Variability in oxygen delivery with bag-valve-mask devices: An observational laboratory simulation study

被引:5
|
作者
Dodds, Jarron M. [1 ,5 ]
Appelqvist, Dylan, I [2 ]
Paleologos, Michael S. [3 ,4 ]
Downey, Ryan G. [3 ,4 ]
机构
[1] St Vincents Hosp, Dept Anaesthesia & Acute Pain Med, Melbourne, Australia
[2] Univ Notre Dame, Sydney, Australia
[3] Royal Prince Alfred Hosp, Dept Anaesthesia, Sydney, Australia
[4] Univ Sydney, Sydney Med Sch, Sydney, Australia
[5] St Vincents Hosp Melbourne, Dept Anaesthesia & Acute Pain Med, 41 Victoria Parade, Fitzroy, Vic 3065, Australia
关键词
Bag-valve-mask; BVM; self-inflating resuscitation device; non-rebreather valve; pre-oxygenation; oxygen delivery; emergency intubation; PRE-OXYGENATION; SPONTANEOUS VENTILATION; PREOXYGENATION; ANESTHESIA; VOLUME; ADULT;
D O I
10.1177/0310057X221119824
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
A bag-valve-mask (BVM) is a portable handheld medical device commonly used in airway management and manual ventilation. Outside of the operating theatre, BVM devices are often used to pre-oxygenate spontaneously breathing patients before intubation to reduce the risk of hypoxaemia. Pre-oxygenation is considered adequate when the end-tidal expiratory fraction of oxygen is greater than 0.85. There are reports that some BVM devices fail to deliver a satisfactory inspired oxygen (FiO(2)) in spontaneously breathing patients due to variability in design. The primary aim of this study was to evaluate the efficacy of oxygen delivery of a broad range of adult and paediatric BVM devices at increasing tidal volumes using a mechanical lung to simulate spontaneous ventilation. The secondary aim was to evaluate the effect of BVM design on performance. Forty BVM devices were evaluated in a laboratory setting as part of a safety assessment requested by HealthShare New South Wales. The oxygen inlet of each BVM device was primed with 100% oxygen (15 l/min) for two min. The BVM device was then attached to the mechanical lung and commenced spontaneous breathing at a fixed respiratory rate of 12 breaths/min with an inspiratory: expiratory ratio of 1:2. For each device FiO(2) was measured after two min of spontaneous breathing. This process was repeated with small (250 ml), medium (500 ml) and large (750 ml) tidal volumes simulating adult breathing in adult BVM devices, and small (150 ml), medium (300 ml) and large (450 ml) tidal volumes simulating paediatric breathing in paediatric BVM devices. The test was repeated using up to five BVM devices of the same model (where supplied) at each tidal volume as a manufacturing quality control measure. Eight of the 40 devices tested failed to deliver a FiO(2) above 0.85 for at least one tidal volume, and five models failed to achieve this at any measured tidal volume. Concerningly, three of these devices delivered a FiO(2) below 0.55. Six of the eight poorly performing devices delivered reducing concentrations of inspired oxygen with increasing tidal volumes. Devices which performed the worst were those with a duckbill non-rebreather valve and without a dedicated expiratory valve. Several BVM devices available for clinical use in Australia did not deliver sufficient oxygen for reliable pre-oxygenation in a spontaneously breathing in vitro model. Devices with a duckbill non-rebreather valve and without a dedicated expiratory valve performed the worst. It is imperative that clinicians using BVM devices to deliver oxygen to spontaneously breathing patients are aware of the characteristics and limitations of the BVM devices, and that the standards for manufacture are updated to require safe performance in all clinical circumstances.
引用
收藏
页码:130 / 140
页数:11
相关论文
共 50 条
  • [1] Influence of face mask design on bag-valve-mask ventilation performance: a randomized simulation study
    Na, J. U.
    Han, S. K.
    Choi, P. C.
    Cho, J. H.
    Shin, D. H.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2013, 57 (09) : 1186 - 1192
  • [2] Understanding your bag-valve-mask resuscitator
    Lee, C. P.
    Yip, Y. Y.
    BJA EDUCATION, 2023, 23 (06) : 208 - 211
  • [3] Bag-Valve-Mask versus Laryngeal Mask Airway Ventilation in Cardiopulmonary Resuscitation with Continuous Compressions: A Simulation Study
    Dundar, Zerrin Defne
    Ayranci, Mustafa Kursat
    Kocak, Sedat
    Girisgin, Abdullah Sadik
    PREHOSPITAL AND DISASTER MEDICINE, 2021, 36 (02) : 189 - 194
  • [4] Bag-Valve-Mask (BVM): A performance Evaluation Report
    Kazlovich, Kate
    Lembrikov, Ilya
    Anwari, Vahid
    Mashari, Azad
    ANESTHESIA AND ANALGESIA, 2022, 134 : 786 - 786
  • [5] Does the Number of Fingers on the Bag Influence Volume Delivery? A Randomized Model Study of Bag-Valve-Mask Ventilation in Infants
    Zweiker, David
    Schwaberger, Hanna
    Urlesberger, Berndt
    Mileder, Lukas P.
    Baik-Schneditz, Nariae
    Pichler, Gerhard
    Schmoelzer, Georg M.
    Schwaberger, Bernhard
    CHILDREN-BASEL, 2018, 5 (10):
  • [6] Gastric rupture following bag-valve-mask ventilation
    Smally, AJ
    Ross, MJ
    Huot, CP
    JOURNAL OF EMERGENCY MEDICINE, 2002, 22 (01): : 27 - 29
  • [7] Training Novice Users in Bag-Valve-Mask Technique
    Matioc, Adrian A.
    PREHOSPITAL AND DISASTER MEDICINE, 2010, 25 (04) : 370 - 371
  • [8] Optimizing bag-valve-mask ventilation with a new mouth-to-bag resuscitator
    Wagner-Berger, HG
    Wenzel, V
    Stallinger, A
    Voelckel, WG
    Rheinberger, K
    Augenstein, S
    Herff, H
    Idris, AH
    Dörges, V
    Lindner, KH
    Hörmann, C
    RESUSCITATION, 2003, 56 (02) : 191 - 198
  • [9] Evaluation of the self-inflating bag-valve-mask and non-rebreather mask as preoxygenation devices in volunteers
    Robinson, Amelia
    Ercole, Ari
    BMJ OPEN, 2012, 2 (05):
  • [10] Optimizing bag-valve-mask ventilation with a new mouth-to-bag resuscitator
    Wagner-Berger, HG
    Wenzel, V
    Stallinger, A
    Voelckel, WG
    Rheinberger, K
    Augenstein, S
    Herff, H
    Idris, AK
    Doerges, V
    Lindner, KH
    Hoermann, C
    CIRCULATION, 2002, 106 (19) : 538 - 538