Agreement of Corneal Diameter and Anterior Chamber Depth Measurements with the IOLMaster 500 and the IOLMaster 700 Optical Biometers in Myopic Eyes

被引:1
|
作者
Tana-Rivero, Pedro [1 ,2 ]
Tello-Elordi, Cristina [1 ]
Orts-Vila, Paz [1 ]
Tana-Sanz, Pedro [1 ]
Tana-Sanz, Santiago [1 ]
机构
[1] Oftalvist Alicante, Cataract Dept, Alicante, Spain
[2] Oftalvist Alicante, Cataract Dept, C Angel Lozano N 11, Alicante 03001, Spain
来源
CLINICAL OPHTHALMOLOGY | 2023年 / 17卷
关键词
anterior chamber depth; white-to-white; IOLMaster; 500; 700; optical biometry; INTRAOCULAR-LENS POWER; PRECISION REPEATABILITY; REPRODUCIBILITY;
D O I
10.2147/OPTH.S409132
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the agreement in automated corneal diameter (CD) and anterior chamber depth (ACD, measured from corneal epithelium to lens) distances measurements between the IOLMaster 500 and 700 optical biometers in myopic eyes.Methods: One hundred and sixteen eyes of 116 myopic patients aged between 21 to 60 years were included in this study. Measurements of CD and ACD distances were taken with each biometer for all patients in the same session. A t-test and a p-value less than 0.05 was considered statistically significant when the measurements were compared. The agreement between biometers was studied by applying a Bland-Altman analysis.Results: The mean CD values obtained using the IOLMaster 500 and 700 biometers were 12.26 +/- 0.35 mm and 12.13 +/- 0.34 mm, respectively. The mean ACD distance values were 3.61 +/- 0.29 mm and 3.62 +/- 0.31, for the IOLMaster 500 and 700 biometers, respectively. There were statistically significant differences between the two devices only for CD measurements (p<0.001), but not for ACD measurements (p=0.26). The limits of agreement obtained were wide in both types of measurements, being 0.422 mm for the CD distance and 0.389 mm for the ACD distance.Conclusion: There were statistically significant differences between the IOLMaster 500 and 700 biometers regarding CD but not in ACD measurements in healthy myopic eyes. These differences could be clinically significant in some cases. According to these results, medical judgement should be used to assess whether the two devices could be used interchangeably for CD and ACD measurements during the clinical practice.
引用
收藏
页码:1245 / 1253
页数:9
相关论文
共 32 条
  • [11] Comparability and intra-/interobserver reliability of anterior chamber depth measurements with the Pentacam and IOLMaster
    Savant, Vijay
    Chavan, Randhir
    Pushpoth, Sreekumari
    Ilango, B.
    JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (06) : 615 - 618
  • [12] Comparison of Anterior Chamber Depth Measurements from the Galilei Dual Scheimpflug Analyzer with IOLMaster
    Patel, Roma P.
    Pandit, Rahul T.
    JOURNAL OF OPHTHALMOLOGY, 2012, 2012
  • [13] Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000
    Liao, Xuan
    Peng, Yue
    Liu, Bo
    Tan, Qing-Qing
    Lan, Chang-Jun
    SCIENTIFIC REPORTS, 2020, 10 (01) : 3134
  • [14] Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000
    Xuan Liao
    Yue Peng
    Bo Liu
    Qing-Qing Tan
    Chang-Jun Lan
    Scientific Reports, 10
  • [15] Comparison of Horizontal Corneal Diameter Measurements Using Orbscan IIz, OPD Scan III, and IOLMaster 700
    Cruz, Sebastian
    Valenzuela, Felipe
    Stoppel, Juan
    Maul, Eugenio
    Gibbons, Allister
    EYE & CONTACT LENS-SCIENCE AND CLINICAL PRACTICE, 2021, 47 (10): : 533 - 538
  • [16] Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes
    Dong, Jing
    Tang, Maolong
    Zhang, Yaqin
    Jia, Yading
    Zhang, Haining
    Jia, Zhijie
    Wang, Xiaogang
    PLOS ONE, 2015, 10 (11):
  • [17] Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children
    Huri Sabur
    Omer Takes
    International Ophthalmology, 2023, 43 : 475 - 482
  • [18] Agreement of axial length and anterior segment parameters measured with the MYAH device compared to Pentacam AXL and IOLMaster 700 in myopic children
    Sabur, Huri
    Takes, Omer
    INTERNATIONAL OPHTHALMOLOGY, 2023, 43 (02) : 475 - 482
  • [19] Anterior Chamber Depth Measurements Following Cycloplegia Using Photographic Image Analysis (IOLMaster™) and Anterior Segment Optical Coherence Tomography (Visante™) and Their Relationship to the Haigis Formula Derived Postop Anterior Chamber Depth
    Lyles, G. W.
    Cohen, K. L.
    Aschbrenner, M. W.
    El-Khorazaty, J.
    Lam, D.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [20] To compare central corneal thickness measurements obtained by Pentacam with those obtained by IOLMaster 700, Cirrus anterior segment optical coherence tomography and Tomey specular microscopy in normal healthy eyes
    Kumar, Kiran K.
    Prakash, Akshata A.
    Neeraja, T. G.
    Adappa, Karishma T.
    Prabha, Chandra T. S.
    Gangasagara, Suresh Babu
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2021, 69 (07) : 1713 - 1717