Analytical and clinical validation of a novel, laboratory-developed, modular multiplex-PCR panel for fully automated high-throughput detection of 16 respiratory viruses

被引:2
|
作者
Tang, Hui Ting [1 ]
Noerz, Dominik [1 ]
Grunwald, Moritz [1 ]
Giersch, Katja [1 ]
Pfefferle, Susanne [1 ]
Fischer, Nicole [1 ]
Aepfelbacher, Martin [1 ]
Rohde, Holger [1 ]
Lutgehetmann, Marc [1 ]
机构
[1] Univ Med Ctr Hamburg Eppendorf, Inst Med Microbiol Virol & Hyg, Martinistr 52, D-20246 Hamburg, Germany
关键词
RT-PCR; Respiratory viruses; Cobas x800; COVID-19; IDENTIFICATION; INFLUENZA; SEVERITY; ASSAYS; LOAD; RNA;
D O I
10.1016/j.jcv.2024.105693
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Background: : Viral respiratory Infections pose a health risk, especially to vulnerable patient populations. Effective testing programs can detect and differentiate these infections at an early stage, which is particularly important for high-risk clinical departments. The objective of this study was to develop and validate a multiplex PCR-panel for 16 different respiratory viruses on a fully-automated high-throughput platform. Methods: Three multiplex-PCR assays were designed to run on the cobas5800/6800/8800 systems, consolidating 16 viral targets: RESP1: SARS-CoV-2, influenza-A/B, RSV; RESP2: hMPV, hBoV, hAdV, rhino-/ENV; RESP3: HPIV-1-4, hCoV-229E, hCoV-NL63, hCoV-OC43, hCoV-HKU1. Analytic performance was evaluated using digitalPCR based standards and international reference material. Clinical performance was determined by comparing results from clinical samples with reference assays. Results: Analytical sensitivity (i.e. lower limit of detection (LoD), 95 % probability of detection) was determined as follows: SARS-CoV-2: 29.3 IU/ml, influenza-A: 179.9 cp/ml, influenza-B: 333.9 cp/ml and RSV: 283.1 cp/ml. LoDs of other pathogens ranged between 9.4 cp/ml (hCoV-NL63) and 21,419 cp/ml (HPIV-2). Linearity was verified over 4-7 log-steps with pooled standard differentials (SD) ranging between 0.18-0.70ct. Inter-/intra-run variability (precision) was assessed for all targets over 3 days. SDs ranged between 0.13-0.74ct. Positive agreement in clinical samples was 99.4 % and 95 % for SARS-CoV-2 and influenza-A respectively. Other targets were in the 80-100 % range. Negative agreement varied between 96.3-100 %. Discussion: Lab-developed tests are a key factor for effective clinical diagnostics. The multiplex panel presented in this study demonstrated high performance and provides an easily scalable high-throughput solution for respiratory virus testing, e.g. for testing in high-risk patient populations.
引用
收藏
页数:7
相关论文
共 11 条
  • [1] Analytical and clinical validation of a multiplex PCR assay for detection of Neisseria gonorrhoeae and Chlamydia trachomatis including simultaneous LGV serotyping on an automated high-throughput PCR system
    Pflueger, Lisa Sophie
    Noerz, Dominik
    Grunwald, Moritz
    Pfefferle, Susanne
    Giersch, Katja
    Christner, Martin
    Weber, Beatrice
    Aepfelbacher, Martin
    Rohde, Holger
    Luetgehetmann, Marc
    MICROBIOLOGY SPECTRUM, 2024, 12 (03):
  • [2] Clinical Evaluation of the New High-Throughput Luminex NxTAG Respiratory Pathogen Panel Assay for Multiplex Respiratory Pathogen Detection
    Chen, Jonathan H. K.
    Lam, Ho-Yin
    Yip, Cyril C. Y.
    Wong, Sally C. Y.
    Chan, Jasper F. W.
    Ma, Edmond S. K.
    Cheng, Vincent C. C.
