Digital Breast Tomosynthesis: Towards Dose Reduction through Image Quality Improvement

被引:2
|
作者
Mota, Ana M. [1 ]
Mendes, Joao [1 ,2 ]
Matela, Nuno [1 ]
机构
[1] Univ Lisbon, Fac Ciencias, Inst Biofis & Engn Biomed, P-1749016 Lisbon, Portugal
[2] Univ Lisbon, Fac Ciencias, LASIGE, P-1749016 Lisbon, Portugal
关键词
digital breast tomosynthesis; total variation minimization; image quality; dose reduction; TOTAL VARIATION MINIMIZATION; MAMMOGRAPHY; RECONSTRUCTION;
D O I
10.3390/jimaging9060119
中图分类号
TB8 [摄影技术];
学科分类号
0804 ;
摘要
Currently, breast cancer is the most commonly diagnosed type of cancer worldwide. Digital Breast Tomosynthesis (DBT) has been widely accepted as a stand-alone modality to replace Digital Mammography, particularly in denser breasts. However, the image quality improvement provided by DBT is accompanied by an increase in the radiation dose for the patient. Here, a method based on 2D Total Variation (2D TV) minimization to improve image quality without the need to increase the dose was proposed. Two phantoms were used to acquire data at different dose ranges (0.88-2.19 mGy for Gammex 156 and 0.65-1.71 mGy for our phantom). A 2D TV minimization filter was applied to the data, and the image quality was assessed through contrast-to-noise ratio (CNR) and the detectability index of lesions before and after filtering. The results showed a decrease in 2D TV values after filtering, with variations of up to 31%, increasing image quality. The increase in CNR values after filtering showed that it is possible to use lower doses (-26%, on average) without compromising on image quality. The detectability index had substantial increases (up to 14%), especially in smaller lesions. So, not only did the proposed approach allow for the enhancement of image quality without increasing the dose, but it also improved the chances of detecting small lesions that could be overlooked.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Digital breast tomosynthesis: Dose and image quality assessment
    Maldera, A.
    De Marco, P.
    Colombo, P. E.
    Origgi, D.
    Torresin, A.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2017, 33 : 56 - 67
  • [2] RELATIONSHIP BETWEEN RADIATION DOSE AND IMAGE QUALITY IN DIGITAL BREAST TOMOSYNTHESIS
    Alakhras, Maram M.
    Mello-Thoms, Claudia
    Bourne, Roger
    Rickard, Mary
    Diffey, Jennifer
    Brennan, Patrick C.
    [J]. RADIATION PROTECTION DOSIMETRY, 2017, 173 (04) : 351 - 360
  • [3] Simulation of Dose Reduction in Digital Breast Tomosynthesis
    Borges, Lucas R.
    Guerrero, Igor
    Bakic, Predrag R.
    Maidment, Andrew D. A.
    Schiabel, Homero
    Vieira, Marcelo A. C.
    [J]. BREAST IMAGING, IWDM 2016, 2016, 9699 : 343 - 350
  • [4] Image quality and dose assessment in digital breast tomosynthesis: A Monte Carlo study
    Baptista, M.
    Di Maria, S.
    Oliveira, N.
    Matela, N.
    Janeiro, L.
    Almeida, P.
    Vaz, P.
    [J]. RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 158 - 162
  • [5] EFFECT OF THE GLANDULAR COMPOSITION ON DIGITAL BREAST TOMOSYNTHESIS IMAGE QUALITY AND DOSE OPTIMISATION
    Marques, T.
    Ribeiro, A.
    Di Maria, S.
    Belchior, A.
    Cardoso, J.
    Matela, N.
    Oliveira, N.
    Janeiro, L.
    Almeida, P.
    Vaz, P.
    [J]. RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 337 - 341
  • [6] Method for Simulating Dose Reduction in Digital Breast Tomosynthesis
    Borges, Lucas R.
    Guerrero, Igor
    Bakic, Predrag R.
    Foi, Alessandro
    Maidment, Andrew D. A.
    Vieira, Marcelo A. C.
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2017, 36 (11) : 2331 - 2342
  • [7] Image quality evaluation and improvement of a digital tomosynthesis system
    Kang, ST
    Cho, HS
    [J]. MATERIALS EVALUATION, 1999, 57 (08) : 841 - 845
  • [8] Dose Reduction in Digital Breast Tomosynthesis with the Dose Reduction Prior Image Constrained Compressed Sensing (DR-PICCS) Algorithm
    Garrett, J.
    Tang, J.
    Zhang, Y.
    Ruth, C.
    Jing, Z.
    Chen, G. H.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3916 - 3916
  • [9] Evaluating the effect of dose on reconstructed image quality in digital tomosynthesis
    Kempston, Michael P.
    Mainprize, James G.
    Yaffe, Martin J.
    [J]. DIGITAL MAMOGRAPHY, PROCEEDINGS, 2006, 4046 : 490 - 497
  • [10] Comparison of effects of dose on image quality in digital breast tomosynthesis systems across multiple vendors
    Zhao, Amy
    Santana, Maira
    Samei, Ehsan
    Lo, Joseph
    [J]. MEDICAL IMAGING 2017: PHYSICS OF MEDICAL IMAGING, 2017, 10132