Pulse pileup model for spectral resolved X-ray photon-counting detectors with dead time and retrigger capability

被引:1
|
作者
Zambon, P. [1 ]
Amato, C. [1 ]
机构
[1] DECTRIS Ltd, Baden, Switzerland
关键词
counting detector; non-paralyzable; retrigger; dead time; pulse pileup; spectral distortion; CT; ENERGY; PERFORMANCE; READOUT;
D O I
10.3389/fphy.2023.1205638
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
We developed an analytical model to evaluate the effect of signal pileup on the recorded energy spectrum in X-ray photon-counting detectors affected by dead time and equipped with retrigger capability. The retrigger function allows the system to work in a specific non-paralyzable counting mode by counting the time-over-threshold of piled-up signals in multiples of a predefined and selectable retrigger time. The model, designed for rectangle-like-shaped signals, allows for arbitrary input energy spectra and can significantly help understand and optimize the behavior of counting detectors with spectral capabilities and retrigger mechanisms in applications involving polychromatic beams, e.g., spectral X-ray imaging and computed tomography (CT), in a time-efficient way. Dedicated numerical simulations were used to validate the model under several conditions of incoming flux and threshold energy, with excellent results.
引用
收藏
页数:10
相关论文
共 50 条
  • [31] Absolute dose calibration of an X-ray system and dead time investigations of photon-counting techniques
    Carpentieri, C
    Schwarz, C
    Ludwig, J
    Ashfaq, A
    Fiederle, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 487 (1-2): : 71 - 77
  • [32] Sensitivity Analysis of Pulse Pileup Model Parameter in Photon Counting Detectors
    Shunhavanich, Picha
    Pelc, Norbert J.
    MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132
  • [33] Energy-sensitive scatter estimation and correction for spectral x-ray imaging with photon-counting detectors
    Lewis, Cale E.
    Das, Mini
    JOURNAL OF INSTRUMENTATION, 2022, 17 (12)
  • [34] Development of a pixel readout ASIC for CZT detectors for spectral X-ray photon-counting imaging applications
    Wang, X.
    Deng, Z.
    Liu, W.
    Liu, Y.
    JOURNAL OF INSTRUMENTATION, 2020, 15 (01):
  • [35] Spectral X-Ray CT Image Reconstruction with a Combination of Energy-Integrating and Photon-Counting Detectors
    Yang, Qingsong
    Cong, Wenxiang
    Xi, Yan
    Wang, Ge
    PLOS ONE, 2016, 11 (05):
  • [36] Noise in energy-discriminating photon-counting x-ray imaging detectors
    Tanguay, Jesse
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622
  • [37] Characterization of energy response for photon-counting detectors using x-ray fluorescence
    Ding, Huanjun
    Cho, Hyo-Min
    Barber, William C.
    Iwanczyk, Jan S.
    Molloi, Sabee
    MEDICAL PHYSICS, 2014, 41 (12)
  • [38] Assessment of the spectral performance of hybrid photon counting x-ray detectors
    Trueb, Peter
    Zambon, Pietro
    Broennimann, Christian
    MEDICAL PHYSICS, 2017, 44 (09) : E207 - E214
  • [39] Quasi-monochromatic imaging in x-ray CT via spectral deconvolution using photon-counting detectors
    Wang, Sen
    Gao, Hewei
    Zhang, Li
    Wu, Dufan
    Xu, Xiaofei
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (06): : 2208 - 2223
  • [40] Dead-time correction for spectroscopic photon-counting pixel detectors
    Blaj, Gabriel