Calibration of a compact ASIC-based data acquisition system for neutron/γ discrimination and spectroscopy with organic scintillators

被引:0
|
作者
Wu, T. C. [1 ]
Sutanto, F. [2 ]
Li, V. A. [2 ]
Classen, T. M. [2 ]
Dazeley, S. [2 ]
Jovanovic, I. [1 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
[2] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
关键词
Fast-neutron detection; Pulse-shape discrimination; SiPM readout; PLASTIC SCINTILLATORS; MODEL;
D O I
10.1016/j.nima.2023.168699
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Segmented neutron detectors that use silicon photomultipliers (SiPMs) are receiving significant attention in nuclear security applications. Some of these detectors employ hundreds of channels and would therefore benefit from the use of high-channel-density and low-cost-per-channel data acquisition (DAQ) systems. Candidate DAQ systems that meet these requirements exist, but few perform full waveform digitization, which permits neutron and gamma-ray interaction discrimination via pulse shape. In this work, we study the performance of the TOFPET2 (PETsys Electronics), an ASIC-based DAQ designed for positron emission tomography, which has been adapted to provide sensitivity to pulse shape by the use of variable-period charge integration. We use a light-emitting diode to calibrate a combination of an ON Semiconductor (SensL) 60035-64P J-Series SiPM and TOPFET2 DAQ and evaluate the linearity of their response and dynamic range. The calibration curve was obtained by comparing the DAQ response to that measured with a photodiode interfaced with a traditional waveform digitizer. This calibration process was used for rudimentary spectroscopy of various neutron and gamma sources.
引用
收藏
页数:7
相关论文
共 34 条
  • [1] PCA-based neutron/gamma discrimination with organic scintillators
    Zhou, Hongzhao
    Xiao, Wuyun
    Liu, Haixia
    Sun, Tao
    Li, Zhiyuan
    Wang, Dongxi
    RADIATION PHYSICS AND CHEMISTRY, 2023, 212
  • [2] Evaluation of a digital data acquisition system and optimization of n-γ discrimination for a compact neutron spectrometer
    Giacomelli, L.
    Zimbal, A.
    Reginatto, M.
    Tittelmeier, K.
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (01):
  • [3] An ASIC-based system-in-package MEMS gas sensor with impedance spectroscopy readout and AI-enabled identification capabilities
    Zampolli, S.
    Elmi, I.
    Bruschi, P.
    Ria, A.
    Magliocca, F.
    Vitelli, M.
    Piotto, M.
    SENSORS AND ACTUATORS B-CHEMICAL, 2025, 424
  • [4] Portable data acquisition system based on compact flash interface
    Yuan, Yang
    Yong, Gao
    Yaowu, Gao
    ICEMI 2007: PROCEEDINGS OF 2007 8TH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS, VOL IV, 2007, : 942 - 946
  • [5] Compact spectroscopy system based on tunable organic semiconductor lasers
    Woggon, T.
    Klinkhammer, S.
    Lemmer, U.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2010, 99 (1-2): : 47 - 51
  • [6] Compact spectroscopy system based on tunable organic semiconductor lasers
    T. Woggon
    S. Klinkhammer
    U. Lemmer
    Applied Physics B, 2010, 99 : 47 - 51
  • [7] A MICROCOMPUTER-BASED SLEEP SYSTEM - DATA ACQUISITION AND SYSTEM CALIBRATION PROGRAMS
    VIVALDI, EA
    PASTEL, RH
    BAKALIAN, MJ
    FERNSTROM, JD
    BRAIN RESEARCH BULLETIN, 1988, 20 (01) : 133 - 138
  • [8] A USB-based data acquisition system for neutron TOF measurements
    Khalil, MI
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2006, 17 (01) : N1 - N7
  • [9] MPA-SSA, design and test of a 65 nm ASIC-based system for particle tracking at HL-LHC featuring on-chip particle discrimination
    Ceresa, Davide
    Bergamin, Gianmario
    Caratelli, Alessandro
    Kaplon, Jan
    Kloukinas, Kostas
    Scarfi, Simone
    Leblebici, Yusuf
    2019 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2019,
  • [10] COMPUTER-BASED CALIBRATION AND MEASUREMENT OF AN EEG DATA ACQUISITION-SYSTEM
    LENZ, JE
    KELLY, EF
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1981, 28 (05) : 396 - 402