users
    Spitzer, Emily R.
    Landsberger, David M.
    Lichtl, Alexandra J.
    Waltzman, Susan B.
    COCHLEAR IMPLANTS INTERNATIONAL, 2024, 25 (01) : 69 - 80
  • [27] Bilateral and Unilateral Cochlear Implant Users Compared on Speech Perception in Noise
    Dunn, Camille C.
    Noble, William
    Tyler, Richard S.
    Kordus, Monika
    Gantz, Bruce J.
    Ji, Haihong
    EAR AND HEARING, 2010, 31 (02): : 296 - 298
  • [28] Long-term Speech Perception in Elderly Cochlear Implant Users
    Dillon, Margaret T.
    Buss, Emily
    Adunka, Marcia C.
    King, English R.
    Pillsbury, Harold C., III
    Adunka, Oliver F.
    Buchman, Craig A.
    JAMA OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2013, 139 (03) : 279 - 283
  • [29] Evaluation of the Effect of Musical Perception Activities on Speech Perception in Adult Cochlear Implant Users
    Mutlu, Barak
    Topcu, Merve Torun
    Yuksel, Mustafa
    Kalcioglu, Mahmut Tayyar
    TURKISH ARCHIVES OF OTORHINOLARYNGOLOGY, 2022, 60 (04) : 188 - 198
  • [30] Deactivating cochlear implant electrodes to improve speech perception: A computational approach
    Sagi, Elad
    Svirsky, Mario A.
    HEARING RESEARCH, 2018, 370 : 316 - 328