Immunoassay System Based on the Technology of Time-Resolved Fluorescence Resonance Energy Transfer

被引:0
|
作者
Xu, Zhengping [1 ]
Zhou, Hong [1 ]
Li, Li [1 ]
Chen, Zhang [1 ]
Zhang, Xin [1 ]
Feng, Yongtong [1 ]
Wang, Jianping [1 ]
Li, Yuan [2 ]
Wu, Yanfan [1 ]
机构
[1] Chinese Acad Sci, Suzhou Inst Biomed Engn & Technol, Suzhou 215163, Peoples R China
[2] Chongqing Guoke Med Innovat Technol Dev Co Ltd, Chongqing 401122, Peoples R China
关键词
time resolved; fluorescence resonance energy transfer; immunoassay; in vitro diagnosis; PROTEIN; INFLAMMATION; MARKERS;
D O I
10.3390/s24051430
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
To enhance the specificity and sensitivity, cut the cost, and realize joint detection of multiple indicators, an immunoassay system based on the technology of time-resolved fluorescence resonance energy transfer (TR-FRET) was studied. Due to the FRET of the reagent, the donor probe and acceptor probe emitted specific fluorescence to enhance specificity. Long-lifetime specific fluorescence from the acceptor probe was combined with time-resolved technology to enhance sensitivity. A xenon flash lamp and a photomultiplier tube (PMT) were selected as the light source and detector, respectively. A filter-switching mechanism was placed in the light path, so the fluorescence signal from the donor and acceptor was measured alternately. The instrument's design is given, and some specificI parts are described in detail. Key technical specifications of the instrument and procalcitonin (PCT), C-reactive protein (CRP), and interleukin-6(IL-6) were tested, and the test results were presented subsequently. The CV value of the self-designed counting module is better than 0.01%, and the instrument noises for 620 nm and 665 nm are 41.44 and 10.59, respectively. When set at 37 degrees C, the temperature bias (B) is 0.06 degrees C, and the temperature fluctuation is 0.10 degrees C. The CV and bias are between +/- 3% and 5%, respectively, when pipetting volumes are between 10 mu L and 100 mu L. Within the concentration range of 0.01 nM to 10 nM, the luminescence values exhibit linear regression correlation coefficients greater than 0.999. For PCT detection, when the concentration ranges from 0.02 ng/mL to 50 ng/mL, the correlation coefficient of linear fitting exceeds 0.999, and the limit of quantification is 0.096 ng/mL. For CRP and IL-6, the detection concentration ranges from 0 ng/mL to 500 ng/mL and 0 ng/mL to 20 ng/mL, respectively, with limits of quantification of 2.70 ng/mL and 2.82 ng/mL, respectively. The experimental results confirm the feasibility of the technical and instrumental solutions.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Immunoassay for serodiagnosis of Zika virus infection based on time-resolved Forster resonance energy transfer
    Kareinen, Lauri
    Hepojoki, Satu
    Huhtamo, Eili
    Korhonen, Essi M.
    Schmidt-Chanasit, Jonas
    Hedman, Klaus
    Hepojoki, Jussi
    Vapalahti, Olli
    [J]. PLOS ONE, 2019, 14 (07):
  • [2] The fluorescence resonance energy transfer (FRET) gate: A time-resolved study
    Xu, QH
    Wang, S
    Korystov, D
    Mikhailovsky, A
    Bazan, GC
    Moses, D
    Heeger, AJ
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2005, 102 (03) : 530 - 535
  • [3] TIME-RESOLVED FLUORESCENCE ENERGY-TRANSFER
    VANDERMEER, BW
    CHEN, SY
    [J]. BIOPHYSICAL JOURNAL, 1993, 64 (02) : A242 - A242
  • [4] On the possibility of ephedrine detection: time-resolved fluorescence resonance energy transfer (FRET)-based approach
    Varriale, Antonio
    Marzullo, Vincenzo Manuel
    Di Giovanni, Stefano
    Scala, Andrea
    Capo, Alessandro
    Majoli, Adelia
    Pennacchio, Angela
    Staiano, Maria
    D'Auria, Sabato
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2016, 408 (23) : 6329 - 6336
  • [5] On the possibility of ephedrine detection: time-resolved fluorescence resonance energy transfer (FRET)-based approach
    Antonio Varriale
    Vincenzo Manuel Marzullo
    Stefano Di Giovanni
    Andrea Scala
    Alessandro Capo
    Adelia Majoli
    Angela Pennacchio
    Maria Staiano
    Sabato D’Auria
    [J]. Analytical and Bioanalytical Chemistry, 2016, 408 : 6329 - 6336
  • [6] Cell surface detection of membrane protein interaction with homogeneous time-resolved fluorescence resonance energy transfer technology
    Maurel, D
    Kniazeff, J
    Mathis, G
    Trinquet, E
    Pin, JP
    Ansanay, H
    [J]. ANALYTICAL BIOCHEMISTRY, 2004, 329 (02) : 253 - 262
  • [7] Rapid Homogeneous Immunoassay Based on Time-Resolved Forster Resonance Energy Transfer for Serodiagnosis of Acute Hantavirus Infection
    Hepojoki, Satu
    Hepojoki, Jussi
    Hedman, Klaus
    Vapalahti, Olli
    Vaheri, Antti
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (02) : 636 - 640
  • [8] Determination of resonance energy transfer efficiency by global analysis of time-resolved fluorescence spectra
    Biskup, C
    Zimmer, T
    Benndorf, K
    [J]. BIOPHYSICAL JOURNAL, 2003, 84 (02) : 475A - 476A
  • [9] Conformational flexibility in DNA nanoconstructs: A time-resolved fluorescence resonance energy transfer study
    Sandin, Peter
    Lincoln, Per
    Albinsson, Bo
    [J]. JOURNAL OF PHYSICAL CHEMISTRY C, 2008, 112 (34): : 13089 - 13094
  • [10] Time-resolved fluorescence resonance energy transfer: A versatile tool for the analysis of nucleic acids
    Klostermeier, D
    Millar, DP
    [J]. BIOPOLYMERS, 2001, 61 (03) : 159 - 179