    Tang, Bone S. F.
    Yuen, Kwok-Yung
    JOURNAL OF CLINICAL MICROBIOLOGY, 2016, 54 (07) : 1820 - 1825
  • [3] Multiplex PCR for detection of respiratory viruses: Can the laboratory performing a respiratory viral panel (RVP) assay trigger better patient care and clinical outcomes?
    Chapin, Kimberle
    CLINICAL BIOCHEMISTRY, 2011, 44 (07) : 496 - 497
  • [4] Novel automated sample-to-result SARS-CoV-2 laboratory-developed RT-PCR assay for high-throughput testing using LabTurbo AIO 48 system
    Jian, Ming-, Jr.
    Chung, Hsing-Yi
    Chang, Chih-Kai
    Lin, Jung-Chung
    Yeh, Kuo-Ming
    Chiu, Sheng-Kang
    Wang, Yi-Hui
    Liao, Shu-Jung
    Li, Shih-Yi
    Hsieh, Shan-Shan
    Perng, Cherng-Lih
    Chang, Feng-Yee
    Shang, Hung-Sheng
    CLINICA CHIMICA ACTA, 2021, 514 : 54 - 58
  • [5] Development and validation of a fully automated laboratory-developed test for detection of HSV-1/2 and VZV in clinical samples run on the Panther Fusion® system
    Vahidnia, Ali
    Voll, Soeraya
    Zwart, Nick
    Tangali, Faried
    Oosterlaken, Tom
    Sinnige, Jan C.
    Kalpoe, Jayant
    Herpers, Bjorn L.
    DIAGNOSTIC MICROBIOLOGY AND INFECTIOUS DISEASE, 2023, 107 (02)
  • [6] High-throughput, sensitive, and accurate multiplex PCR-microsphere flow Cytometry system for large-scale comprehensive detection of respiratory viruses
    Lee, Wai-Ming
    Grindle, Kris
    Pappas, Tressa
    Marshall, David J.
    Moser, Michael J.
    Beaty, Edward L.
    Shult, Peter A.
    Prudent, James R.
    Gern, James E.
    JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (08) : 2626 - 2634
  • [7] Clinical evaluation of a fully automated, laboratory- developed multiplex RT-PCR assay integrating dual- target SARS-CoV-2 and influenza A/B detection on a high- throughput platform
    Noerz, Dominik
    Hoffmann, Armin
    Aepfelbacher, Martin
    Pfefferle, Susanne
    Luetgehetmann, Marc
    JOURNAL OF MEDICAL MICROBIOLOGY, 2021, 70 (02)
  • [8] Detection of C. difficile toxin as a model assay for performing fully automated high-throughput RT-PCR on clinical stool samples
    Eigner, Ulrich
    Noerz, Dominik
    Reucher, Svenja
    Furrer, Jan
    Sun, Jingtao
    Chu, Kristina
    Kolb, Melissa
    Hefner, Nadine
    Pfefferle, Susanne
    Luetgehetmann, Marc
    JOURNAL OF MICROBIOLOGICAL METHODS, 2020, 172
  • [9] Development and Validation of a Laboratory-Developed Multiplex Real-Time PCR Assay on the BD Max System for Detection of Herpes Simplex Virus and Varicella-Zoster Virus DNA in Various Clinical Specimens
    Pillet, Sylvie
    Verhoeven, Paul O.
    Epercieux, Amelie
    Bourlet, Thomas
    Pozzetto, Bruno
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (06) : 1921 - 1926
  • [10] Validation of a novel molecular diagnostic panel for pediatric musculoskeletal infections: Integration of the Cepheid Xpert MRSA/SA SSTI and laboratory-developed real-time PCR assays for clindamycin resistance genes and Kingella kingae detection
    Searns, Justin B.
    Robinson, Christine C.
    Wei, Qi
    Yuan, Ji
    Hamilton, Stacey
    Pretty, Kristin
    Donaldson, Nathan
    Parker, Sarah K.
    Dominguez, Samuel R.