Continuous Vertical Inside-Out Versus Traditional Vertical Inside-Out Meniscal Repair: A Biomechanical Comparison

被引:1
|
作者
de Faria, Jose Leonardo Rocha [1 ,2 ,3 ,4 ]
Santos, Arthur Paiva Grimaldi [1 ]
Pavao, Douglas Mello [1 ,3 ,4 ]
Radulski, Mariana [1 ,5 ]
Leal, Ana Carolina [1 ,6 ]
More, Ari Digiacomo Ocampo [1 ,5 ]
de Paula Mozella, Alan [1 ,3 ,4 ]
Guimaraes, Joao Antonio Matheus [1 ,6 ]
LaPrade, Robert F. [1 ,7 ,8 ]
de Mello Roesler, Carlos Rodrigo [1 ]
Salim, Rodrigo [1 ,4 ]
机构
[1] Univ Fed Santa Catarina, Biomech Engn Lab, Florianopolis, SC, Brazil
[2] Natl Inst Traumatol & Orthoped INTO, Ave Brasil 500, BR-20940070 Rio De Janeiro, RJ, Brazil
[3] Natl Inst Traumatol & Orthoped INTO, Knee Surg Ctr, Rio De Janeiro, Brazil
[4] Univ Sao Paulo, Sch Med, Ribeirao Preto, SP, Brazil
[5] Univ Fed Santa Catarina, Med Sch, Florianopolis, SC, Brazil
[6] Natl Inst Traumatol & Orthoped INTO, Res Div, Rio De Janeiro, Brazil
[7] Twin Cities Orthopaed, Edina, MN USA
[8] Univ Minnesota, Sch Med, Minneapolis, MN USA
关键词
biomechanics; inside-out; meniscal repair; primary fixation; SUTURE MATERIAL; FAILURE; STRENGTH; SYSTEM; LOAD;
D O I
10.1177/23259671231209951
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background:Biomechanical assessment of meniscal repairs is essential for evaluating different meniscal suturing methods and techniques. The continuous meniscal suture technique is a newer method of meniscal repair that may have biomechanical differences compared with traditional techniques.Purpose:To evaluate the displacement, stiffness after cyclical loading, and load to failure for a continuous vertical inside-out meniscal suture versus a traditional vertical inside-out meniscal suture in a porcine medial meniscus.Study Design:Controlled laboratory study.Methods:A total of 28 porcine knees were acquired and divided into 2 test groups of 14 medial meniscus each. A 2.0-cm longitudinal red-white zone cut was made in the body of the medial meniscus for each knee. The continuous suture (CS) group received 4 vertical stitches performed with a continuous vertical meniscal suture technique, and the inside-out suture (IO) group received a traditional vertical suture with 4 stitches. Two traction tapes were passed between the sutures and positioned in the biomechanical testing fixture device. Each specimen underwent load-to-failure testing at 5 mm/s, and displacement, system stiffness, and maximum load to failure were compared between the groups.Results:The displacement after the cyclic test was 0.53 +/- 0.12 and 0.48 +/- 0.07 mm for the CS and IO groups, respectively. There was no significant difference between the groups (P = .2792). The stiffness at the ultimate load testing was 36.3 +/- 1.9 and 35.3 +/- 2.4 N/mm for groups CS and IO, respectively, with no significant difference between the groups (P = .2557). In the load-to-failure test, the ultimate load was 218.2 +/- 63.9 and 238.3 +/- 71.3 N in the CS and IO groups, respectively, with no significant group differences (P = .3062).Conclusion:A continuous vertical meniscal suture created a configuration for treating longitudinal meniscal lesions that was beneficial and biomechanically similar to a traditional vertical suture technique.Clinical Relevance:The study findings indicate that use of the continuous vertical inside-out meniscal suture technique is a possible therapeutic option.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Modified inside-out technique for meniscal repair
    Ahn, JH
    Wang, JH
    Oh, I
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2004, 20 (06): : 178 - 182
  • [2] A Comparison of All-inside and Inside-out Meniscal Repair in Elite Athletes
    Borque, Kyle A.
    Laughlin, Mitzi S.
    Webster, Emily
    Jones, Mary
    Pinheiro, Vitor Hugo
    Williams, Andy
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2023, 51 (03): : 579 - 584
  • [3] Meniscal cyst formation after inside-out meniscal repair
    Choi, NH
    Kim, SJ
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2004, 20 (01): : e1
  • [4] Comparison of Arthroscopic Medial Meniscal Suture Repair Techniques Inside-Out Versus All-Inside Repair
    Choi, Nam-Hong
    Kim, Tae-Hyung
    Victoroff, Brian N.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2009, 37 (11): : 2144 - 2150
  • [5] Meniscal Repair Using the Inside-Out Suture Technique
    Johnson, Don
    Weiss, William M.
    [J]. CLINICS IN SPORTS MEDICINE, 2012, 31 (01) : 15 - +
  • [6] Arthroscopic repair of meniscal tears with inside-out technique
    Raza, Hasnain
    Abbas, Kashif
    Umer, Masood
    [J]. JOURNAL OF THE PAKISTAN MEDICAL ASSOCIATION, 2011, 61 (01) : 10 - 14
  • [7] Inside-Out Meniscal Repair: Medial and Lateral Approach
    Chahla, Jorge
    Cruz, Raphael Serra
    Cram, Tyler R.
    Dean, Chase S.
    LaPrade, Robert F.
    [J]. ARTHROSCOPY TECHNIQUES, 2016, 5 (01): : E163 - E168
  • [8] Arthroscopic meniscal repair using the inside-out technique
    Van der Reis, W
    Cannon, WD
    [J]. SPORTS MEDICINE AND ARTHROSCOPY REVIEW, 1999, 7 (01): : 8 - 19
  • [9] Meniscal Repair Using the Inside-Out Suture Technique
    Johnson, Donald
    Weiss, Bill
    [J]. SPORTS MEDICINE AND ARTHROSCOPY REVIEW, 2012, 20 (02): : 68 - 76
  • [10] Comparison of All-Inside Meniscal Repair Devices With Matched Inside-Out Suture Repair
    Rosso, Claudio
    Kovtun, Konstantin
    Dow, William
    McKenzie, Brett
    Nazarian, Ara
    DeAngelis, Joseph P.
    Ramappa, Arun J.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2011, 39 (12): : 2634 - 2639