Characteristics of detection accuracy of the patient setup using InBore optical patient positioning system

被引:1
|
作者
Oku, Yoshifumi [1 ]
Toyota, Masahiko [1 ]
Saigo, Yasumasa [1 ]
机构
[1] Kagoshima Univ Hosp, Div Radiol, Dept Clin Technol, 8-35-1 Sakuragaoka, Kagoshima, Kagoshima 8908520, Japan
关键词
Surface-guided radiation therapy; InBore system; Patient setup; Detection accuracy; SURFACE IMAGING-SYSTEM; LOCALIZATION;
D O I
10.1007/s12194-023-00741-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study aimed to evaluate the detection accuracy of the AlignRT-InBore system in surface-guided radiation therapy using a phantom and to determine the feasibility of the system by conducting a comparative analysis with cone-beam computed tomography (CBCT) registration. The AlignRT-InBore system integrated with the ETHOS Therapy was used. A phantom and a QUASAR phantom were employed to examine the specific areas of interest relevant to clinical cases. The evaluation involved monitoring translations for approximately 30 min and assessing the position detection accuracy for static and moving objects. Fifty clinical cases were used to evaluate the position detection accuracy and its relationship with the localization accuracy of CBCT before treatment. The detection accuracy of static and moving objects was within 1.0 mm using the phantom. However, the longitudinal direction tended to be larger than the other directions. Regarding the accuracy of localization in clinical cases, a strong and statistically significant (p < 0.01) correlation was observed in each direction. A detection accuracy within 1.0 mm is possible for static and moving objects. The detection accuracy of the patient setup using the InBore optical patient positioning system was extremely high, and the patient could be detected with high precision, suggesting its usefulness.
引用
收藏
页码:532 / 542
页数:11
相关论文
共 50 条
  • [41] Three-dimensional accuracy and interfractional reproducibility of patient fixation and positioning using a stereotactic head mask system
    Karger, CP
    Jäkel, O
    Debus, J
    Kuhn, S
    Hartmann, GH
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (05): : 1493 - 1504
  • [42] Eliminating Errors in Setup and Positioning for Patient-Specific IMRT QA
    Kamath, S.
    Palta, J.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [43] Accuracy of patient setup for extracranial stereotactic radiotherapy for lung cancer
    Mourillon, T
    Kuijper, IT
    Meijburg-van den Hoek, MM
    Cuijpers, JP
    Slotman, BJ
    Lagerwaard, FJ
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S158 - S159
  • [44] Dosimetric Evaluation of HexaPOD as a Tool for Patient Setup Verification Accuracy
    Ahmad, S.
    Spencer, K.
    Brahmavar, S.
    Hyder, J.
    Saleeby, J.
    Bornstein, L.
    MEDICAL PHYSICS, 2012, 39 (06) : 3618 - 3618
  • [45] Individual and Group Setup Margins with KV Imaging for Daily Patient Positioning
    Niedermayr, T.
    Herman, M. G.
    MEDICAL PHYSICS, 2008, 35 (06) : 2922 - +
  • [46] Assessment of Patient Setup Accuracy at a Treatment Position after the Patient Has Been Registered at Another Location Using Robotic Positioner and a Dual KV Imaging System
    Ghebremedhin, A.
    Wang, N.
    Patyal, B.
    MEDICAL PHYSICS, 2011, 38 (06) : 3603 - U73
  • [47] Automated patient positioning in CT using a 3D camera for body contour detection: accuracy in pediatric patients
    Ronald Booij
    Marcel van Straten
    Andreas Wimmer
    Ricardo P.J. Budde
    European Radiology, 2021, 31 : 131 - 138
  • [48] Comparison of automatic tomotherapy CT setup verification vs. an optically guided patient positioning system
    Boswell, S
    Tome, W
    Jaradat, H
    Mackie, TR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S532 - S532
  • [49] Automated patient positioning in CT using a 3D camera for body contour detection: accuracy in pediatric patients
    Booij, Ronald
    van Straten, Marcel
    Wimmer, Andreas
    Budde, Ricardo P. J.
    EUROPEAN RADIOLOGY, 2021, 31 (01) : 131 - 138
  • [50] Accuracy of a patient positioning system for image-guided frameless stereotactic IMRT treatments
    Dohm, O.
    Hoenes, A.
    Paulsen, F.
    Alber, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S288 - S288