Biomechanical comparative finite element analysis between a conventional proximal interphalangeal joint flexible hinge implant and a novel implant design using a rolling contact joint mechanism

被引:1
|
作者
Kim, Yong-Jae [1 ]
Bae, Hyun-ah [1 ]
Hong, Seok Woo [2 ]
机构
[1] Korea Univ Technol & Educ, Sch Elect Elect & Commun Engn, 1600 ChungJeol Ro, Cheonan 31253, South Korea
[2] Sungkyunkwan Univ, Kangbuk Samsung Hosp, Dept Orthopaed Surg, Sch Med, 29 Saemunan Ro, Seoul 03181, South Korea
关键词
Proximal interphalangeal joint prosthesis; Flexible hinge implant; Rolling contact joint mechanism; Finite element analysis; Von-Mises stress; Von-Mises strain; Moment reactions; ARTHROPLASTY; RUBBER; OSTEOARTHRITIS; NEUFLEX; MODELS;
D O I
10.1186/s13018-023-04477-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundThe rolling contact joint (RCJ) mechanism is a system of constraint that allows two circular bodies connected with flexible straps to roll relative to one another without slipping. This study aims to compare the biomechanical characteristics between the conventional proximal interphalangeal joint (PIPJ) flexible hinge (FH) implant and the novel PIPJ implant adopting a RCJ mechanism during PIPJ range of motion using finite element (FE) analysis.MethodsThe three-dimensional (3D) surface shape of a conventional PIPJ FH implant was obtained using a 3D laser surface scanning system. The configuration and parameters of the novel PIPJ implant were adapted from a previous study. The two implants were assumed to have the same material characteristics and each implant was composed of a hyperelastic material, silicone elastomers. The configuration data for both implants were imported to a computer-aided design program to generate 3D geometrical surface and hyperelastic models of both implants. The hyperelastic models of both implants were imported into a structural engineering software to produce the FE mesh and to perform FE analysis. The FE analysis modeled the changes of mechanics during flexion-extension motion between 0 degrees and 90 degrees of two PIPJ implants. The mean and maximum values of von-Mises stress and strain as well as the total moment reaction based on the range of motion of the PIPJs were calculated. The mean values within the PIPJ's functional range of motion of the mean and maxinum von-Mises stress and strain and the total moment reaction were also determined.ResultsThe maximum values for the von-Mises stress, and strain, as well as the total moment reactions of the conventional PIPJ FH and novel PIPJ implants were all at 90 degrees of PIPJ flexion. The maximum value of each biomechanical property for the novel PIPJ implant was considerably lower compared with that of the conventional PIPJ FH implant. The mean values within the PIPJ's functional range of motion of the maximum von-Mises stress and strain for the novel PIPJ implant was approximately 6.43- and 6.46-fold lower compared with that of the conventional PIPJ FH implant, respectively. The mean value within a PIPJ's functional range of motion of the total moment reaction of the novel PIPJ implant was approximately 49.6-fold lower compared with that of the conventional PIPJ FH implant.ConclusionsThe novel PIPJ implant with an RCJ mechanism may offer improved biomechanical performance compared with conventional PIPJ FH implant.
引用
收藏
页数:13
相关论文
共 27 条
  • [21] Analysis and optimal design a new flexible hinge displacement amplifier mechanism by using Finite element analysis based on Taguchi method
    Viet-Hung Hoang
    Ngoc-Thai Huynh
    Ho Nguyen
    Huang, Shyh-Chour
    PROCEEDINGS OF THE 2019 IEEE EURASIA CONFERENCE ON IOT, COMMUNICATION AND ENGINEERING (ECICE), 2019, : 259 - 262
  • [22] Reduction of Dorsal Displacement of the Proximal and Middle Phalanges Using a Neutral or Angled Implant for Joint Arthrodesis to Treat Hammertoe Deformity A Finite Element Study
    Bayod Lopez, Javier
    de Bengoa Vallejo, Ricardo Becerro
    Losa Iglesias, Marta Elena
    Jules, Kevin T.
    Doblare, Manuel
    JOURNAL OF THE AMERICAN PODIATRIC MEDICAL ASSOCIATION, 2015, 105 (06) : 493 - 502
  • [23] Letter to the editor regarding "Finite element analysis of subtalar joint arthroereisis on adult-acquired flexible flatfoot deformity using customised sinus tarsi implant"
    Chen, Cheng
    Shi, ZhongMin
    JOURNAL OF ORTHOPAEDIC TRANSLATION, 2022, 33 : 70 - 71
  • [24] Response to the letter to editor regarding "finite element analysis of subtalar joint arthroereisis on adult acquired flexible flatfoot deformity using customized sinus tarsi implant"
    Wong, Duo Wai-Chi
    Wang, Yan
    Niu, Wenxin
    Zhang, Ming
    JOURNAL OF ORTHOPAEDIC TRANSLATION, 2022, 37 : 173 - 174
  • [25] Biomechanical assessment of a novel L4/5 level interspinous implant using three dimensional finite element analysis
    Song, C.
    Li, X. -F.
    Liu, Z. -D.
    Zhong, G. -B.
    EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 2014, 18 (01) : 86 - 94
  • [26] COMPARATIVE STRESS ANALYSIS OF LINGUALIZED AND CONVENTIONAL BALANCED OCCLUSION SCHEMES IN A FULL-ARCH FIXED IMPLANT PROSTHESIS USING FINITE ELEMENT ANALYSIS
    Abu Hasan, Md
    Shiakolas, Panos S.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2017, 17 (04)
  • [27] Finite Element Analysis to determine the optimum contact pressure between the components of a hip implant made by using the Selective Laser Sintering and the Selective Laser Melting Technologies
    Pacurar, Razvan
    Pacurar, Ancuta
    Petrilak, Anna
    MODERN TECHNOLOGIES IN MANUFACTURING (MTEM 2017 - AMATUC), 2017, 137