An adaptive targeting algorithm for magnetic resonance-guided high-intensity focused ultrasound controlled hyperthermia

被引:3
|
作者
Wong, Suzanne M. [1 ,2 ]
Luo, Phoebe [1 ]
Keunen, Benjamin [1 ]
Pichardo, Samuel [3 ]
Drake, James M. [1 ,2 ]
Waspe, Adam C. [1 ,4 ]
机构
[1] Hosp Sick Children, Posluns Ctr Image Guided Innovat & Therapeut Inte, Toronto, ON M5G 1X8, Canada
[2] Univ Toronto, Inst Biomed Engn, Toronto, ON, Canada
[3] Univ Calgary, Hotchkiss Brain Inst, Calgary, AB, Canada
[4] Univ Toronto, Dept Med Imaging, Toronto, ON, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
controlled hyperthermia; high intensity focused ultrasound; magnetic resonance thermometry; ABLATION; TISSUE;
D O I
10.1002/mp.16414
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Mild hyperthermia has been demonstrated to improve the efficacy of chemotherapy, radiation, and immunotherapy in various cancer types. One localized, non-invasive method of administering mild hyperthermia is magnetic resonance-guided high-intensity focused ultrasound (MRgHIFU). However, challenges for ultrasound such as beam deflection, refraction and coupling issues may result in a misalignment of the HIFU focus and the tumor during hyperthermia. Currently, the best option is to stop the treatment, wait for the tissue to cool, and redo the treatment planning before restarting the hyperthermia. This current workflow is both time-consuming and unreliable. Purpose: An adaptive targeting algorithm was developed for MRgHIFU controlled hyperthermia treatments for cancer therapeutics. This algorithm executes in real time while hyperthermia is being administered to ensure that the focus is within our target region. If a mistarget is detected, the HIFU system will electronically steer the focus of the HIFU beam to the correct target. The goal of this study was to quantify the accuracy and precision of the adaptive targeting algorithm's ability to correct a purposely misplanned hyperthermia treatment in real-time using a clinical MRgHIFU system. Methods: A gelatin phantom with acoustic properties matched to the average speed of sound in human tissue was used to test the adaptive targeting algorithm's accuracy and precision. The target was purposely offset 10 mm away from the focus at the origin, in four orthogonal directions, allowing the algorithm to correct for this mistarget. In each direction, 10 data sets were collected for a total sample size of 40. Hyperthermia was administered with a target temperature set at 42 degrees C. The adaptive targeting algorithm was run during the hyperthermia treatment and 20 thermometry images were collected after the beam steering occurred. The location of the focus was quantified by calculating the center of heating on the MR thermometry data. Results: The average calculated trajectory passed to the HIFU system was 9.7 mm +/- 0.4 mm where the target trajectory was 10 mm. The accuracy of the adaptive targeting algorithm after the beam steering correction was 0.9 mm and the precision was 1.6 mm. Conclusion: The adaptive targeting algorithm was implemented successfully and was able to correct the 10 mm mistargets with high accuracy and precision in gelatin phantoms. The results demonstrate the capability to correct the MRgHIFU focus location during controlled hyperthermia.
引用
收藏
页码:3347 / 3358
页数:12
相关论文
共 50 条
  • [1] Mild hyperthermia with magnetic resonance-guided high-intensity focused ultrasound for applications in drug delivery
    Partanen, Ari
    Yarmolenko, Pavel S.
    Viitala, Antti
    Appanaboyina, Sunil
    Haemmerich, Dieter
    Ranjan, Ashish
    Jacobs, Genevieve
    Woods, David
    Enholm, Julia
    Wood, Bradford J.
    Dreher, Matthew R.
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2012, 28 (04) : 320 - 336
  • [2] Robotic Active Positioning For Magnetic Resonance-guided High-intensity Focused Ultrasound
    Xiao, Xu
    Huang, Zhihong
    Volovick, Alexander
    Melzer, Andreas
    12TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND, 2012, 1503 : 217 - 221
  • [3] Transcranial focused ultrasound. Neurological applications of magnetic resonance-guided high-intensity focused ultrasound
    Werner, B.
    Martin, E.
    RADIOLOGE, 2015, 55 (11): : 976 - 983
  • [4] Focused ultrasound using a novel targeting method four-tract tractography for magnetic resonance-guided high-intensity focused ultrasound targeting
    Feltrin, Fabricio S.
    Chopra, Rajiv
    Pouratian, Nader
    Elkurd, Mazen
    El-Nazer, Rasheda
    Lanford, Lauren
    Dauer, William
    Shah, Bhavya R.
    BRAIN COMMUNICATIONS, 2022, 4 (06)
  • [5] Magnetic resonance-guided interstitial high-intensity focused ultrasound for brain tumor ablation
    MacDonell, Jacquelyn
    Patel, Niravkumar
    Rubino, Sebastian
    Ghoshal, Goutam
    Fischer, Gregory
    Burdette, E. Clif
    Hwang, Roy
    Pilitsis, Julie G.
    NEUROSURGICAL FOCUS, 2018, 44 (02)
  • [6] Segmentation Method for Magnetic Resonance-Guided High-Intensity Focused Ultrasound Therapy Planning
    Vargas-Olivares, A.
    Navarro-Hinojosa, O.
    Maqueo-Vicencio, M.
    Curiel, L.
    Alencastre-Miranda, M.
    Chong-Quero, J. E.
    JOURNAL OF HEALTHCARE ENGINEERING, 2017, 2017
  • [7] Magnetic resonance-guided, high-intensity focused ultrasound sonolysis: potential applications for stroke
    Ilyas, Adeel
    Chen, Ching-Jen
    Ding, Dale
    Romeo, Andrew
    Buell, Thomas J.
    Wang, Tony R.
    Kalani, M. Yashar S.
    Park, Min S.
    NEUROSURGICAL FOCUS, 2018, 44 (02)
  • [8] Magnetic Resonance-Guided High-Intensity Focused Ultrasound Hyperthermia for Recurrent Rectal Cancer: MR Thermometry Evaluation and Preclinical Validation
    Chu, William
    Staruch, Robert M.
    Pichardo, Samuel
    Tillander, Matti
    Kohler, Max O.
    Huang, Yuexi
    Ylihautala, Mika
    McGuffin, Merrylee
    Czarnota, Gregory
    Hynynen, Kullervo
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 95 (04): : 1259 - 1267
  • [10] Feasibility of magnetic resonance-guided high-intensity focused ultrasound treatment targeting distinct nodular lesions in neurofibromatosis type 1
    Tydings, Caitlin
    Yarmolenko, Pavel
    Bornhorst, Miriam
    Dombi, Eva
    Myseros, John
    Keating, Robert
    Bost, James
    Sharma, Karun
    Kim, AeRang
    NEURO-ONCOLOGY ADVANCES, 2021, 3 (01)