Performance evaluation of the bio-rad BioPlex 2200 multiplex system in the detection of measles, mumps, rubella, and varicella-zoster antibodies

被引:0
|
作者
Lafreniere, Matthew A. [1 ]
Badr, Eley [2 ]
Beattie, John [2 ]
Macri, Joseph [2 ]
Khan, Waliul I. [2 ]
机构
[1] McMaster Univ, Dept Pathol & Mol Med, 1280 Main St West, Hamilton, ON L8S 4L8, Canada
[2] Hamilton Hlth Sci, Hamilton Reg Lab Med Program, 237 Barton St East, Hamilton, ON L8L2 X2, Canada
来源
JOURNAL OF CLINICAL VIROLOGY PLUS | 2023年 / 3卷 / 01期
关键词
Bio-Rad BioPlex 2200; MMRV; Bio-Rad Evolis Plus IgG MMV; Precision; Abbott Architect IgG Rubella; IGG; DISEASE;
D O I
10.1016/j.jcvp.2022.100131
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Background: Measles, mumps, rubella, and varicella-zoster (MMRV) immunity testing is important in occupational screening of healthcare workers and at-risk populations.Objectives: The goal of this study was to compare the performance of the Bio-Rad BioPlex 2200 MMRV multiplex fluorescence immunoassay (MFI) against two enzyme immunoassay (EIA) methods: the Bio-Rad Evolis Twin Plus measles, mumps, and varicella-zoster (MMV) IgG assay and the Abbott Architect Rubella IgG assay.Study Design: Clinically uncharacterized serum specimens were obtained and analyzed using the Bio-Rad BioPlex assay and compared against the EIA methods.Results: The Bio-Rad BioPlex demonstrated total agreement of 85.5% (95% confidence interval (CI), 78.0 to 90.7%), 92.7% (95% confidence interval (CI), 86.7 to 96.1%), 92.4% (95% confidence interval (CI), 86.9 to 95.7%), and 98.8% (95% confidence interval (CI), 93.7 to 99.8%) for measles, mumps, varicella-zoster, and rubella, respectively, against the current EIA methods. Furthermore, precision testing agreed with the manufac-turer's package insert in 10 of 13 pooled samples. Conclusion: These data indicate that the Bio-Rad BioPlex has comparable performance to the EIA methods.
引用
收藏
页数:4
相关论文
共 46 条
  • [1] Evaluation of the Bio-Rad BioPlex Measles, Mumps, Rubella, and Varicella-Zoster Virus IgG Multiplex Bead Immunoassay
    Binnicker, Matthew J.
    Jespersen, Deborah J.
    Rollins, Leonard O.
    CLINICAL AND VACCINE IMMUNOLOGY, 2011, 18 (09) : 1524 - 1526
  • [2] Comparison of a new multiplex BioPlex® 2200 assay for the detection of measles, mumps, and varicella-zoster virus IgM antibodies in human specimens
    King, D.
    Del Rosario, C. Renato
    JOURNAL OF CLINICAL VIROLOGY, 2015, 70 : S50 - S50
  • [3] Evaluation of three different sample populations on a new multiplex BioPlex® 2200 assay for the detection of measles, Mumps, and Varicella-Zoster virus IgM antibodies
    King, D.
    Del Rosario, R.
    Tatyosian, C.
    Todd, C.
    Flanagan, J.
    Vogel, J.
    JOURNAL OF CLINICAL VIROLOGY, 2016, 82 : S50 - S51
  • [4] Evaluation of the Bio-Rad BioPlex 2200 Toxoplasma gondii IgM Multiplex Flow Immunoassay
    Theel, Elitza S.
    Yarbrough, Melanie L.
    Hilgart, Heather
    Gronowski, Ann M.
    JOURNAL OF APPLIED LABORATORY MEDICINE, 2019, 3 (06): : 1022 - 1027
  • [5] Detection of IgG-class antibodies to measles, mumps, rubella, and varicella-zoster virus using a multiplex bead immunoassay
    Dhiman, Neelam
    Jespersen, Deborah J.
    Rollins, Leonard O.
    Harring, Julie A.
    Beito, Elaine M.
    Binnicker, Matthew J.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2010, 67 (04) : 346 - 349
  • [6] Evaluation of the Bio-Rad BioPlex 2200 Syphilis Multiplex Flow Immunoassay for the Detection of IgM- and IgG-Class Antitreponemal Antibodies
    Gomez, E.
    Jespersen, D. J.
    Harring, J. A.
    Binnicker, M. J.
    CLINICAL AND VACCINE IMMUNOLOGY, 2010, 17 (06) : 966 - 968
  • [7] Seroprevalence of measles, mumps, rubella, and varicella-zoster antibodies in new female nurses in the Republic of Korea
    Yoo, Yeongjae
    Park, Won-Ju
    Cho, Seunghyeon
    Lim, Dae-Young
    Kim, Suwhan
    Kang, Wonyang
    Kim, Hyeonjun
    Moon, Jai-Dong
    ANNALS OF OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 2021, 33
  • [8] Performance Characteristics of a Multiplex Flow Immunoassay for Detection of IgG-Class Antibodies to Measles, Mumps, Rubella, and Varicella-Zoster Viruses in Presumptively Immune Health Care Workers
    Theel, Elitza S.
    Sorenson, Marisa
    Rahman, Crystal
    Granger, Dane
    Vaughn, Andrew
    Breeher, Laura
    JOURNAL OF CLINICAL MICROBIOLOGY, 2020, 58 (04)
  • [9] Development of a Bead-Based Multiplex Immunoassay for Simultaneous Quantitative Detection of IgG Serum Antibodies against Measles, Mumps, Rubella, and Varicella-Zoster Virus
    Smits, Gaby P.
    van Gageldonk, Pieter G.
    Schouls, Leo M.
    van der Klis, Fiona R. M.
    Berbers, Guy A. M.
    CLINICAL AND VACCINE IMMUNOLOGY, 2012, 19 (03) : 396 - 400
  • [10] EVALUATION OF THE FULLY AUTOMATED BIO-RAD BIOPLEX 2200 SYPHILIS TOTAL & RPR ANTIBODY DETECTION ASSAY
    Boissonneault, Vincent
    Martel-Laferriere, Valerie
    Serhir, Bouchra
    Fortin, Claude
    SEXUALLY TRANSMITTED INFECTIONS, 2019, 95 : A326 - A326