Quasi-static pipeline in electroconvulsive therapy computational modeling

被引:4
|
作者
Unal, Gozde [1 ]
Poon, Cynthia [1 ]
FallahRad, Mohamad [1 ]
Thahsin, Myesha [1 ]
Argyelan, Miklos [2 ]
Bikson, Marom [1 ]
机构
[1] CUNY, Dept Biomed Engn, New York, NY 10031 USA
[2] North Shore Long Isl Jewish Hlth Syst, Feinstein Inst Med Res, Ctr Neurosci, Manhasset, NY 11030 USA
基金
美国国家卫生研究院;
关键词
DEEP BRAIN-STIMULATION; SEIZURE THRESHOLD; ELECTRICAL-PROPERTIES; UNIFORM ASSUMPTION; SKIN IMPEDANCE; DC STIMULATION; ECT; STIMULUS; EXCITATION; TISSUE;
D O I
10.1016/j.brs.2023.03.007
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background: Computational models of current flow during Electroconvulsive Therapy (ECT) rely on the quasi-static assumption, yet tissue impedance during ECT may be frequency specific and change adaptively to local electric field intensity.Objectives: We systematically consider the application of the quasi-static pipeline to ECT under conditions where 1) static impedance is measured before ECT and 2) during ECT when dynamic impedance is measured. We propose an update to ECT modeling accounting for frequency-dependent impedance.Methods: The frequency content on an ECT device output is analyzed. The ECT electrode-body impedance under low-current conditions is measured with an impedance analyzer. A framework for ECT modeling under quasi-static conditions based on a single device-specific frequency (e.g., 1 kHz) is proposed.Results: Impedance using ECT electrodes under low-current is frequency dependent and subject specific, and can be approximated at >100 Hz with a subject-specific lumped parameter circuit model but at <100 Hz increased non-linearly. The ECT device uses a 2 mA 800 Hz test signal and reports a static impedance that approximate 1 kHz impedance. Combined with prior evidence suggesting that conductivity does not vary significantly across ECT output frequencies at high-currents (800-900 mA), we update the adaptive pipeline for ECT modeling centered at 1 kHz frequency. Based on individual MRI and adaptive skin properties, models match static impedance (at 2 mA) and dynamic impedance (at 900 mA) of four ECT subjects.Conclusions: By considering ECT modeling at a single representative frequency, ECT adaptive and nonadaptive modeling can be rationalized under a quasi-static pipeline.& COPY; 2023 The Authors. Published by Elsevier Inc. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
下载
收藏
页码:607 / 618
页数:12
相关论文
共 50 条
  • [1] A computational method for quasi-static fracture
    Tabbara, MR
    Stone, CM
    COMPUTATIONAL MECHANICS, 1998, 22 (02) : 203 - 210
  • [2] A computational method for quasi-static fracture
    M. R. Tabbara
    C. M. Stone
    Computational Mechanics, 1998, 22 : 203 - 210
  • [3] COMPUTATIONAL ASPECTS OF QUASI-STATIC CRACK PROPAGATION
    Knees, Dorothee
    Schroeder, Andreas
    DISCRETE AND CONTINUOUS DYNAMICAL SYSTEMS-SERIES S, 2013, 6 (01): : 63 - 99
  • [4] Modeling of quasi-static ferromagnetic behavior
    Arib, M.
    Ille, J.L.
    Journal De Physique. IV : JP, 1998, 8 (02): : 607 - 610
  • [5] Quasi-static modeling of defected ground structure
    Karmakar, Nemai Chandra
    Roy, Sushim Mukul
    Balbin, Isaac
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2006, 54 (05) : 2160 - 2168
  • [6] A quasi-static modeling approach of electromechanical actuators
    Torabzadeh-Tari, M.
    Engdahl, G.
    JOURNAL OF MAGNETISM AND MAGNETIC MATERIALS, 2004, 272 : E1705 - E1707
  • [7] MAGNETIC DRAG IN THE QUASI-STATIC LIMIT - A COMPUTATIONAL METHOD
    MARCUSO, M
    GASS, R
    JONES, D
    ROWLETT, C
    AMERICAN JOURNAL OF PHYSICS, 1991, 59 (12) : 1118 - 1123
  • [8] Numerical modeling of quasi-static rock testing
    Bhide, R.J.
    McLennan, J.D.
    Guilkey, J.E.
    Green, S.J.
    45th US Rock Mechanics / Geomechanics Symposium, 2011,
  • [9] Quasi-Static Modeling of the da Vinci Instrument
    Anooshahpour, Farshad
    Polushin, Ilia G.
    Patel, Rajni V.
    2014 IEEE/RSJ INTERNATIONAL CONFERENCE ON INTELLIGENT ROBOTS AND SYSTEMS (IROS 2014), 2014, : 1308 - 1313
  • [10] Design and Modeling of a Quasi-Static Peristaltic Piezoelectric Micropump
    Beckers, Guillaume
    Dehez, Bruno
    2013 INTERNATIONAL CONFERENCE ON ELECTRICAL MACHINES AND SYSTEMS (ICEMS), 2013, : 1301 - 1306