High amplitude pulses on the same charge condition efficiently elicit bipolar cell-mediated retinal ganglion cell responses in the degenerate retina

被引:3
|
作者
Ahn, Jungryul [1 ]
Jeong, Yurim [1 ]
Cha, Seongkwang [1 ]
Lee, Joo Yong [2 ]
Yoo, Yongseok [3 ]
Goo, Yong Sook [1 ]
机构
[1] Chungbuk Natl Univ, Dept Physiol, Sch Med, Cheongju, South Korea
[2] Univ Ulsan, Asan Med Ctr, Dept Ophthalmol, Coll Med, Seoul, South Korea
[3] Soongsil Univ, Sch Comp Sci & Engn, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Retinal prosthesis; Retinal degeneration; Pde6b knockout rat; Electrical stimulation; Retinal ganglion cell; Stimulus parameters; ELECTRICAL-STIMULATION; MOUSE; PIGMENTOSA; TISSUE; MODEL;
D O I
10.1007/s13534-023-00260-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Retinal pigmentosa (RP) patients lose vision due to the loss of photoreceptors. Retinal prostheses bypass the dead photoreceptors by electrically stimulating surviving retinal neurons, such as bipolar cells or retinal ganglion cells (RGCs). In previous studies, stimulus charge has been mainly optimized to maximize the RGC response to electrical stimulation. This study aimed to investigate the effect of amplitude and duration even under the same charge condition on eliciting RGC spikes in the wild-type and degenerate retinas. Wild-type (WT) Sprague-Dawley rats were used as the normal retinal model, and Pde6b knockout rats were used as a retinal degeneration (RD) model. Electrically-evoked RGC spikes were recorded from isolated rat retinas using an 8 x 8 multielectrode array. The same charge was maintained (10 or 20 nC), and electrical stimulation was applied to WT and RD retinas, adjusting the amplitude and duration of the 1st phase of biphasic pulses. In the pulse modulation of the 1st phase, high amplitude (short duration) pulses induced more RGC spikes than low amplitude (long duration) pulses. Both WT and RD retinas showed a significant reduction in the number of RGC spikes upon stimulation with lower amplitude (longer duration) pulses. In clinical trials where stimulus charges are delivered to the degenerate retina of blind patients, high amplitude (short duration) pulses would help elicit more RGC spikes.
引用
收藏
页码:129 / 140
页数:12
相关论文
共 1 条
  • [1] High amplitude pulses on the same charge condition efficiently elicit bipolar cell-mediated retinal ganglion cell responses in the degenerate retina
    Jungryul Ahn
    Yurim Jeong
    Seongkwang Cha
    Joo Yong Lee
    Yongseok Yoo
    Yong Sook Goo
    Biomedical Engineering Letters, 2023, 13 : 129 - 140