Comparison of corneal measurements using two different Scheimpflug analyzers in Sirius and Pentacam devices

被引:5
|
作者
Abdi, Parisa [1 ]
Atighehchian, Mehrnaz [1 ]
Farsiani, Amir Reza [2 ]
机构
[1] Univ Tehran Med Sci, Farabi Eye Hosp, Tehran, Iran
[2] Zanjan Univ Med Sci, Sch Med, Dept Ophthalmol, Zanjan, Iran
关键词
THICKNESS MEASUREMENT; REPEATABILITY; ANTERIOR; POWER; AGREEMENT; GALILEI; PARAMETERS; ULTRASOUND; DIAMETER; BIOMETRY;
D O I
10.1038/s41598-023-44133-3
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
The aim was to compare measurements of anterior segment biometry parameters using two Scheimpflug tomographies, Pentacam and Sirius to assess the agreement. Prospective cross-sectional observational study. A total of 60 eyes of 30 healthy subjects were included and evaluated with Pentacam followed by Sirius imaging. Corneal indices were performed with two modalities in both eyes including; apical corneal thickness (ACT), corneal thickness at pupil site(PCT), thinnest corneal thickness (TCT), anterior chamber depth (ACD), chamber angle, chamber volume, cornea volume, mean front keratometry (FKm), the radius of corneal curvature at the anterior and posterior surface in steep and flat meridian, anterior astigmatism values, pupil diameter, and horizontal corneal diameter. The Bland-Altman graph and ICC (intra-class correlation were used to establish an agreement and present the similarity of the findings. Most parameters showed perfect agreement. In both devices, the ICC was more than 0.91 in all measurements except for ACD (ICC = 0.820), cylinder axis (ICC = 0.520), TCT(ICC = 0.881), ACT(ICC = 0.672), PCT (ICC = 0.882), chamber angle (ICC = 0.362), pupil diameter(ICC = 0.137). Pentacam yielded higher values that were significant in five parameters including 3.47 mu m for TCT, 4.29 mu m for PCT, 10.03 mm(3) for chamber volume,2.67 mm(3) for cornea volume, and 1.49 mm for pupil diameter but there was only a statistically significant difference in cornea volume and pupil diameter (p-value < 0.001). However, in Pentacam only the chamber angle value was 6.44 mm3 lower than Sirius, with a statistically significant difference (p-value < 0.001). Although these two devices had some statistically different results, it seems that they have a good agreement and correlation in most parameters.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Comparison of Angle Kappa Measurements: Repeatability of Sirius, Keratron, and Pentacam and Agreement With Amaris 750S
    Wang, Yuzhou
    Hu, Xiaojian
    Gong, Qianwen
    Yu, Ye
    Lu, Fan
    Hu, Liang
    JOURNAL OF REFRACTIVE SURGERY, 2025, 41 (03) : e213 - e219
  • [42] Comparison of Central Corneal Thickness Measurements with Pentacam, Orbscan II, and Ultrasound Pachymeter
    Yekta, Abbas-Ali
    Hashemi, Hassan
    KhabazKhoob, Mehdi
    Dostdar, Asghar
    Mehravaran, Shiva
    Heravian, Javad
    Fotouhi, Akbar
    IRANIAN JOURNAL OF OPHTHALMOLOGY, 2009, 21 (02): : 51 - 57
  • [43] Comparison of Central Corneal Thickness Measurements in Normal and Keratoconic Eyes Using Ultrasonic Pachymetry and OCULUS Pentacam
    Buyuk, Kursat
    Bozkurt, Banu
    Kamis, Umit
    Ozkagnici, Ahmet
    Okudan, Suleyman
    TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2011, 41 (02): : 104 - 107
  • [44] Comparison between central corneal thickness measurements by oculus pentacam and ultrasonic pachymetry
    Al-Mezaine H.S.
    Al-Amro S.A.
    Kangave D.
    Sadaawy A.
    Wehaib T.A.
    Al-Obeidan S.
    International Ophthalmology, 2008, 28 (5) : 333 - 338
  • [45] Comparison between corneal elevation maps using different reference surfaces with Scheimpflug–Placido topographer
    Engy Mohamed Mostafa
    International Ophthalmology, 2017, 37 : 553 - 558
  • [46] Corneal Power Measurements Using Scheimpflug Imaging in Eyes With Prior Corneal Refractive Surgery
    Holladay, Jack T.
    Hill, Warren E.
    Steinmueller, Andreas
    JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (10) : 862 - 868
  • [47] Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes
    Irmak Karaca
    Suzan Guven Yilmaz
    Melis Palamar
    Halil Ates
    International Ophthalmology, 2018, 38 : 1601 - 1609
  • [48] Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
    Mohammad Naderan
    Saeed Shoar
    Morteza Naderan
    Mohammad Amin Kamaleddin
    Mohammad Taher Rajabi
    International Journal of Ophthalmology, 2015, (02) : 275 - 280
  • [49] Comparison of central corneal thickness and endothelial cell measurements by Scheimpflug camera system and two noncontact specular microscopes
    Karaca, Irmak
    Yilmaz, Suzan Guven
    Palamar, Melis
    Ates, Halil
    INTERNATIONAL OPHTHALMOLOGY, 2018, 38 (04) : 1601 - 1609
  • [50] Comparison of corneal measurements in keratoconic eyes using rotating Scheimpflug camera and scanning-slit topography
    Naderan, Mohammad
    Shoar, Saeed
    Naderan, Morteza
    Kamaleddin, Mohammad Anmin
    Rajabi, Mohammad Taher
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2015, 8 (02) : 275 - 280