Development and validation of an ultra-high performance liquid chromatography-tandem mass spectrometry method to quantify antiretroviral drug concentrations in human plasma for therapeutic monitoring

被引:0
|
作者
West III, Raymond E. [1 ]
Oberly, Patrick J. [1 ]
Riddler, Sharon A. [2 ]
Nolin, Thomas D. [1 ]
Devanathan, Aaron S. [1 ,3 ]
机构
[1] Univ Pittsburgh, Ctr Clin Pharmaceut Sci, Sch Pharm, Dept Pharm & Therapeut, Pittsburgh, PA USA
[2] Univ Pittsburgh Sch Med, Dept Med, Pittsburgh, PA USA
[3] Univ Pittsburgh Sch Pharm, Sch Pharm, 208 Salk Pavil,335 Sutherland Dr, Pittsburgh, PA 15261 USA
关键词
Antiretrovirals; Concentrations; HIV; UPLC-MS/MS; Therapeutic drug monitoring; Clinical pharmacology; LC-MS/MS METHOD;
D O I
10.1016/j.jpba.2023.115932
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Antiretroviral therapy (ART) is highly effective for the treatment of HIV-1 infection. ART previously consisted of concomitant administration of many drugs, multiple times per day. Currently, ART generally consists of two- or three-drug regimens once daily as fixed-dose combinations. Drug monitoring may be necessary to ensure adequate concentrations are achieved in the plasma over the dosing interval and prevent further HIV resistance formation. Additionally, nonadherence remains an issue, highlighting the need to ensure sufficient ART exposure. Towards this effort, we developed and validated a highly selective ultra performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for the simultaneous quantification of a panel of nine antiretrovirals: abacavir, bictegravir, cabotegravir, dolutegravir, doravirine, emtricitabine, lamivudine, raltegravir, and tenofovir in human plasma. Using only 50 mu L of plasma, a simple protein precipitation with acetonitrile with internal standards followed by reconstitution in 50 uL (high) or 400 uL (low) was performed. Analyte separation was achieved using a multistep UPLC gradient mixture of (A: 0.1% formic acid in water and B: acetonitrile) and a Waters CORTECS T3 (2.1 x100 mm) column. The method was comprehensively validated according to the United States Food and Drug Administration Bioanalytical Guidelines over two clinically relevant ranges (1-250 ng/mL and 100-5000 ng/mL) with excellent linearity (R-2 > 0.99 for all). The assay run time was 7.5 min. This method achieves acceptable performance of trueness (89.7-104.1%), repeatability, and precision (CV <15%), and allows for simultaneous quantification of guideline-recommended ART regimens. This method can be utilized for the therapeutic monitoring of antiretrovirals in human plasma.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Ultra-Performance Liquid Chromatography-Tandem Mass Spectrometry for Simultaneous Determination of Antipsychotic Drugs in Human Plasma and Its Application in Therapeutic Drug Monitoring
    Qi, Yingjie
    Liu, Guangxuan
    DRUG DESIGN DEVELOPMENT AND THERAPY, 2021, 15 : 463 - 479
  • [42] High performance liquid chromatography-tandem mass spectrometry for the determination of bile acid concentrations in human plasma
    Xiang, Xiaoqiang
    Han, Yi
    Neuvonen, Mikko
    Laitila, Jouko
    Neuvonen, Pertti J.
    Niemi, Mikko
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (01): : 51 - 60
  • [43] Development and validation of a high-performance liquid chromatography-tandem mass spectrometry method to quantify LY-354,740 in rat and marmoset plasma
    Gaudette, Fleur
    Hamadjida, Adjia
    Bedard, Dominique
    Nuara, Stephen G.
    Beaudry, Francis
    Huot, Philippe
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2017, 1061 : 392 - 398
  • [44] High-Throughput Ultra Performance Liquid Chromatography-Tandem Mass Spectrometry Method Validation for the Estimation of Atorvastatin and Active Metabolites in Human Plasma
    Agrawal, Nikhil
    Mittal, Amit
    ASSAY AND DRUG DEVELOPMENT TECHNOLOGIES, 2023, 21 (03) : 110 - 125
  • [45] Validation of a method for quantification of ketobemidone in human plasma with liquid chromatography-tandem mass spectrometry
    Lampinen, M
    Bondesson, U
    Fredriksson, E
    Hedeland, M
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 789 (02): : 347 - 354
  • [46] Development and validation of a method for quantitative determination of valsartan in human plasma by liquid chromatography-tandem mass spectrometry
    Koseki, Nozomu
    Kawashita, Hiroto
    Hara, Hisanori
    Niina, Miyuki
    Tanaka, Makoto
    Kawai, Ryosei
    Nagae, Yusuke
    Masuda, Naoki
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (05) : 1769 - 1774
  • [47] Development and validation of a high-performance liquid chromatography-tandem mass spectrometry assay for the quantification of Dexamphetamine in human plasma
    Herbrink, M.
    Thijssen, B.
    Hillebrand, M. J. X.
    Rosing, H.
    Schellens, J. H. M.
    Nuijen, B.
    Beijnen, J. H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2018, 148 : 259 - 264
  • [48] Determination of fluoroquinolones in dried plasma spots by using microwave-assisted extraction coupled to ultra-high performance liquid chromatography-tandem mass spectrometry for therapeutic drug monitoring
    Brahmadhi, Ageng
    Chen, Michael X.
    Wang, San-Yuan
    Cho, Yun-Yu
    Yu, Ming-Chih
    Lee, Chih-Hsin
    Tsai, I-Lin
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2021, 195
  • [49] Development and validation of a high-performance liquid chromatography-tandem mass spectrometry assay quantifying olaparib in human plasma
    Nijenhuis, C. M.
    Lucas, L.
    Rosing, H.
    Schellens, J. H. M.
    Beijnen, J. H.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2013, 940 : 121 - 125
  • [50] Development and validation of a high-performance liquid chromatography-tandem mass spectrometry assay quantifying vemurafenib in human plasma
    Nijenhuis, C. M.
    Rosing, H.
    Schellens, J. H. M.
    Beijnen, J. H.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 88 : 630 - 635