Immunoassay System Based on the Technology of Time-Resolved Fluorescence Resonance Energy Transfer

被引:0
|
作者
Xu, Zhengping [1 ]
Zhou, Hong [1 ]
Li, Li [1 ]
Chen, Zhang [1 ]
Zhang, Xin [1 ]
Feng, Yongtong [1 ]
Wang, Jianping [1 ]
Li, Yuan [2 ]
Wu, Yanfan [1 ]
机构
[1] Chinese Acad Sci, Suzhou Inst Biomed Engn & Technol, Suzhou 215163, Peoples R China
[2] Chongqing Guoke Med Innovat Technol Dev Co Ltd, Chongqing 401122, Peoples R China
关键词
time resolved; fluorescence resonance energy transfer; immunoassay; in vitro diagnosis; PROTEIN; INFLAMMATION; MARKERS;
D O I
10.3390/s24051430
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To enhance the specificity and sensitivity, cut the cost, and realize joint detection of multiple indicators, an immunoassay system based on the technology of time-resolved fluorescence resonance energy transfer (TR-FRET) was studied. Due to the FRET of the reagent, the donor probe and acceptor probe emitted specific fluorescence to enhance specificity. Long-lifetime specific fluorescence from the acceptor probe was combined with time-resolved technology to enhance sensitivity. A xenon flash lamp and a photomultiplier tube (PMT) were selected as the light source and detector, respectively. A filter-switching mechanism was placed in the light path, so the fluorescence signal from the donor and acceptor was measured alternately. The instrument's design is given, and some specificI parts are described in detail. Key technical specifications of the instrument and procalcitonin (PCT), C-reactive protein (CRP), and interleukin-6(IL-6) were tested, and the test results were presented subsequently. The CV value of the self-designed counting module is better than 0.01%, and the instrument noises for 620 nm and 665 nm are 41.44 and 10.59, respectively. When set at 37 degrees C, the temperature bias (B) is 0.06 degrees C, and the temperature fluctuation is 0.10 degrees C. The CV and bias are between +/- 3% and 5%, respectively, when pipetting volumes are between 10 mu L and 100 mu L. Within the concentration range of 0.01 nM to 10 nM, the luminescence values exhibit linear regression correlation coefficients greater than 0.999. For PCT detection, when the concentration ranges from 0.02 ng/mL to 50 ng/mL, the correlation coefficient of linear fitting exceeds 0.999, and the limit of quantification is 0.096 ng/mL. For CRP and IL-6, the detection concentration ranges from 0 ng/mL to 500 ng/mL and 0 ng/mL to 20 ng/mL, respectively, with limits of quantification of 2.70 ng/mL and 2.82 ng/mL, respectively. The experimental results confirm the feasibility of the technical and instrumental solutions.
引用
收藏
页数:15
相关论文
共 50 条
  • [21] Development of a time-resolved fluorescence resonance energy transfer (TR-FRET) assay for IKKε
    Gao, Yi
    Schartner, Caitlin
    Riddle, Steve
    Vogel, Kurt
    [J]. MOLECULAR CANCER THERAPEUTICS, 2007, 6 (12) : 3406S - 3407S
  • [22] Time-resolved spectroscopy and fluorescence resonance energy transfer in the study of excimer laser damage of chromatin
    Radu, L.
    Mihailescu, I.
    Radu, S.
    Gazdaru, D.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01): : 591 - 593
  • [23] Determination of Dimethachlon in Tobacco by Time-resolved Fluorescence Immunoassay
    Liu Yuan
    Liu Rong-Rong
    Liu Xian-Jin
    Fang Dun-Huang
    Song Cun-Man
    Huovinen, Tuomas
    Vehnianen, Markus
    Lovgren, Timo
    [J]. CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2012, 40 (07) : 1114 - 1117
  • [24] Time-resolved fluorescence immunoassay: An experiment for instrumental analysis
    Ritzert, Nicole L.
    Zapien, Donald C.
    [J]. ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2007, 233 : 612 - 612
  • [25] A time-resolved fluorescence immunoassay for insulin in rodent plasma
    Daijo, JE
    Sportsman, JR
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 19 (3-4) : 335 - 342
  • [26] SENSITIVE TIME-RESOLVED FLUORESCENCE IMMUNOASSAY OF SOMATOTROPIN IN SERUM
    KAHAN, I
    PAPANASTASIOUDIAMANDI, A
    ELLIS, G
    MAKELA, SK
    MCLAURIN, JA
    DCOSTA, M
    DIAMANDIS, EP
    [J]. CLINICAL CHEMISTRY, 1990, 36 (03) : 503 - 508
  • [27] A substrate for deubiquitinating enzymes based on time-resolved fluorescence resonance energy transfer between terbium and yellow fluorescent protein
    Horton, Robert A.
    Strachan, Elizabeth A.
    Vogel, Kurt W.
    Riddle, Steven M.
    [J]. ANALYTICAL BIOCHEMISTRY, 2007, 360 (01) : 138 - 143
  • [28] Homogeneous assay based on anti-Stokes' shift time-resolved fluorescence resonance energy-transfer measurement
    Laitala, V
    Hemmila, L
    [J]. ANALYTICAL CHEMISTRY, 2005, 77 (05) : 1483 - 1487
  • [29] THEORETICAL CONSIDERATIONS ON THE FLUORESCENCE LIFETIME OF LABELS FOR TIME-RESOLVED FLUORESCENCE IMMUNOASSAY
    BAILEY, MP
    ROCKS, BF
    RILEY, C
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1987, 5 (07) : 741 - 746
  • [30] Full Genotyping of a Highly Polymorphic Human Gene Trait by Time-Resolved Fluorescence Resonance Energy Transfer
    Tote, Edoardo
    Lamperti, Marco
    Bondani, Maria
    Salerno, Domenico
    Cassina, Valeria
    Nardo, Luca
    [J]. PLOS ONE, 2014, 9 (09):