Biomechanical comparative finite element analysis between a conventional proximal interphalangeal joint flexible hinge implant and a novel implant design using a rolling contact joint mechanism

被引:1
|
作者
Kim, Yong-Jae [1 ]
Bae, Hyun-ah [1 ]
Hong, Seok Woo [2 ]
机构
[1] Korea Univ Technol & Educ, Sch Elect Elect & Commun Engn, 1600 ChungJeol Ro, Cheonan 31253, South Korea
[2] Sungkyunkwan Univ, Kangbuk Samsung Hosp, Dept Orthopaed Surg, Sch Med, 29 Saemunan Ro, Seoul 03181, South Korea
关键词
Proximal interphalangeal joint prosthesis; Flexible hinge implant; Rolling contact joint mechanism; Finite element analysis; Von-Mises stress; Von-Mises strain; Moment reactions; ARTHROPLASTY; RUBBER; OSTEOARTHRITIS; NEUFLEX; MODELS;
D O I
10.1186/s13018-023-04477-y
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
BackgroundThe rolling contact joint (RCJ) mechanism is a system of constraint that allows two circular bodies connected with flexible straps to roll relative to one another without slipping. This study aims to compare the biomechanical characteristics between the conventional proximal interphalangeal joint (PIPJ) flexible hinge (FH) implant and the novel PIPJ implant adopting a RCJ mechanism during PIPJ range of motion using finite element (FE) analysis.MethodsThe three-dimensional (3D) surface shape of a conventional PIPJ FH implant was obtained using a 3D laser surface scanning system. The configuration and parameters of the novel PIPJ implant were adapted from a previous study. The two implants were assumed to have the same material characteristics and each implant was composed of a hyperelastic material, silicone elastomers. The configuration data for both implants were imported to a computer-aided design program to generate 3D geometrical surface and hyperelastic models of both implants. The hyperelastic models of both implants were imported into a structural engineering software to produce the FE mesh and to perform FE analysis. The FE analysis modeled the changes of mechanics during flexion-extension motion between 0 degrees and 90 degrees of two PIPJ implants. The mean and maximum values of von-Mises stress and strain as well as the total moment reaction based on the range of motion of the PIPJs were calculated. The mean values within the PIPJ's functional range of motion of the mean and maxinum von-Mises stress and strain and the total moment reaction were also determined.ResultsThe maximum values for the von-Mises stress, and strain, as well as the total moment reactions of the conventional PIPJ FH and novel PIPJ implants were all at 90 degrees of PIPJ flexion. The maximum value of each biomechanical property for the novel PIPJ implant was considerably lower compared with that of the conventional PIPJ FH implant. The mean values within the PIPJ's functional range of motion of the maximum von-Mises stress and strain for the novel PIPJ implant was approximately 6.43- and 6.46-fold lower compared with that of the conventional PIPJ FH implant, respectively. The mean value within a PIPJ's functional range of motion of the total moment reaction of the novel PIPJ implant was approximately 49.6-fold lower compared with that of the conventional PIPJ FH implant.ConclusionsThe novel PIPJ implant with an RCJ mechanism may offer improved biomechanical performance compared with conventional PIPJ FH implant.
引用
收藏
页数:13
相关论文
共 27 条
  • [1] Biomechanical comparative finite element analysis between a conventional proximal interphalangeal joint flexible hinge implant and a novel implant design using a rolling contact joint mechanism
    Yong-Jae Kim
    Hyun-ah Bae
    Seok Woo Hong
    Journal of Orthopaedic Surgery and Research, 18
  • [2] Novel implant design of the proximal interphalangeal joint using an optimized rolling contact joint mechanism
    Seok Woo Hong
    Junsuk Yoon
    Yong-Jae Kim
    Hyun Sik Gong
    Journal of Orthopaedic Surgery and Research, 14
  • [3] Novel implant design of the proximal interphalangeal joint using an optimized rolling contact joint mechanism
    Hong, Seok Woo
    Yoon, Junsuk
    Kim, Yong-Jae
    Gong, Hyun Sik
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2019, 14 (1)
  • [4] Two Implant/Abutment Joint Designs: A Comparative Finite Element Analysis
    Khraisat, Ameen
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (02) : E83 - E87
  • [5] Proximal Interphalangeal Joint Arthroplasty Using a Silicone Implant: A Comparison Between Integra and NeuFlex in 72 Cases
    Weistra, Kelly
    Kan, Hester J.
    van Alebeek, Veerle A. H. J.
    Ritt, Marco J. P. F.
    HAND-AMERICAN ASSOCIATION FOR HAND SURGERY, 2024, 19 (02): : 236 - 240
  • [6] Optimal design of a hip joint implant with hallow stem using finite element method
    Zakeri, Nasser
    Farahmand, Farzam
    Katoozian, Hamid
    Proceedings of the 8th Biennial Conference on Engineering Systems Design and Analysis, Vol 2, 2006, : 505 - 510
  • [7] Biomechanical Comparison of a New Triple Cylindrical Implant Design and a Conventional Cylindrical Implant Design on the Mandible by Three-Dimensions Finite Element Analysis
    Bayrak, Ayben
    Yaramanoglu, Panur
    Kilicarslan, Mehmet Ali
    Yaramanoglu, Betul
    Akat, Bora
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2020, 35 (02) : 257 - 264
  • [8] Finite element analysis of knee joint implant for varying bio material using ANSYS
    Kohli, Anirudh
    Mathad, Manoj
    Hosamani, Santoshkumar, V
    Adagimath, Mrutyunjay K.
    Kotturshettar, Basavaraj B.
    MATERIALS TODAY-PROCEEDINGS, 2022, 59 : 941 - 950
  • [9] Evaluation of the glenoid implant survival using a biomechanical finite element analysis: Influence of the implant design, bone properties, and loading location
    Mansat, P.
    Briot, J.
    Mansat, M.
    Swider, P.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2007, 16 (03) : 79S - 83S
  • [10] Biomechanical Analysis of Patient-Specific Temporomandibular Joint Implant and Comparison with Natural Intact Jaw Bone Using Finite Element Method
    Tiwari, Anubhav
    Gupta, Vijay Kumar
    Haldkar, Rakesh Kumar
    Parinov, Ivan A.
    APPLIED SCIENCES-BASEL, 2022, 12 (06):