Validation of a Monte Carlo model of the uEXPLORER total-body PET scanner using GATE code

被引:0
|
作者
EL Katib, Mahmoud [1 ,2 ,5 ]
Chakir, El Mahjoub [1 ]
Sebihi, Rajaa [3 ]
Saikouk, Hind [3 ,4 ]
Nhila, Oussama [1 ]
机构
[1] Ibn Tofail Univ, Fac Sci, Phys Dept, Mat & Subatom Phys Lab LPMS, Kenitra, Morocco
[2] Mohammed VI Univ Hosp, Oncol & Hematol Hosp, Radiotherapy Dept, Marrakech, Morocco
[3] Mohammed V Univ, Fac Sci, Phys Dept, ESMAR Team, Rabat, Morocco
[4] Mohammed VI Univ Hosp, Oncol & Hematol Hosp, Nucl Med Dept, Marrakech, Morocco
[5] Univ Ibn Tofail, Dept Phys, Lab Phys Mat & Subatom, Fac Sci, Campus Univ,BP 133, Kenitra, Morocco
关键词
PET; Monte Carlo simulations; GATE; Performance; uEXPLORER; FIELD-OF-VIEW; AXIAL-FIELD; SIMULATION; PERFORMANCE; SYSTEMS; IMPACT; IMAGE;
D O I
10.1016/j.radphyschem.2023.111052
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Purpose: To develop and validate a GATE model of the newly manufactured total-body PET scanner (uEXPLORER), in order to evaluate the capability of GATE to simulate PET scanners with long AFOV, and to provide a reliable model to be used in further optimization studies. Methods: The detector geometry of our model consists of 564,480 LYSO crystals grouped into blocks and modules in accordance with the actual scanner's specifications. Signal processing chain was modeled directly in GATE through the digitizer modules. For model validation, we performed spatial resolution, sensitivity, count rate and scatter fraction tests following the National Electrical Manufactures Association (NEMA) NU 2-2018 protocol, in addition to other devised tests representative of total-body acquisitions. The simulation results were then compared to published measurements. Results: Overall, our simulation model demonstrated good agreement with the experimental results, within 6% for sensitivity, 7% for scatter fraction and 17% for noise equivalent count rate. For the spatial resolution test, the model matched the experimental results to within 0.1-1.3 mm. Conclusions: We have shown that our developed model predicts the uEXPLORER PET scanner's performances with good accuracy. Consequently, we conclude that GATE is reasonably suitable for simulating long AFOV PET scanners.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Development of a Monte Carlo-based scatter correction method for total-body PET using the uEXPLORER PET/CT scanner
    Bayerlein, Reimund
    Spencer, Benjamin A.
    Leung, Edwin K.
    Omidvari, Negar
    Abdelhafez, Yasser G.
    Wang, Qian
    Nardo, Lorenzo
    Cherry, Simon R.
    Badawi, Ramsey D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (04):
  • [2] Development and Validation of a Monte Carlo Simulation Workflow for a Total-Body PET Scanner
    Pommranz, C.
    Elmoujarkach, E.
    Cabello, J.
    Lan, W.
    Rafecas, M.
    Mannheim, J.
    Linder, P.
    Santangelo, A.
    la Fougere, C.
    Pichler, B.
    Schmidt, F.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2023, 50 (SUPPL 1) : S690 - S691
  • [3] Quantitative accuracy in total-body imaging using the uEXPLORER PET/CT scanner
    Leung, Edwin K.
    Berg, Eric
    Omidvari, Negar
    Spencer, Benjamin A.
    Li, Elizabeth
    Abdelhafez, Yasser G.
    Schmall, Jeffrey P.
    Liu, Weiping
    He, Liuchun
    Tang, Songsong
    Liu, Yilin
    Dong, Yun
    Jones, Terry
    Cherry, Simon R.
    Badawi, Ramsey D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (20):
  • [4] Accurate modeling and performance evaluation of a total-body pet scanner using Monte Carlo simulations
    Rezaei, Hadi
    Sheikhzadeh, Peyman
    Ghafarian, Pardis
    Zaidi, Habib
    Ay, Mohammad Reza
    [J]. MEDICAL PHYSICS, 2023, 50 (11) : 6815 - 6827
  • [5] Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE
    Lu, Lijun
    Zhang, Houjin
    Bian, Zhaoying
    Ma, Jianhua
    Feng, Qiangjin
    Chen, Wufan
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 : 170 - 175
  • [6] Validation of a Clinical PET Scanner Using Monte Carlo Simulation Code: MCNP5
    Musarudin, Marianie
    Saripan, M. Iqbal
    Mashohor, Syamsiah
    Saad, Wira Hidayat Mohd
    Hashim, Suhairul
    Nordin, Abdul Jalil
    [J]. 2013 8TH EUROSIM CONGRESS ON MODELLING AND SIMULATION (EUROSIM), 2013, : 36 - 41
  • [7] Assessment of PET scanner SNR, and CNR factors: validation of Gate Monte Carlo code and STIR reconstruction with experimental study
    Karimipourfard, M.
    Sina, S.
    Sadeghi, M.
    Shobeiry, M.
    Safari, R.
    Alavi, M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2022, 49 (SUPPL 1) : S598 - S598
  • [8] Monte-Carlo simulation of the photodetection systems prototype PET scanner using GATE: A validation study
    Groiselle, CJ
    Kudrolli, HA
    Glick, SJ
    [J]. 2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3130 - 3132
  • [9] Considerations for Positioning Arthritic Patients on the uExplorer Total-Body PET/CT System
    Hunt, Heather
    Nguyen, Mikey
    Caudle, Denise
    Emerson, Nikki
    Sarkar, Soumajyoti
    Abdelhafez, Yasser
    Nardo, Lorenzo
    Cherry, Simon
    Badawi, Ramsey
    Raychaudhuri, Siba
    Chaudhari, Abhijit
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2020, 61
  • [10] Monte Carlo validation of the microPET FOCUS PET scanner using FLUKA
    Toufique, Y.
    Ferrari, A.
    Bouhali, O.
    Ortega, P. G.
    Augusto, R. Santos
    Vlachoudis, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S102 - S102