Non-diffractive, toric, extended depth-of-focus intraocular lenses in eyes with low corneal astigmatism

被引:0
|
作者
Pastor-Pascual, Francisco [1 ]
Orts-Vila, Paz [2 ]
Tana-Sanz, Pedro [2 ]
Tana-Sanz, Santiago [2 ]
Ruiz-Mesa, Ramon [3 ]
Tana-Rivero, Pedro [2 ]
机构
[1] Oftalvist, C Ruzafa 19, Valencia 46004, Spain
[2] Oftalvist, Alicante, Spain
[3] Oftalvist, Jerez de la Frontera, Spain
关键词
EDOF; Toric; Intraocular lens; Astigmatism; Cataract; CATARACT-SURGERY; PREVALENCE;
D O I
10.1186/s40662-024-00380-7
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background To assess clinical outcomes after implanting toric, extended-depth-of-focus intraocular lenses (IOLs) to correct low corneal astigmatism in eyes with cataracts.Methods 47 eyes were implanted with the AcrySof IQ Vivity Toric DFT215 IOL. Main outcome measures were refractive error, monocular uncorrected and corrected distance (UDVA/CDVA), uncorrected and distance-corrected intermediate (UIVA/DCIVA), and uncorrected near and distance-corrected near (UNVA/DCNVA) visual acuities, monocular defocus curve, rotational stability, and IOLSAT and QUVID questionnaires. Patients were assessed at 3 months postsurgery.Results All eyes had a postoperative spherical equivalent (SE) within +/- 0.50 D and 97.87% (n = 46) had a refractive cylinder <= 0.50 D. The mean SE and refractive cylinder were - 0.10 +/- 0.17 D and - 0.16 +/- 0.24 D, respectively. The CDVA was >= 20/25 and >= 20/32 in 95.74% (n = 45) and 97.87% (n = 46) of eyes, respectively. The DCIVA was >= 20/32 in 85.11% (n = 40) of eyes and the DCNVA was >= 20/40 in 74.47% (n = 35). The mean values of CDVA, DCIVA, and DCNVA were - 0.02 +/- 0.08, 0.14 +/- 0.09, and 0.23 +/- 0.12 logMAR, respectively. The defocus curve revealed good visual acuity at far and intermediate distances with a depth-of-focus of about 1.75 D. IOL rotation was 0.74 +/- 1.13 degrees and all eyes had a rotation of less than 5 degrees. Patients reported either good or very good postoperative vision without eyeglasses under bright-light-conditions at distance (87.80%, 36/41) and intermediate distance (92.68%, 38/41). Between about 63.83%-72.34% (30-34) of patients reported no starburst, halos, or glare, or if experienced, were not bothersome.Conclusions The Vivity toric IOL implanted in eyes with low-astigmatism provides accurate refractive outcomes, good visual acuity at different distances and excellent rotational stability. Trial Registration The study was registered with the German Clinical Trials Register (DRKS00030579)Conclusions The Vivity toric IOL implanted in eyes with low-astigmatism provides accurate refractive outcomes, good visual acuity at different distances and excellent rotational stability. Trial Registration The study was registered with the German Clinical Trials Register (DRKS00030579)
引用
下载
收藏
页数:11
相关论文
共 50 条
  • [1] Non-diffractive, toric, extended depth-of-focus intraocular lenses in eyes with low corneal astigmatism
    Francisco Pastor-Pascual
    Paz Orts-Vila
    Pedro Tañá-Sanz
    Santiago Tañá-Sanz
    Ramón Ruiz-Mesa
    Pedro Tañá-Rivero
    Eye and Vision, 11
  • [2] Clinical Outcomes of a Non-Diffractive Extended Depth-of-Focus IOL in Eyes with Mild Glaucoma
    Ferguson, Tanner J.
    Wilson, Caroline W.
    Shafer, Brian M.
    Berdahl, John P.
    Terveen, Daniel C.
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 861 - 868
  • [3] Comparative Study Between Non-Diffractive Extended Depth of Focus and Monofocal Intraocular Lenses
    Kandavel, Rom
    Colvard, Michael
    Dredge, Justin
    Bayle, Michael
    Alam, Tamanna
    Snyder, James
    Hall, Brad
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 1161 - 1168
  • [4] Trifocal toric intraocular lenses in eyes with low amount of corneal astigmatism
    Orts-Vila, Paz
    Aguilar-Corcoles, Salvador
    Tello-Elordi, Cristina
    Ramos-Alzamora, Maria
    Montes-Mico, Robert
    Tana-Rivero, Pedro
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2020, 13 (10) : 1567 - 1573
  • [5] Trifocal toric intraocular lenses in eyes with low amount of corneal astigmatism
    Paz Orts-Vila
    Salvador Aguilar-Córcoles
    Cristina Tello-Elordi
    María Ramos-Alzamora
    Robert Montés-Micó
    Pedro Ta?á-Rivero
    International Journal of Ophthalmology, 2020, (10) : 1567 - 1573
  • [6] Visual Performance of Non-Diffractive Extended Depth of Focus and Neutral Aspheric Monofocal Intraocular Lenses
    O'Rourke, Tara
    Pantanelli, Seth
    Bolognia, Olga
    Longenecker, Amy
    Lehman, Erik
    Scruggs, Kathleen
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [7] Extended depth-of-focus technology in intraocular lenses
    Kohnen, Thomas
    Suryakumar, Rajaraman
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (02): : 298 - 304
  • [8] Visual function in eyes with diffractive extended depth-of-focus and monofocal intraocular lenses: 2-year comparison
    Takahiro Hisai
    Keita Takada
    Shota Tokuda
    Ryohei Nejima
    Yosai Mori
    Keiichiro Minami
    Kazunori Miyata
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2023, 261 : 2567 - 2573
  • [9] Visual function in eyes with diffractive extended depth-of-focus and monofocal intraocular lenses: 2-year comparison
    Hisai, Takahiro
    Takada, Keita
    Tokuda, Shota
    Nejima, Ryohei
    Mori, Yosai
    Minami, Keiichiro
    Miyata, Kazunori
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2023, 261 (09) : 2567 - 2573
  • [10] Clinical Outcomes of a New Non-Diffractive Extended Depth-of-Focus Intraocular Lens Targeted for Mini-Monovision
    Tomagova, Nikola
    Elahi, Sina
    Vandekerckhove, Kristof
    CLINICAL OPHTHALMOLOGY, 2023, 17 : 981 - 990