Fast and Sensitive Method for Simultaneous Quantification of Meropenem and Vaborbactam in Human Plasma Microsamples by Liquid Chromatography-Tandem Mass Spectrometry for Therapeutic Drug Monitoring

被引:4
|
作者
Barone, Rossella [1 ,2 ]
Conti, Matteo [1 ]
Giorgi, Beatrice [1 ]
Gatti, Milo [1 ,2 ]
Cojutti, Pier Giorgio [1 ,2 ]
Viale, Pierluigi [2 ,3 ]
Pea, Federico [1 ,2 ]
机构
[1] IRCCS Azienda Osped Univ Bologna, Clin Pharmacol Unit, I-40138 Bologna, Italy
[2] Univ Bologna, Dept Med & Surg Sci, I-40138 Bologna, Italy
[3] IRCCS Azienda Osped Univ Bologna, Infect Dis Unit, I-40138 Bologna, Italy
来源
ANTIBIOTICS-BASEL | 2023年 / 12卷 / 04期
关键词
meropenem; vaborbactam; therapeutic drug monitoring; plasma micro samples; liquid chromatography-tandem mass spectrometry; BETA-LACTAM ANTIBIOTICS; CIPROFLOXACIN; PIPERACILLIN; CEFEPIME;
D O I
10.3390/antibiotics12040719
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
Meropenem (MRP)-Vaborbactam (VBR) is a novel beta-lactam/beta-lactamase inhibitor used for the management of difficult-to-treat Gram-negative infections. Among critically ill patients, MRP-VBR shows remarkable inter-individual variability in pharmacokinetic behavior, thus justifying the implementation of therapeutic drug monitoring (TDM) for improving real-time management in different challenging scenarios. In this study, we developed and validated a fast and sensitive Liquid Chromatography-Tandem Mass Spectrometry (LC-MS/MS) method for the simultaneous quantification of MRP and VBR in human plasma microsamples of 3 microliters. The analysis required only a single-step sample preparation and was performed by means of a fast chromatographic run of 4 min, followed by positive electrospray ionization and detection on a high-sensitivity triple quadrupole tandem mass spectrometer operated in multiple reaction monitoring modes. The straightforward analytical procedure was successfully validated, based on the EMA guidelines, in terms of specificity, sensitivity, linearity, precision, accuracy, matrix effect, extraction recovery, the limit of quantification, and stability. The novel method was successfully applied for simultaneously measuring MRP and VBR concentrations in more than 42 plasma samples collected from critically ill patients affected by carbapenem-resistant Gram-negative bacteria infections.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Quantification of rabeprazole in human plasma by liquid chromatography-tandem mass spectrometry
    Zhang, Y
    Chen, XY
    Gu, Q
    Zhong, DF
    ANALYTICA CHIMICA ACTA, 2004, 523 (02) : 171 - 175
  • [22] Rapid liquid chromatography-tandem mass spectrometry method for quantification of ziprasidone in human plasma
    Al-Dirbashi, OY
    Aboul-Enein, HY
    Al-Odaib, A
    Jacob, M
    Rashed, MS
    BIOMEDICAL CHROMATOGRAPHY, 2006, 20 (04) : 365 - 368
  • [23] Quantification of micafungin in human plasma by liquid chromatography-tandem mass spectrometry
    Giuliana Cangemi
    Sebastiano Barco
    Roberto Bandettini
    Elio Castagnola
    Analytical and Bioanalytical Chemistry, 2014, 406 : 1795 - 1798
  • [24] A validated method for the quantification of fosfomycin in human plasma by liquid chromatography-tandem mass spectrometry
    Martens-Lobenhoffer, Jens
    Bode-Boeger, Stefanie M.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2015, 990 : 164 - 168
  • [25] Quantification of micafungin in human plasma by liquid chromatography-tandem mass spectrometry
    Cangemi, Giuliana
    Barco, Sebastiano
    Bandettini, Roberto
    Castagnola, Elio
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2014, 406 (06) : 1795 - 1798
  • [26] Liquid chromatography-tandem mass spectrometry method for simultaneous quantification of four triazole antifungal agents in human plasma
    Verdier, Marie-Clemence
    Bentue-Ferrer, Daniele
    Tribut, Olivier
    Bellissant, Eric
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2010, 48 (10) : 1515 - 1522
  • [27] Rapid and Sensitive Liquid Chromatography-Tandem Mass Spectrometry for Quantification of Glycyrrhetic Acid in Human Plasma
    TIAN Li1
    3. Center for Drug Metabolism and Pharmacokinetic Research
    ChemicalResearchinChineseUniversities, 2011, 27 (02) : 185 - 189
  • [28] Rapid and Sensitive Liquid Chromatography-Tandem Mass Spectrometry for Quantification of Glycyrrhetic Acid in Human Plasma
    Tian Li
    Gao Xiao-li
    Chen Xiao-yan
    Zhong Da-fang
    Zhang Yi-fan
    Da Xiao-jian
    CHEMICAL RESEARCH IN CHINESE UNIVERSITIES, 2011, 27 (02) : 185 - 189
  • [29] Sensitive quantification of diphemanil methyl sulphate in human plasma by liquid chromatography-tandem mass spectrometry
    Do, B.
    Goulay-Dufay, S.
    Le Hoang, M. D.
    Adoui, N.
    Graffard, H.
    Guyon, F.
    Pradeau, D.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 845 (01): : 104 - 108
  • [30] Highly sensitive liquid chromatography-tandem mass spectrometry method for quantification of TAK-448 in human plasma
    Kuze, Yoji
    Jinno, Fumihiro
    Kondo, Takahiro
    Asahi, Satoru
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2013, 83 : 75 - 81