Agreement of Corneal Diameter and Anterior Chamber Depth Measurements with the IOLMaster 500 and the IOLMaster 700 Optical Biometers in Myopic Eyes

被引:1
|
作者
Tana-Rivero, Pedro [1 ,2 ]
Tello-Elordi, Cristina [1 ]
Orts-Vila, Paz [1 ]
Tana-Sanz, Pedro [1 ]
Tana-Sanz, Santiago [1 ]
机构
[1] Oftalvist Alicante, Cataract Dept, Alicante, Spain
[2] Oftalvist Alicante, Cataract Dept, C Angel Lozano N 11, Alicante 03001, Spain
来源
CLINICAL OPHTHALMOLOGY | 2023年 / 17卷
关键词
anterior chamber depth; white-to-white; IOLMaster; 500; 700; optical biometry; INTRAOCULAR-LENS POWER; PRECISION REPEATABILITY; REPRODUCIBILITY;
D O I
10.2147/OPTH.S409132
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the agreement in automated corneal diameter (CD) and anterior chamber depth (ACD, measured from corneal epithelium to lens) distances measurements between the IOLMaster 500 and 700 optical biometers in myopic eyes.Methods: One hundred and sixteen eyes of 116 myopic patients aged between 21 to 60 years were included in this study. Measurements of CD and ACD distances were taken with each biometer for all patients in the same session. A t-test and a p-value less than 0.05 was considered statistically significant when the measurements were compared. The agreement between biometers was studied by applying a Bland-Altman analysis.Results: The mean CD values obtained using the IOLMaster 500 and 700 biometers were 12.26 +/- 0.35 mm and 12.13 +/- 0.34 mm, respectively. The mean ACD distance values were 3.61 +/- 0.29 mm and 3.62 +/- 0.31, for the IOLMaster 500 and 700 biometers, respectively. There were statistically significant differences between the two devices only for CD measurements (p<0.001), but not for ACD measurements (p=0.26). The limits of agreement obtained were wide in both types of measurements, being 0.422 mm for the CD distance and 0.389 mm for the ACD distance.Conclusion: There were statistically significant differences between the IOLMaster 500 and 700 biometers regarding CD but not in ACD measurements in healthy myopic eyes. These differences could be clinically significant in some cases. According to these results, medical judgement should be used to assess whether the two devices could be used interchangeably for CD and ACD measurements during the clinical practice.
引用
收藏
页码:1245 / 1253
页数:9
相关论文
共 32 条
  • [1] Agreement of phakic and pseudophakic anterior chamber depth measurements in IOLMaster and Pentacam
    Hamoudi, Hassan
    Christensen, Ulrik Correll
    la Cour, Morten
    ACTA OPHTHALMOLOGICA, 2018, 96 (03) : E403 - E403
  • [2] Comparison of anterior chamber depth measurements taken with the Pentacam, Orbscan IIz and IOLMaster in myopic and emmetropic eyes
    Utine, Canan Asli
    Altin, Feyza
    Cakir, Hanefi
    Perente, Irfan
    ACTA OPHTHALMOLOGICA, 2009, 87 (04) : 386 - 391
  • [3] Corneal radii and anterior chamber depth measurements using the IOLMaster versus the Pentacam
    Reuland, Mirjarn S.
    Reuland, Andreas J.
    Nishi, Yutaro
    Auffarth, Gerd U.
    JOURNAL OF REFRACTIVE SURGERY, 2007, 23 (04) : 368 - 373
  • [4] Measurement of the internal diameter and depth of the anterior chamber: IOLMaster versus anterior chamber optical coherence tomographer
    Baikoff, G
    Jodai, HJ
    Bourgeon, G
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09): : 1722 - 1728
  • [5] Comparison of central corneal thickness measured in myopic eyes by Pentacam, Sirius and IOLMaster 700
    Song, Han
    Liu, Chengyang
    Yang, Wei
    Yang, Chunliu
    Cheng, Xiaodong
    PHOTODIAGNOSIS AND PHOTODYNAMIC THERAPY, 2024, 49
  • [6] Agreement of Anterior Segment Parameter Measurements With CASIA 2 and IOLMaster 700
    Ruan, Xiaoting
    Yang, GuangYao
    Xia, Zhaoxia
    Zhang, Jiaqing
    Gu, Xiaoxun
    Tan, Yuan
    Liu, Zhenzhen
    Luo, Lixia
    FRONTIERS IN MEDICINE, 2022, 9
  • [7] Comparison of Anterior Chamber Depth Measurements Conducted With Pentacam HR® and IOLMaster®
    Nemeth, Gabor
    Hassan, Ziad
    Modis, Laszlo, Jr.
    Szalai, Eszter
    Katona, Kristof
    Berta, Andras
    OPHTHALMIC SURGERY LASERS & IMAGING, 2011, 42 (02) : 144 - 147
  • [8] Comparison of anterior chamber depth measurements using Orbscan II and IOLMaster
    Frisch, I. B.
    Rabsilber, T. M.
    Becker, K. A.
    Reuland, A. J.
    Auffarth, G. U.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2007, 17 (03) : 327 - 331
  • [9] Comparison of anterior chamber depth measurements using the IOLMaster, scanning peripheral anterior chamber depth analyser, and anterior segment optical coherence tomography
    Lavanya, Raghavan
    Teo, Livia
    Friedman, David S.
    Aung, Han T.
    Baskaran, Mani
    Gao, Hong
    Alfred, Tamuno
    Seah, Steve K.
    Kashiwagi, Kenji
    Foster, Paul J.
    Aung, Tin
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2007, 91 (08) : 1023 - 1026
  • [10] The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™
    Lam, AKC
    Chan, R
    Pang, PCK
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2001, 21 (06) : 477 - 483