Development and validation of a fully automated laboratory-developed test for detection of HSV-1/2 and VZV in clinical samples run on the Panther Fusion® system

被引:0
|
作者
Vahidnia, Ali [1 ]
Voll, Soeraya [1 ]
Zwart, Nick [1 ]
Tangali, Faried [2 ]
Oosterlaken, Tom [2 ]
Sinnige, Jan C. [1 ]
Kalpoe, Jayant [1 ]
Herpers, Bjorn L. [1 ]
机构
[1] Reg Publ Hlth Lab Kennemerland, Haarlem, Netherlands
[2] Hologic Inc, Sci Affairs, Amsterdam, Netherlands
关键词
Laboratory-developed test; Herpes simplex virus (HSV) 1 & 2; Varicella-zoster virus (VZV); Multiplex PCR;
D O I
10.1016/j.diagmicrobio.2023.116022
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The purpose of this study was to develop a laboratory developed test (LDT) for HSV1/2 and VZV to run on fully automated Hologic Panther Fusion (R) System. The Panther Fusion System is a fully automated walkaway system, providing end-to-end workflow from sample input to DNA/RNA extraction, amplification, automated analysis, and reporting to a laboratory information system (LIS). The LDT was developed and validated on 230 clinical and 20 reference samples (n = 250) and compared to a commercially available kit. Assessment of the analytical and diagnostic performances of the LDT revealed >98% accuracy, sensitivity, and specificity, which is consistent with or better than many of the commercial or laboratory-developed tests available. The advantage of this LDT is that it is designed to perform a single-run full female health screening in parallel with 4 commercially available Hologic kits for Chlamydia trachomatis/Neisseria gonorrhea (CT/NG), Trichomonas vaginalis (TV), Mycoplasma genitalium (MG), and bacterial vaginosis (BV).(C) 2023 Elsevier Inc. All rights reserved.
引用
收藏
页数:4
相关论文
共 13 条
  • [1] Validation of a Laboratory-Developed Test for HSV Types 1 and 2 in Cerebrospinal Fluid
    Bissaillon, C. M.
    Fienman, J. A.
    Lebel, K. A.
    Moore, F. R.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2013, 15 (06): : 892 - 892
  • [2] Detection of Group A Streptococcus Using a Laboratory Developed Test on the Fully Automated Panther Fusion Open Access System
    Vang, L.
    Wilkins, B.
    Panuganti, S.
    Srinivasan, V.
    Worlock, A.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2018, 20 (06): : 947 - 947
  • [3] Analysis of sputum/tracheal aspirate and nasopharyngeal samples for SARS-CoV-2 detection by laboratory-developed test and Panther Fusion system
    Thwe, Phyu M.
    Ren, Ping
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2021, 99 (01)
  • [4] Detection of Mycoplasma pneumoniae and Chlamydia pneumoniae using a laboratory-developed high-throughput test for the Panther Fusion system
    Nilsson, Stephen Strunge
    Fahrmann, Linda
    Sundelin, Thomas
    Nielsen, Lene
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2025, 112 (02)
  • [5] Simultaneous detection of HSV-1, HSV-2 and VZV in clinical samples by multiplex polymerase chain reaction (MPCR)
    Sharifzad, F.
    Behbahani, A. Behzad
    Sadeghi, M. R.
    JOURNAL OF CLINICAL VIROLOGY, 2006, 36 : S25 - S26
  • [6] Evaluation of a laboratory-developed test for simultaneous detection of norovirus and rotavirus by real-time RT-PCR on the Panther Fusion® system
    Kulis-Horn, Robert K.
    Tiemann, Carsten
    EUROPEAN JOURNAL OF CLINICAL MICROBIOLOGY & INFECTIOUS DISEASES, 2020, 39 (01) : 103 - 112
  • [7] Evaluation of a laboratory-developed test for simultaneous detection of norovirus and rotavirus by real-time RT-PCR on the Panther Fusion® system
    Robert K. Kulis-Horn
    Carsten Tiemann
    European Journal of Clinical Microbiology & Infectious Diseases, 2020, 39 : 103 - 112
  • [8] Design and Development of a Fully Automated Assay for Detection and Differentiation of HSV-1 and HSV-2: The IMDx HSV-1/2 for Abbott m2000
    Bradley, S.
    Reiske, H.
    Salcius, J.
    Lauzier, B.
    Laslin, C.
    Jacobs, A.
    Del Mastro, R.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2012, 14 (06): : 694 - 694
  • [9] Performance Comparison of the Lyra Multiplex HSV1/2+VZV PCR with a Laboratory Developed Single-Plex HSV-1 and HSV-2 PCR from Retrospectively Tested Samples
    Mancini, J.
    Svetcharnik, C.
    Rocklin, K.
    Payne, D.
    JOURNAL OF MOLECULAR DIAGNOSTICS, 2014, 16 (06): : 733 - 733
  • [10] Analytical and clinical validation of a novel, laboratory-developed, modular multiplex-PCR panel for fully automated high-throughput detection of 16 respiratory viruses
    Tang, Hui Ting
    Noerz, Dominik
    Grunwald, Moritz
    Giersch, Katja
    Pfefferle, Susanne
    Fischer, Nicole
    Aepfelbacher, Martin
    Rohde, Holger
    Lutgehetmann, Marc
    JOURNAL OF CLINICAL VIROLOGY, 2024, 173