Accuracy of maxillary full-arch digital impressions of tooth and implant models made by two intraoral scanners

被引:1
|
作者
Siadat, Hakimeh [1 ]
Chitsaz, Foujan [2 ]
Zeighami, Somayeh [1 ,4 ]
Esmaeilzadeh, Alireza [3 ]
机构
[1] Univ Tehran Med Sci, Dent Res Inst, Dept Prosthodont, Sch Dent, Tehran, Iran
[2] Univ New South Wales, Sch Mech & Mfg Engn, Sydney, NSW, Australia
[3] Arak Univ Med Sci, Sch Dent, Dept Periodont, Arak, Iran
[4] Univ Tehran Med Sci, Dent Res Inst, Dept Prosthodont, Tehran, Iran
来源
关键词
dental implants; dental impression technique; dimensional measurement accuracy; IN-VIVO PRECISION; CONVENTIONAL IMPRESSIONS; MULTIPLE IMPLANTS; SCANNING SYSTEMS; TECHNOLOGY; VITRO; TRUENESS; LIGHT; TIME;
D O I
10.1002/cre2.857
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: Limited studies are available on the accuracy of intraoral scanners (IOSs) for full-arch implant and tooth models. This study aimed to assess the accuracy of maxillary full-arch digital impressions of tooth and implant models made by two IOSs. Materials and Methods: This in vitro, experimental study was conducted on two maxillary dentiform models: one with six prepared natural teeth and the other with six implants at the site of canine, first premolar, and first molar teeth, bilaterally. A highly accurate industrial scanner was used for actual measurements on the models that served as the reference scan. TS (Trios3) and CO (CEREC Omnicam) IOSs were then used to scan each model 10 times according to the manufacturer's instructions. All scans were saved in STL format. The GOM Inspect software was used according to the best-fit algorithm to compare the accuracy of measurements in the groups with the reference scan. The trueness and precision were calculated. Statistical analyses were carried out using SPSS by one-way analysis of variance and t-test (alpha = .05). Results: TS showed a significantly higher trueness than CO for both tooth and implant models (p < .05). TS also revealed significantly higher precision than CO for the tooth model; however, the difference in precision for the implant model was not significant between the two IOSs (p > .05). Conclusions: TS showed higher accuracy than CO in both tooth and implant models.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Comparison of 3D accuracy of three different digital intraoral scanners in full-arch implant impressions
    Akkal, Ozcan
    Korkmaz, Ismail Hakki
    Bayindir, Funda
    [J]. JOURNAL OF ADVANCED PROSTHODONTICS, 2023, 15 (04): : 179 - 188
  • [2] Accuracy of full-arch digital implant impressions taken using intraoral scanners and related variables: A systematic review
    Zhang, Yi-Jie
    Shi, Jun-Yu
    Qian, Shu-Jiao
    Qiao, Shi-Chong
    Lai, Hong-Chang
    [J]. INTERNATIONAL JOURNAL OF ORAL IMPLANTOLOGY, 2021, 14 (02) : 157 - 179
  • [3] In vitro and in vivo accuracy of full-arch digital implant impressions
    Rutkunas, Vygandas
    Gedrimiene, Agne
    Akulauskas, Mykolas
    Fehmer, Vincent
    Sailer, Irena
    Jegelevicius, Darius
    [J]. CLINICAL ORAL IMPLANTS RESEARCH, 2021, 32 (12) : 1444 - 1454
  • [4] Accuracy of full-arch scans using intraoral scanners
    Patzelt, Sebastian B. M.
    Emmanouilidi, Archontia
    Stampf, Susanne
    Strub, Joerg R.
    Att, Wael
    [J]. CLINICAL ORAL INVESTIGATIONS, 2014, 18 (06) : 1687 - 1694
  • [5] Accuracy of full-arch scans using intraoral scanners
    Sebastian B. M. Patzelt
    Archontia Emmanouilidi
    Susanne Stampf
    Joerg R. Strub
    Wael Att
    [J]. Clinical Oral Investigations, 2014, 18 : 1687 - 1694
  • [6] Accuracy, reliability, and efficiency of intraoral scanners for full-arch impressions: a systematic review of the clinical evidence
    Goracci, Cecilia
    Franchi, Lorenzo
    Vichi, Alessandro
    Ferrari, Marco
    [J]. EUROPEAN JOURNAL OF ORTHODONTICS, 2016, 38 (04) : 422 - 428
  • [7] Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study
    Di Fiore, Adolfo
    Graiff, Lorenzo
    Savio, Gianpaolo
    Granata, Stefano
    Basilicata, Michele
    Bollero, Patrizio
    Meneghello, Roberto
    [J]. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 2022, 19 (08)
  • [8] Effect of Scan Pattern on the Accuracy of Complete-Arch Digital Implant Impressions with Two Intraoral Scanners
    Li, Zhipeng
    Huang, Ruoxuan
    Wu, Xiayi
    Chen, Zetao
    Huang, Baoxin
    Chen, Zhuofan
    [J]. INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (04) : 731 - 739
  • [9] In vivo full-arch accuracy of intraoral scanners: a narrative review
    Jennes, Marie-Elise
    Soetebeer, Maren
    Beuer, Florian
    [J]. INTERNATIONAL JOURNAL OF COMPUTERIZED DENTISTRY, 2022, 25 (01) : 9 - 16
  • [10] Linear Accuracy of Intraoral Scanners for Full-Arch Impressions of Implant-Supported Prostheses: A Systematic Review and Meta-Analysis
    Floriani, Franciele
    Lopes, Guilherme Carpena
    Cabrera, Alexandre
    Duarte, Wagner
    Zoidis, Panagiotis
    Oliveira, Dayane
    Rocha, Mateus Garcia
    [J]. EUROPEAN JOURNAL OF DENTISTRY, 2023, 17 (04) : 964 - 973