Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review

被引:10
|
作者
Aoki, Kawaiola C. [1 ]
Barrant, Maya [1 ]
Gai, Mam Jarra [1 ]
Handal, Marina [1 ]
Xu, Vivian [1 ]
Mayrovitz, Harvey N. [2 ]
机构
[1] Nova Southeastern Univ, Dr Kiran C Patel Coll Osteopath Med, Med Sch, Ft Lauderdale, FL 33314 USA
[2] Nova Southeastern Univ, Dr Kiran C Patel Coll Allopath Med, Med Educ & Simulat, Cardiopulm Physiol, Ft Lauderdale, FL 33328 USA
关键词
racial disparity; racial bias; skin color; occult hypoxia; electronic monitoring devices; false hypoxemia; skin pigmentation; pulse oximetry; PULSE OXIMETRY; ACCURACY; PIGMENTATION; DISPARITIES; HEMOGLOBIN; AGREEMENT; MELANIN;
D O I
10.7759/cureus.46078
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Standard pulse oximeters estimate arterial blood saturation (SaO(2)) non-invasively by emitting and detecting light of a specific wavelength through a cutaneous vascular bed, such as a digit or the ear lobe. The quantity measured at these peripheral sites is designated as oxygen saturation (SpO(2)). Most reliable pulse oximeters are calibrated from measurements of healthy volunteers using some form of oxygen desaturation method. As the degree of inducible hypoxemia is limited, the calibration below achievable desaturation levels is usually extrapolated, leading to potential measurement error at low SaO(2) values, especially in highly pigmented skin. Such skin color-related errors (SCRE) are the topic of this scoping review. Specifically, this study aimed to identify the combined impact of skin color and reduced SaO2 on the non-invasive assessment of SpO(2) and report the consequences of potential inaccuracies. Three databases were searched (Cumulated Index to Nursing and Allied Health Literature (CINAHL), PubMed, and Web of Science) for peer-reviewed prospective and retrospective studies published in English between 2000 and 2022 involving human patients with hypoxemia that included a measure of skin color (Fitzpatrick scale or race/ethnicity). Ten studies met the criteria and were included in the final review. Eight of these studies reported statistically significant higher pulse oximeter readings in darker-skinned patients with hypoxia compared to their arterial blood gas measurements. Occult hypoxia was more prevalent in Black and Hispanic patients than in White patients. Minority patients overall (Black, Asian, and American Indian) were more likely to have a SaO(2) < 88% that was not detected by pulse oximetry (occult hypoxemia) during hospitalization. With greater levels of hypoxemia, the differences between SpO(2) and SaO(2) were greater. If SaO(2) was < 90%, then SpO(2) was overestimated in all ethnicities but worse in minorities. In conclusion, the bias found in pulse oximeter readings in the skin of color broadly impacts patients with hypoxemia. The failure of SpO(2) measuring devices to detect occult hypoxemia can delay the delivery of lifesaving treatment to critically ill patients requiring respiratory rehabilitation and supplemental oxygen therapy. This may lead to adverse health outcomes, increased in-hospital mortality, and complications such as organ dysfunction. An improvement in pulse oximeter detection mechanisms that would include all skin pigmentations is therefore much desired to optimize individual healthcare status and minimize disparities in treatment.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Assessment of blood oxygen saturation using photoacoustic technique
    Saha R.K.
    Karmakar S.
    Roy M.
    Saha, R.K. (ratank.saha@saha.ac.in), 2013, Optical Society of India (42): : 141 - 147
  • [22] NONINVASIVE MEASUREMENT OF BLOOD-OXYGEN SATURATION WITH MAGNETIC-RESONANCE-IMAGING
    PELC, LR
    LI, KC
    WRIGHT, GA
    WEGMUELLER, H
    DALMAN, R
    BRITTAIN, J
    CIRCULATION, 1993, 88 (04) : 237 - 237
  • [23] Technique for noninvasive monitoring of blood loss via oxygen saturation measurements in the eye
    Smith, MH
    Denninghoff, KR
    Hillman, LW
    Hughes, CE
    Minnich, TE
    Chipman, RA
    OPTICAL DIAGNOSTICS OF BIOLOGICAL FLUIDS AND ADVANCED TECHNIQUES IN ANALYTICAL CYTOLOGY, PROCEEDINGS OF, 1997, 2982 : 46 - 52
  • [24] Application of Optical Analysis Methods for Noninvasive Monitoring of the Blood Oxygen Saturation Level
    Guzenko M.M.
    Mazing M.S.
    Zaitseva A.Y.
    Biophysics, 2023, 68 (2) : 306 - 311
  • [25] AI-Based Noninvasive Blood Glucose Monitoring: Scoping Review
    Chan, Pin Zhong
    Jin, Eric
    Jansson, Miia
    Chew, Han Shi Jocelyn
    JOURNAL OF MEDICAL INTERNET RESEARCH, 2024, 26
  • [26] Saturation Oxygen Pressure Index for Assessment of Pulmonary Disease in Neonates on Noninvasive Ventilation
    Doreswamy, Srinivasa M.
    Chakkarapani, Aravanan A.
    Murthy, Prashanth
    INDIAN PEDIATRICS, 2015, 52 (01) : 74 - 75
  • [27] A Review of Monitoring Methods for Cerebral Blood Oxygen Saturation
    Zhong, Wentao
    Ji, Zhong
    Sun, Changlong
    HEALTHCARE, 2021, 9 (09)
  • [28] Systematic instrumental errors between oxygen saturation analysers in fetal blood during deep hypoxemia
    Porath, M
    Sinha, P
    Dudenhausen, JW
    Luttkus, AK
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 151 - 157
  • [29] SKIN OXYGEN TENSION AND ARTERIAL OXYGEN SATURATION INVIVO - SIMILARITY TO THE OXYGEN DISSOCIATION CURVE OF THE BLOOD
    PENNEYS, R
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 1951, 221 (01): : 115 - 115
  • [30] Assessment and Identification of Skin Disorders in Skin of Color An Integrative Review