A Study of Fixed Current and Automatic Tube Current Modulation at a Fixed Low Tube Voltage in Low-dose Computed Tomography (CT) Imaging of Infant Lungs

被引:0
|
作者
Chen, Wushuang [1 ]
Tang, Shilong [1 ]
Xu, Zhen [1 ]
Chen, Xin [1 ]
Zheng, Yinglan [1 ]
He, Ling [1 ]
机构
[1] Chongqing Med Univ, Childrens Hosp, Dept Radiol, Chongqing, Peoples R China
关键词
Multilayer Spiral CT; Infant Chest; Low Dose; iDose4 Iterative Reconstruction; STATISTICAL ITERATIVE RECONSTRUCTION; QUALITY; CHEST;
D O I
10.5812/iranjradiol-126362
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: With the development of computed tomography (CT) technology, the number of infants undergoing CT examination has increased. Recently, problems associated with radiation exposure have attracted the public attention, and low-dose CT examination has become the research focus. Objectives: This study aimed to explore the feasibility of a fixed ultralow tube current combined with iDose(4) iterative reconstruction at a low tube voltage (80 kV) in low-dose CT scanning of infant lungs. Patients and Methods: Ninety infants, aged 0 - 3 years, undergoing enhanced chest CT scan (including plain CT scan and arterial scan), were included in the experimental group. In the experimental group, plain CT scan was performed at 80 kV under a fixed tube current (30 mA), and arterial phase CT scan was carried out at 80 kV with automatic tube current modulation (30 - 150 mA). Meanwhile, 90 infants, aged 0 - 3 years, undergoing non-enhanced chest CT scan, were randomly recruited as the control group. For the control group, plain CT scan was performed at 80 kV with automatic tube current modulation (30-150 mA). In each examination, the iDose4 iterative reconstruction technique was used for reconstruction. The volume CT dose index (CTDIvol), dose-length product (DLP), effective dose (ED, measured using a 32-cm diameter phantom), signal-to-noise ratio (SNR), and overall subjective quality of plain CT scans were analyzed and compared between the experimental and control groups. Results: Regarding the radiation dose in different age groups, the ED (mSv) was significantly lower in the experimental group compared to the control group (0-1-year-old subgroup: 0.48 +/- 0.05 mSv in the experimental group vs. 0.91 +/- 0.19 mSv in the control group, P < 0.05; 1-2-year-old subgroup: 0.50 +/- 0.01 mSv in the experimental group vs. 0.75 +/- 0.10 mSv in the control group, P < 0.05; and 2-3-year-old subgroup: 0.46 +/- 0.05 mSv in the experimental group vs. 0.79 +/- 0.11 mSv in the control group, P < 0.05). The image SNR was higher in the control group compared to the experimental group (P < 0.05), while the difference in subjective image quality was not significant between the experimental and control groups (P < 0.05). Conclusion: The chest CT examination of infants, aged 0 - 3 years, at a low tube voltage (80 kV) and a fixed tube current (30 mA), combined with the iDose4 iterative reconstruction technique, and the obtained images met the diagnostic criteria...
引用
收藏
页数:10
相关论文
共 50 条
  • [1] Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography A Phantom Study on Radiation Dose and Image Quality
    Papadalds, Antonios E.
    Damilakis, John
    INVESTIGATIVE RADIOLOGY, 2019, 54 (05) : 265 - 272
  • [2] Determining Tube Current, Tube Voltage and Pitch Suitable for Low-Dose Lung Screening CT
    Williams, K.
    Matthews, K.
    MEDICAL PHYSICS, 2014, 41 (06)
  • [3] Comparison of Image Quality and Radiation Dose between Fixed Tube Current and Combined Automatic Tube Current Modulation in Craniocervical CT Angiography
    Lee, E. J.
    Lee, S. K.
    Agid, R.
    Howard, P.
    Bae, J. M.
    terBrugge, K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (09) : 1754 - 1759
  • [4] Effects of tube potential selection together with computed tomography automatic tube current modulation on CT imaging performance
    Sookpeng, S.
    Martin, C. J.
    Krisanachinda, A.
    JOURNAL OF RADIOLOGICAL PROTECTION, 2021, 41 (04) : 809 - 831
  • [5] Low-Dose CT With Automatic Tube Current Modulation, Adaptive Statistical Iterative Reconstruction, and Low Tube Voltage for the Diagnosis of Renal Colic: Impact of Body Mass Index
    Gervaise, Alban
    Naulet, Pierre
    Beuret, Florence
    Henry, Christelle
    Pernin, Matthieu
    Portron, Yann
    Lapierre-Combes, Marie
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2014, 202 (03) : 553 - 560
  • [6] Effect of Automatic Tube Current Modulation on Radiation Dose and Image Quality for Low Tube Voltage Multidetector Row CT Angiography: Phantom Study
    Schindera, Sebastian T.
    Nelson, Rendon C.
    Yoshizumi, Terry
    Toncheva, Greta
    Nguyen, Giao
    DeLong, David M.
    Szucs-Farkas, Zsolt
    ACADEMIC RADIOLOGY, 2009, 16 (08) : 997 - 1002
  • [7] Comparison of image quality and radiation dose between combined automatic tube current modulation and fixed tube current technique in CT of abdomen and pelvis
    Lee, Sanghee
    Yoon, Sang-Wook
    Yoo, Seung-Min
    Ji, Young Geon
    Kim, Kyoung Ah
    Kim, Sang Heum
    Lee, Jong Tae
    ACTA RADIOLOGICA, 2011, 52 (10) : 1101 - 1106
  • [8] Combining Automatic Tube Current Modulation with Adaptive Statistical Iterative Reconstruction for Low-Dose Chest CT Screening
    Chen, Jiang-Hong
    Jin, Er-Hu
    He, Wen
    Zhao, Li-Qin
    PLOS ONE, 2014, 9 (04):
  • [9] Comparison of z-axis automatic tube current modulation technique with fixed tube current CT scanning of abdomen and pelvis
    Kalra, MK
    Maher, MM
    Toth, TL
    Kamath, RS
    Halpern, EF
    Saini, S
    RADIOLOGY, 2004, 232 (02) : 347 - 353
  • [10] Considerations on an automatic computed tomography tube current modulation system
    Moro, Luca
    Panizza, Denis
    D'Ambrosio, Daniela
    Carne, Irene
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (04) : 525 - 530