A comparison of the visual performance of refractive rotationally asymmetric multifocal intraocular lenses and trifocal intraocular lenses

被引:0
|
作者
Qiu, Jing-Jing [1 ]
Zhang, Qian [1 ]
Fu, Shu-Hua [1 ]
Huang, Jun-Qi [2 ]
Wang, Zhi-Hong [3 ]
You, Zhi-Peng [1 ]
Liu, Fei [1 ]
Sun, Tao [2 ]
机构
[1] Nanchang Univ, Dept Ophthalmol, Affiliated Hosp 2, Nanchang 330006, Peoples R China
[2] Nanchang Univ, Dept Ophthalmol, Affiliated Eye Hosp, Nanchang 330006, Peoples R China
[3] Dept Jiujiang 1 Peoples Hosp, Ophthalmol, Jiujiang 332000, Peoples R China
关键词
Cataract; Phacoemulsification; Multifocal Lens; Trifocal Lens; Intraocular Lens; BILATERAL IMPLANTATION; OUTCOMES; EXPERIENCE; CATARACT;
D O I
10.1166/mex.2023.2445
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Objective: The aim of this study is to compare the short term visual outcomes and subjective visual quality of the monocular implantation of a refractive rotationally asymmetric multifocal intraocular lens (IOL) (SBL-3, Lenstec, Inc., USA) with those of a trifocalIOL (AT LISA tri 839 MP, Carl Zeiss Meditec, Jena, Germany). IP: 2038 109 20 On: Tue 08 Aug 2023 05:04:48 Methods: This study involved a totalCopyright:of46 Americapatients (46Sc ntificeyes) Pubwith shrsage-related cataracts, who underwent uncomplicated cataract surgery. Twenty-two patients (22 eys) had SBL-3 multifocal IOLs implanted (group Delivered by ngenta A) and 24 patients (24 eyes) had AT LISA tri 839 MPs implanted (group B). The patients were all evaluated three months after surgery. Results: Group B had a significantly better performance (0.127 & PLUSMN; 0.070) logMAR (20/40 Snellen equivalent) than group A (0.25 & PLUSMN; 0.093) LogMAR for monocular UIVA at 80 cm (p < 0.05). There was no statistically significant difference at other distances. The monocular defocus curves showed that group B had a better performance than group A at a vergence of -1.5 D (p < 0.05). Higher-order aberrations (HOAs) were significantly higher in group A (p < 0.05). The SBL-3 IOLs resulted in much lower MTF cut-off (23.68 & PLUSMN; 9.892) cpd and Strehl ratio (0.13 & PLUSMN; 0.052) than trifocal IOLs (MTF cut-off 26.25 & PLUSMN; 8.751 cpd, Strehl ratio 0.15 & PLUSMN; 0.063). But there was no significant difference between the two IOLs (P > 0.05). Complaints of halos at night were made by 37.5% (9/24) of group A and 9.1% (2/22) of group B, but they did not interfere with their vision. All the patients reported a high level of satisfaction and said they no longer needed to use spectacle correction at either far, intermediate or near distances. Conclusions: Both SBL-3 IOLs and AT LISA tri 839 MP trifocal IOLs provided good short-term visual outcomes at all distances postoperatively with a high level of spectacle independence and patient satisfaction. The AT LISA tri 839 MP lens provided better intermediate vision at 80 cm, while similar visual acuity (VA) was reported at other distances.
引用
下载
收藏
页码:1110 / 1115
页数:6
相关论文
共 50 条
  • [1] Visual quality and performance comparison between 2 refractive rotationally asymmetric multifocal intraocular lenses
    McNeely, Richard N.
    Pazo, Eric
    Spence, Andrew
    Richoz, Olivier
    Nesbit, M. Andrew
    Moore, Tara C. B.
    Moore, Jonathan E.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2017, 43 (08): : 1020 - 1026
  • [2] Comparison of visual and refractive outcomes of 2 trifocal intraocular lenses
    Ribeiro, Filomena J.
    Ferreira, Tiago B.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (05): : 694 - 699
  • [3] Comparative visual performance of diffractive bifocal and rotationally asymmetric refractive intraocular lenses
    Tanabe, Hirotaka
    Shojo, Tomohiro
    Yamauchi, Tomofusa
    Takase, Kosuke
    Akada, Masahiro
    Tabuchi, Hitoshi
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [4] Comparative visual performance of diffractive bifocal and rotationally asymmetric refractive intraocular lenses
    Hirotaka Tanabe
    Tomohiro Shojo
    Tomofusa Yamauchi
    Kosuke Takase
    Masahiro Akada
    Hitoshi Tabuchi
    Scientific Reports, 12
  • [5] Comparison of Visual Performance of Multifocal Intraocular Lenses with Same Material Monofocal Intraocular Lenses
    Yamauchi, Tomofusa
    Tabuchi, Hitoshi
    Takase, Kosuke
    Ohsugi, Hideharu
    Ohara, Zaigen
    Kiuchi, Yoshiaki
    PLOS ONE, 2013, 8 (06):
  • [6] VISUAL AND REFRACTIVE RESULTS OF MULTIFOCAL INTRAOCULAR LENSES
    GIMBEL, HV
    SANDERS, DR
    RAANAN, MG
    OPHTHALMOLOGY, 1991, 98 (06) : 881 - 888
  • [7] Comparison of Visual Performance between Bifocal and Rotationally Asymmetric Multifocal with+1.5 Diopters Near Addition Intraocular Lenses
    Tanabe, Hirotaka
    Shojo, Tomohiro
    Yamauchi, Tomofusa
    Takase, Kosuke
    Tabuchi, Hitoshi
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2021, 62 (08)
  • [8] Visual performance with accommodating and multifocal intraocular lenses
    Lan, Jie
    Huang, Yu-Sen
    Dai, Yun-Hai
    Wu, Xiao-Ming
    Sun, Jia-Jun
    Xie, Li-Xin
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2017, 10 (02) : 235 - 240
  • [9] Visual performance with accommodating and multifocal intraocular lenses
    Jie Lan
    Yu-Sen Huang
    Yun-Hai Dai
    Xiao-Ming Wu
    Jia-Jun Sun
    Li-Xin Xie
    International Journal of Ophthalmology, 2017, 10 (02) : 235 - 240
  • [10] Halos in refractive multifocal intraocular lenses
    Pieh, S
    Zöhrer, R
    Backner, B
    Hanselmayer, G
    Ruhswurm, I
    Scholz, U
    Weghaupt, H
    Skorpik, C
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S298 - S298