Detection and size measurements of kidney stones on virtual non-contrast reconstructions derived from dual-layer computed tomography in an ex vivo phantom setup

被引:3
|
作者
Reimer, R. P. [1 ,2 ]
Zaytoun, H. [1 ,2 ]
Klein, K. [1 ,2 ]
Sonnabend, K. [1 ,2 ,3 ]
Lennartz, S. [1 ,2 ]
Zopfs, D. [1 ,2 ]
Heidenreich, A. [2 ,4 ]
Maintz, D. [1 ,2 ]
Hokamp, N. Grosse [1 ,2 ]
机构
[1] Univ Cologne, Fac Med, Dept Diagnost & Intervent Radiol, Kerpener Str 62, D-50937 Cologne, Germany
[2] Univ Cologne, Univ Hosp Cologne, Kerpener St 62, D-50937 Cologne, Germany
[3] Philips GmbH Market DACH, Hamburg, Germany
[4] Univ Cologne, Fac Med, Dept Urol, Cologne, Germany
关键词
Contrast media; Tomography; spiral computed; Kidney calculi; Radiation dosage; ENERGY CT UROGRAPHY; STATISTICAL ITERATIVE RECONSTRUCTION; IODINE QUANTIFICATION; UROLITHIASIS; ATTENUATION; DIAGNOSIS;
D O I
10.1007/s00330-022-09261-w
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives To systematically investigate the usability of virtual non-contrast reconstructions (VNC) derived from dual-layer CT (DLCT) for detection and size measurements of kidney stones with regards to different degrees of surrounding iodine-induced attenuation and radiation dose. Methods Ninety-two kidney stones of varying size (3-14 mm) and composition were placed in a phantom filled with different contrast media/water mixtures exhibiting specific iodine-induced attenuation (0-1500 HU). DLCT-scans were acquired using CTDIvol of 2 mGy and 10 mGy. Conventional images (CI) and VNC0H-1500HU were reconstructed. Reference stone size was determined using a digital caliper (Man-M). Visibility and stone size were assessed. Statistical analysis was performed using the McNemar test, Wilcoxon test, and the coefficient of determination. Results All stones were visible on CI0HU and VNC200HU. Starting at VNC400 HU, the detection rate decreased with increasing HU and was significantly lower as compared to CI0HU on VNC >= 600HU (100.0 vs. 94.0%, p < 0.05). The overall detection rate was higher using 10 mGy as compared to 2 mGy protocol (87.9 vs. 81.8%; p < 0.001). Stone size was significantly overestimated on all VNC compared to Man-M (7.0 +/- 3.5 vs. 6.6 +/- 2.8 mm, p < 0.001). Again, the 10 mGy protocol tended to show a better correlation with Man-M as compared to 2 mGy protocol (R-2 = 0.39-0.68 vs. R-2 = 0.31-0.57). Conclusions Detection and size measurements of kidney stones surrounded by contrast media on VNC are feasible. The detection rate of kidney stones decreases with increasing iodine-induced attenuation and with decreasing radiation dose as well as stone size, while remaining comparable to CI0HU on VNC (<= 400 HU).
引用
收藏
页码:2995 / 3003
页数:9
相关论文
共 28 条
  • [1] Detection and size measurements of kidney stones on virtual non-contrast reconstructions derived from dual-layer computed tomography in an ex vivo phantom setup
    R. P. Reimer
    H. Zaytoun
    K. Klein
    K. Sonnabend
    S. Lennartz
    D. Zopfs
    A. Heidenreich
    D. Maintz
    N. Große Hokamp
    European Radiology, 2023, 33 : 2995 - 3003
  • [2] Dual-layer spectral computed tomography: Virtual non-contrast in comparison to true non-contrast images
    Sauter, Andreas P.
    Muenzel, Daniela
    Dangelmaier, Julia
    Braren, Rickmer
    Pfeiffer, Franz
    Rummeny, Ernst J.
    Noel, Peter B.
    Fingerle, Alexander A.
    EUROPEAN JOURNAL OF RADIOLOGY, 2018, 104 : 108 - 114
  • [3] Comparison of virtual and true non-contrast images from dual-layer spectral detector computed tomography (CT) in patients with colorectal cancer
    Wen, Deying
    Pu, Qian
    Peng, Pengfei
    Yue, Xun
    Ming, Yue
    Yang, Huiyi
    Yang, Jianyang
    Zhang, Xiaodi
    Liu, Haiwei
    Yang, Lie
    Sun, Jiayu
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (09)
  • [4] Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study
    Jasmin A. Holz
    Hatem Alkadhi
    Kai R. Laukamp
    Simon Lennartz
    Carola Heneweer
    Michael Püsken
    Thorsten Persigehl
    David Maintz
    Nils Große Hokamp
    Scientific Reports, 10
  • [5] Quantitative accuracy of virtual non-contrast images derived from spectral detector computed tomography: an abdominal phantom study
    Holz, Jasmin A.
    Alkadhi, Hatem
    Laukamp, Kai R.
    Lennartz, Simon
    Heneweer, Carola
    Puesken, Michael
    Persigehl, Thorsten
    Maintz, David
    Hokamp, Nils Grosse
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [6] Coronary calcium scoring using virtual non-contrast reconstructions on a dual-layer spectral CT system: Feasibility in the clinical practice
    Langenbach, I. L.
    Wienemann, H.
    Klein, K.
    Scholtz, J. E.
    Pennig, L.
    Langzam, E.
    Pahn, G.
    Holz, J. A.
    Maintz, D.
    Naehle, C. P.
    Langenbach, M. C.
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 159
  • [7] Protocol analysis of dual-energy CT for optimization of kidney stone detection in virtual non-contrast reconstructions
    Matthias Lazar
    Helmut Ringl
    Pascal Baltzer
    Daniel Toth
    Christian Seitz
    Bernhard Krauss
    Ewald Unger
    Stephan Polanec
    Dietmar Tamandl
    Christian J. Herold
    Michael Toepker
    European Radiology, 2020, 30 : 4295 - 4305
  • [8] Protocol analysis of dual-energy CT for optimization of kidney stone detection in virtual non-contrast reconstructions
    Lazar, Matthias
    Ringl, Helmut
    Baltzer, Pascal
    Toth, Daniel
    Seitz, Christian
    Krauss, Bernhard
    Unger, Ewald
    Polanec, Stephan
    Tamandl, Dietmar
    Herold, Christian J.
    Toepker, Michael
    EUROPEAN RADIOLOGY, 2020, 30 (08) : 4295 - 4305
  • [9] Accuracy of Calcium Scoring calculated from contrast-enhanced Coronary Computed Tomography Angiography using a dual-layer spectral CT: A comparison of Calcium Scoring from real and virtual non-contrast data
    Nadjiri, Jonathan
    Kaissis, Georgios
    Meurer, Felix
    Weis, Florian
    Laugwitz, Karl-Ludwig
    Streeter, Alexandra S.
    Muenzel, Daniela
    Noel, Peter B.
    Rummeny, Ernst J.
    Rasper, Michael
    PLOS ONE, 2018, 13 (12):
  • [10] Clinical applications of virtual, non-contrast head images derived from dual-source, dual-energy cerebrovascular computed tomography angiography
    Han, D.
    Xie, X. J.
    Wen, L.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (02): : 159 - 163