Variability in oxygen delivery with bag-valve-mask devices: An observational laboratory simulation study

被引:5
|
作者
Dodds, Jarron M. [1 ,5 ]
Appelqvist, Dylan, I [2 ]
Paleologos, Michael S. [3 ,4 ]
Downey, Ryan G. [3 ,4 ]
机构
[1] St Vincents Hosp, Dept Anaesthesia & Acute Pain Med, Melbourne, Australia
[2] Univ Notre Dame, Sydney, Australia
[3] Royal Prince Alfred Hosp, Dept Anaesthesia, Sydney, Australia
[4] Univ Sydney, Sydney Med Sch, Sydney, Australia
[5] St Vincents Hosp Melbourne, Dept Anaesthesia & Acute Pain Med, 41 Victoria Parade, Fitzroy, Vic 3065, Australia
关键词
Bag-valve-mask; BVM; self-inflating resuscitation device; non-rebreather valve; pre-oxygenation; oxygen delivery; emergency intubation; PRE-OXYGENATION; SPONTANEOUS VENTILATION; PREOXYGENATION; ANESTHESIA; VOLUME; ADULT;
D O I
10.1177/0310057X221119824
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A bag-valve-mask (BVM) is a portable handheld medical device commonly used in airway management and manual ventilation. Outside of the operating theatre, BVM devices are often used to pre-oxygenate spontaneously breathing patients before intubation to reduce the risk of hypoxaemia. Pre-oxygenation is considered adequate when the end-tidal expiratory fraction of oxygen is greater than 0.85. There are reports that some BVM devices fail to deliver a satisfactory inspired oxygen (FiO(2)) in spontaneously breathing patients due to variability in design. The primary aim of this study was to evaluate the efficacy of oxygen delivery of a broad range of adult and paediatric BVM devices at increasing tidal volumes using a mechanical lung to simulate spontaneous ventilation. The secondary aim was to evaluate the effect of BVM design on performance. Forty BVM devices were evaluated in a laboratory setting as part of a safety assessment requested by HealthShare New South Wales. The oxygen inlet of each BVM device was primed with 100% oxygen (15 l/min) for two min. The BVM device was then attached to the mechanical lung and commenced spontaneous breathing at a fixed respiratory rate of 12 breaths/min with an inspiratory: expiratory ratio of 1:2. For each device FiO(2) was measured after two min of spontaneous breathing. This process was repeated with small (250 ml), medium (500 ml) and large (750 ml) tidal volumes simulating adult breathing in adult BVM devices, and small (150 ml), medium (300 ml) and large (450 ml) tidal volumes simulating paediatric breathing in paediatric BVM devices. The test was repeated using up to five BVM devices of the same model (where supplied) at each tidal volume as a manufacturing quality control measure. Eight of the 40 devices tested failed to deliver a FiO(2) above 0.85 for at least one tidal volume, and five models failed to achieve this at any measured tidal volume. Concerningly, three of these devices delivered a FiO(2) below 0.55. Six of the eight poorly performing devices delivered reducing concentrations of inspired oxygen with increasing tidal volumes. Devices which performed the worst were those with a duckbill non-rebreather valve and without a dedicated expiratory valve. Several BVM devices available for clinical use in Australia did not deliver sufficient oxygen for reliable pre-oxygenation in a spontaneously breathing in vitro model. Devices with a duckbill non-rebreather valve and without a dedicated expiratory valve performed the worst. It is imperative that clinicians using BVM devices to deliver oxygen to spontaneously breathing patients are aware of the characteristics and limitations of the BVM devices, and that the standards for manufacture are updated to require safe performance in all clinical circumstances.
