Improved gray-white matter contrast using magnetization prepared fast imaging with steady-state free precession (MP-FISP) brain imaging at 0.55 T

被引:2
|
作者
Schaper, Jessica [1 ,2 ,3 ]
Bauman, Grzegorz [1 ,2 ]
Bieri, Oliver [1 ,2 ]
机构
[1] Univ Basel, Dept Biomed Engn, Basel, Switzerland
[2] Univ Hosp Basel, Dept Radiol, Div Radiol Phys, Basel, Switzerland
[3] Univ Hosp Basel, Dept Radiol, Div Radiol Phys, Petersgraben 4, CH-4031 Basel, Switzerland
基金
瑞士国家科学基金会;
关键词
brain; FISP; GM; WM contrast; low-field; MPRAGE; RAGE; SEGMENTATION; OPTIMIZATION;
D O I
10.1002/mrm.29838
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo improve the gray/white matter contrast of magnetization prepared rapid gradient echo (MP-RAGE) MRI at 0.55 T by optimizing the acquisition and sequence kernel parameters. MethodsA segmented magnetization prepared rapid gradient echo prototype sequence was implemented with (MP-RAGE*) and without (MP-FISP*) radiofrequency spoiling. Optimized parameters were derived with the assistance of an extended phase graph signal simulation as a function of the relaxation times, the flip angle, the delay times, and the effective inversion time using segmentation. The resulting protocols were compared to the MP-RAGE product sequence offered by the vendor in terms of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). A tissue segmentation reproducibility study was performed on three volunteers for the product MP-RAGE and the MP-FISP*. ResultsThe MP-RAGE simulation reproduced the parameters already used in the product MP-RAGE on the scanner. An average CNR improvement of 15% for the custom MP-RAGE* over the product MP-RAGE and additional 22% for the MP-FISP* over the MP-RAGE* were observed, which is in accordance with the simulation results. The total improvement, averaged over all volunteers and regions, was 41%. The reproducibility study did not yield a significant difference between MP-RAGE and MP-FISP*. ConclusionWe presented some easy-to-implement adjustments to the MP-RAGE sequence at 0.55 T, which can lead to an overall average improvement of 41% in CNR.
引用
收藏
页码:162 / 173
页数:12
相关论文
共 50 条
  • [31] Feasibility of fetal cardiac function and anatomy assessment by real-time spiral balanced steady-state free precession magnetic resonance imaging at 0.55T
    Tian, Ye
    Detterich, Jon
    Pruetz, Jay D.
    Yagiz, Ecrin
    Wood, John C.
    Nayak, Krishna S.
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2025, 27 (01)
  • [32] Magnetic resonance portography using contrast-enhanced fat-saturated three-dimensional steady-state free precession imaging
    Amano, Y
    Takahama, K
    Nozaki, A
    Amano, M
    Kumazaki, T
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (02) : 238 - 244
  • [33] Cine Cardiac Imaging Using Black-Blood Steady-State Free Precession (BB-SSFP) at 3T
    Basha, Tamer A.
    Ibrahim, El-Sayed H.
    Weiss, Robert G.
    Osman, Nael F.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (01) : 94 - 103
  • [34] Microstructural correlates of 3D steady-state inhomogeneous magnetization transfer (ihMT) in the human brain white matter assessed by myelin water imaging and diffusion tensor imaging
    Ercan, Ece
    Varma, Gopal
    Maedler, Burkhard
    Dimitrov, Ivan E.
    Pinho, Marco C.
    Xi, Yin
    Wagner, Benjamin C.
    Davenport, Elizabeth M.
    Maldjian, Joseph A.
    Alsop, David C.
    Lenkinski, Robert E.
    Vinogradov, Elena
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (06) : 2402 - 2414
  • [35] Magnetic resonance angiography of coronary arteries with a 3-dimensional magnetization-prepared true fast imaging with steady-state precession sequence compared with conventional coronary angiography
    Man-ching, N
    Lam, WW
    Li, D
    Chan, AK
    Sanderson, JE
    Metreweli, C
    AMERICAN HEART JOURNAL, 2005, 150 (03) : 530 - 535
  • [36] Transverse Relaxation Time (T2) Mapping in the Brain With Off-Resonance Correction Using Phase-Cycled Steady-State Free Precession Imaging
    Deoni, Sean C. L.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 30 (02) : 411 - 417
  • [37] T2-prepared steady-state free precession blood oxygen level-dependent MR imaging of myocardial perfusion in a dog stenosis model
    Shea, SM
    Fieno, DS
    Schirf, BE
    Bi, XM
    Huang, J
    Omary, RA
    Li, DB
    RADIOLOGY, 2005, 236 (02) : 503 - 509
  • [38] Comparison of 3D-diffusion-prepared segmented steady-state free precession and 2D fast spin echo imaging of femoral artery atherosclerosis
    Hayashi, Katsumi
    Mani, Venkatesh
    Nemade, Ajay
    Silvera, Stephane
    Fayad, Zahi A.
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2010, 26 (03): : 309 - 321
  • [39] Comparison of 3D-diffusion-prepared segmented steady-state free precession and 2D fast spin echo imaging of femoral artery atherosclerosis
    Katsumi Hayashi
    Venkatesh Mani
    Ajay Nemade
    Stephane Silvera
    Zahi A. Fayad
    The International Journal of Cardiovascular Imaging, 2010, 26 : 309 - 321
  • [40] Improved cine displacement-encoded MRI using balanced steady-state free precession and time-adaptive sensitivity encoding parallel imaging at 3T
    Kim, Daniel
    Kellman, Peter
    NMR IN BIOMEDICINE, 2007, 20 (06) : 591 - 601