Impacts of Skin Color and Hypoxemia on Noninvasive Assessment of Peripheral Blood Oxygen Saturation: A Scoping Review

被引:10
|
作者
Aoki, Kawaiola C. [1 ]
Barrant, Maya [1 ]
Gai, Mam Jarra [1 ]
Handal, Marina [1 ]
Xu, Vivian [1 ]
Mayrovitz, Harvey N. [2 ]
机构
[1] Nova Southeastern Univ, Dr Kiran C Patel Coll Osteopath Med, Med Sch, Ft Lauderdale, FL 33314 USA
[2] Nova Southeastern Univ, Dr Kiran C Patel Coll Allopath Med, Med Educ & Simulat, Cardiopulm Physiol, Ft Lauderdale, FL 33328 USA
关键词
racial disparity; racial bias; skin color; occult hypoxia; electronic monitoring devices; false hypoxemia; skin pigmentation; pulse oximetry; PULSE OXIMETRY; ACCURACY; PIGMENTATION; DISPARITIES; HEMOGLOBIN; AGREEMENT; MELANIN;
D O I
10.7759/cureus.46078
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Standard pulse oximeters estimate arterial blood saturation (SaO(2)) non-invasively by emitting and detecting light of a specific wavelength through a cutaneous vascular bed, such as a digit or the ear lobe. The quantity measured at these peripheral sites is designated as oxygen saturation (SpO(2)). Most reliable pulse oximeters are calibrated from measurements of healthy volunteers using some form of oxygen desaturation method. As the degree of inducible hypoxemia is limited, the calibration below achievable desaturation levels is usually extrapolated, leading to potential measurement error at low SaO(2) values, especially in highly pigmented skin. Such skin color-related errors (SCRE) are the topic of this scoping review. Specifically, this study aimed to identify the combined impact of skin color and reduced SaO2 on the non-invasive assessment of SpO(2) and report the consequences of potential inaccuracies. Three databases were searched (Cumulated Index to Nursing and Allied Health Literature (CINAHL), PubMed, and Web of Science) for peer-reviewed prospective and retrospective studies published in English between 2000 and 2022 involving human patients with hypoxemia that included a measure of skin color (Fitzpatrick scale or race/ethnicity). Ten studies met the criteria and were included in the final review. Eight of these studies reported statistically significant higher pulse oximeter readings in darker-skinned patients with hypoxia compared to their arterial blood gas measurements. Occult hypoxia was more prevalent in Black and Hispanic patients than in White patients. Minority patients overall (Black, Asian, and American Indian) were more likely to have a SaO(2) < 88% that was not detected by pulse oximetry (occult hypoxemia) during hospitalization. With greater levels of hypoxemia, the differences between SpO(2) and SaO(2) were greater. If SaO(2) was < 90%, then SpO(2) was overestimated in all ethnicities but worse in minorities. In conclusion, the bias found in pulse oximeter readings in the skin of color broadly impacts patients with hypoxemia. The failure of SpO(2) measuring devices to detect occult hypoxemia can delay the delivery of lifesaving treatment to critically ill patients requiring respiratory rehabilitation and supplemental oxygen therapy. This may lead to adverse health outcomes, increased in-hospital mortality, and complications such as organ dysfunction. An improvement in pulse oximeter detection mechanisms that would include all skin pigmentations is therefore much desired to optimize individual healthcare status and minimize disparities in treatment.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Peripheral Oxygen Saturation Misclassifies Hypoxemia
    Matsson, Victor Sjodahl
    Ekstrom, Magnus
    ANNALS OF THE AMERICAN THORACIC SOCIETY, 2023, 20 (07) : 1070 - 1073
  • [2] NONINVASIVE MONITORING OF ARTERIAL BLOOD-OXYGEN SATURATION
    MENDELSON, Y
    CHEUNG, PW
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1984, 31 (08) : 581 - 581
  • [3] Noninvasive Monitoring of Temporal Variation in Transcutaneous Oxygen Saturation for Clinical Assessment of Skin Microcirculatory Activity
    Huong, A. K. C.
    Philimon, S. P.
    Ngu, X. T. I.
    INTERNATIONAL CONFERENCE FOR INNOVATION IN BIOMEDICAL ENGINEERING AND LIFE SCIENCES, ICIBEL2015, 2016, 56 : 248 - 251
  • [4] NONINVASIVE METHODS OF MEASURING OXYGEN SATURATION IN VENOUS BLOOD
    Yesman, S. S.
    Mamilov, S. A.
    Asimov, M. M.
    Gisbreht, A. I.
    JOURNAL OF APPLIED SPECTROSCOPY, 2011, 78 (03) : 406 - 413
  • [5] Noninvasive methods of measuring oxygen saturation in venous blood
    S. S. Yesman
    S. A. Mamilov
    M. M. Asimov
    A. I. Gisbreht
    Journal of Applied Spectroscopy, 2011, 78 : 406 - 413
  • [6] Noninvasive Glucose Evaluation by Human Skin Oxygen Saturation Level
    Wei, Tzu-Ting
    Tsai, Hsin-Yi
    Yang, Ching-Ching
    Hsiao, Wen-Tse
    Huang, Kuo-Cheng
    2016 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE PROCEEDINGS, 2016, : 1384 - 1388
  • [7] Assessment of Moderate to Severe Abdominal Blood Loss Using Peripheral to Central Blood Oxygen Saturation
    Khorasani, Zadeh S.
    Elyassi, H.
    Gharaei, B.
    Yavari, P.
    ADVANCES IN MEDICAL SCIENCES, 2008, 53 (01): : 87 - 93
  • [8] MIRTH AND OXYGEN SATURATION LEVELS OF PERIPHERAL BLOOD
    FRY, WF
    STOFT, PE
    PSYCHOTHERAPY AND PSYCHOSOMATICS, 1971, 19 (1-2) : 76 - &
  • [9] Hypoxia, color vision deficiencies, and blood oxygen saturation
    Hovis, Jeffery K.
    Milburn, Nelda J.
    Nesthus, Thomas E.
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2012, 29 (02) : A268 - A274
  • [10] Oxygen saturation of skin reflects blood flow and stagnation
    Yamamoto S.
    Tsumura N.
    Yoshizaki T.
    Ogawa-Ochiai K.
    Artificial Life and Robotics, 2014, 19 (2) : 170 - 175