引用
收藏
页码:130 / 140
页数:11
相关论文
共 50 条
  • [31] Efficacy of Bag-valve-mask Ventilation for Ventricular Fibrillation Sudden Cardiac Arrest
    Tahara, Yoshio
    Kimura, Kazuo
    Yonemoto, Naohiro
    Nonogi, Hiroshi
    Kimura, Takeshi
    Nagao, Ken
    CIRCULATION, 2013, 128 (22)
  • [32] Comparison of Bag-Valve-Mask Hand-Sealing Techniques in a Simulated Model
    Braude, Darren Alan
    Tawil, Isaac
    Gerstein, Neal Stuart
    Carey, Michael Christopher
    Petersen, Timothy Randal
    ANNALS OF EMERGENCY MEDICINE, 2014, 63 (06) : 784 - 785
  • [33] AIRWAY MANAGEMENT IN UNSKILLED PERSONNEL - A COMPARISON OF LARYNGEAL MASK AIRWAY, POCKET MASK AND BAG-VALVE-MASK TECHNIQUES
    HAYES, A
    MCCARROLL, SM
    ANESTHESIOLOGY, 1995, 83 (3A) : A223 - A223
  • [34] Airway management during cardiopulmonary resuscitation - a comparative study of bag-valve-mask, laryngeal mask airway and combitube in a bench model
    Doerges, V
    Sauer, C
    Ocker, H
    Wenzel, V
    Schmucker, P
    RESUSCITATION, 1999, 41 (01) : 63 - 69
  • [35] VENTILATORY VOLUMES USING MOUTH-TO-MOUTH, MOUTH-TO-MASK, AND BAG-VALVE-MASK TECHNIQUES
    HESS, D
    BARAN, C
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1985, 3 (04): : 292 - 296
  • [36] From Mouth-to-Mouth to Bag-Valve-Mask Ventilation: Evolution and Characteristics of Actual Devices-A Review of the Literature
    Khoury, Abdo
    Hugonnot, Sylvere
    Cossus, Johan
    De Luca, Alban
    Desmettre, Thibaut
    Sall, Fatimata Seydou
    Capellier, Gilles
    BIOMED RESEARCH INTERNATIONAL, 2014, 2014
  • [37] Bag-Valve-Mask Ventilation and Survival From Out-of-Hospital Cardiac Arrest: A Multicenter Study
    Idris, Ahamed H.
    Ecenarro, Elisabete Aramendi
    Leroux, Brian
    Jaureguibeitia, Xabier
    Yang, Betty Y.
    Shaver, Sarah
    Chang, Mary P.
    Rea, Tom
    Kudenchuk, Peter
    Christenson, Jim
    Vaillancourt, Christian
    Callaway, Clifton
    Salcido, David
    Carson, Jonas
    Blackwood, Jennifer
    Wang, Henry E.
    CIRCULATION, 2023, 148 (23) : 1847 - 1856
  • [38] Bag-Valve-Mask Ventilation and Survival From Out-of-Hospital Cardiac Arrest: A Multicenter Study
    Idris, Ahamed H.
    Aramendi, Elisabete
    Leroux, Brian
    Jaureguibeitia, Xabier
    Yang, Betty
    Shaver, Sarah
    Chang, Mary
    Rea, Thomas
    Kudenchuk, Peter
    Christenson, Jim
    Vaillancourt, Christian
    Callaway, Clifton W.
    Salcido, David D.
    Carson, Jonas
    Blackwood, Jennifer
    Wang, Henry E.
    CIRCULATION, 2023, 148 (25) : E311 - E312
  • [39] An evaluation of bag-valve-mask ventilation using an ergonomically designed facemask among novice users: A simulation-based pilot study
    Bauman, Eric B.
    Joffe, Aaron M.
    Lenz, Lynn
    DeVries, Stephen A.
    Hetzel, Scott
    Seider, Samuel P.
    RESUSCITATION, 2010, 81 (09) : 1161 - 1165
  • [40] Airway management during cardiopulmonary resuscitation - a comparative study of bag-valve-mask, laryngeal mask airway and Combitube in a bench model
    Frass, M
    Staudinger, T
    Losert, H
    Krafft, P
    RESUSCITATION, 1999, 43 (01) : 80 - 81