Reliable in vitro method for the evaluation of the primary stability and load transfer of transfemoral prostheses for osseointegrated implantation

被引:1
|
作者
Galteri, Giulia [1 ]
Palanca, Marco [1 ]
Alesi, Domenico [2 ]
Zaffagnini, Stefano [2 ]
Morellato, Kavin [3 ]
Gruppioni, Emanuele [3 ]
Cristofolini, Luca [1 ]
机构
[1] Alma Mater Studiorum Univ Bologna, Sch Engn & Architecture, Dept Ind Engn, Bologna, Italy
[2] IRCCS Ist Ortoped Rizzoli, Bologna, Italy
[3] Ctr Protesi INAIL, Bologna, Italy
关键词
lower-limb amputation; above-the-knee amputation; transfemoral osseointegrated prosthesis; primary stability; load transfer; in vitro mechanical test; micromotion; BONE-ANCHORED PROSTHESES; QUALITY-OF-LIFE; FEMORAL AMPUTATION; DIRECT FIXATION; FOLLOW-UP; AMPUTEES; STRESS; ATTACHMENT; FAILURE; STEM;
D O I
10.3389/fbioe.2024.1360208
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Osseointegrated transfemoral prostheses experience aseptic complications with an incidence between 3% and 30%. The main aseptic risks are implant loosening, adverse bone remodeling, and post-operative periprosthetic fractures. Implant loosening can either be due to a lack of initial (primary) stability of the implant, which hinders bone ingrowth and therefore prevents secondary stability, or, in the long-term, to the progressive resorption of the periprosthetic bone. Post-operative periprosthetic fractures are most often caused by stress concentrations. A method to simultaneously evaluate the primary stability and the load transfer is currently missing. Furthermore, the measurement errors are seldom reported in the literature. In this study a method to reliably quantify the bone implant interaction of osseointegrated transfemoral prostheses in terms of primary stability and load transfer was developed, and its precision was quantified. Micromotions between the prosthesis and the host bone and the strains on the cortical bone were measured on five human cadaveric femurs with a typical commercial osseointegrated implant. To detect the primary stability of the implant and the load transfer, cyclic loads were applied, simulating the peak load during gait. Digital Image Correlation was used to measure displacements and bone strains simultaneously throughout the test. Permanent migrations and inducible micromotions were measured (three translations and three rotations), while, on the same specimen, the full-field strain distribution on the bone surface was measured. The repeatability tests showed that the devised method had an intra-specimen variability smaller than 6 mu m for the translation, 0.02 degrees for the rotations, and smaller than 60 microstrain for the strain distribution. The inter-specimen variability was larger than the intra-specimen variability due to the natural differences between femurs. Altogether, the measurement uncertainties (intrinsic measurement errors, intra-specimen repeatability and inter-specimen variability) were smaller than critical levels of biomarkers for adverse remodelling and aseptic loosening, thus allowing to discriminate between stable and unstable implants, and to detect critical strain magnitudes in the host bone. In conclusion, this work showed that it is possible to measure the primary stability and the load transfer of an osseointegrated transfemoral prosthesis in a reliable way using a combination of mechanical testing and DIC.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] A new method for evaluation of the load transfer capability of trunk power networks
    Takahashi, K
    Tanaka, K
    Kurihara, I
    Inoue, A
    ELECTRICAL ENGINEERING IN JAPAN, 1999, 128 (01) : 17 - 25
  • [22] New method for evaluation of the load transfer capability of trunk power networks
    Takahashi, Kazuhiro
    Tanaka, Kazuyuki
    Kurihara, Ikuo
    Inoue, Atuyuki
    Electrical Engineering in Japan (English translation of Denki Gakkai Ronbunshi), 1999, 128 (01): : 17 - 25
  • [23] An in vitro evaluation of fracture load of implant-supported zirconia-based prostheses fabricated with different veneer materials
    Takata, Hiroki
    Komine, Futoshi
    Honda, Junichi
    Blatz, Markus B.
    Matsumura, Hideo
    CLINICAL ORAL IMPLANTS RESEARCH, 2018, 29 (04) : 396 - 403
  • [24] In vitro evaluation of the influence of the cortical bone on the primary stability of two implant systems
    Andres-Garcia, Rodrigo
    Garcia Vives, Nuria
    Herrero Climent, Federico
    Fernandez Palacin, Ana
    Rios Santos, Vicente
    Herrero Climent, Mariano
    Bullon, Pedro
    MEDICINA ORAL PATOLOGIA ORAL Y CIRUGIA BUCAL, 2009, 14 (02): : E93 - E97
  • [25] Reliability and Correlation of Different Devices for the Evaluation of Primary Implant Stability: An In Vitro Study
    Raz, Perry
    Meir, Haya
    Levartovsky, Shifra
    Peleg, Maia
    Sebaoun, Alon
    Beitlitum, Ilan
    MATERIALS, 2021, 14 (19)
  • [26] An accurate and reliable operational transfer path analysis for transfer path contribution evaluation based on Landweber iterative method
    Cheng, Wei
    Zhu, Yan
    Lu, Jingbai
    Chen, Xuefeng
    Zhou, Guanghui
    Gao, Lin
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2020, 31 (02)
  • [27] Establishment of two complementary in vitro assays for radiocopper complexes achieving reliable and comparable evaluation of in vivo stability
    Zarschler, K.
    Kubeil, M.
    Stephan, H.
    RSC ADVANCES, 2014, 4 (20): : 10157 - 10164
  • [28] Electroporation of primary neural cultures: a simple method for directed gene transfer in vitro
    Kirsten D. Mertz
    Gunnar Weisheit
    Karl Schilling
    Georg H. Lüers
    Histochemistry and Cell Biology, 2002, 118 : 501 - 506
  • [29] Electroporation of primary neural cultures:: a simple method for directed gene transfer in vitro
    Mertz, KD
    Weisheit, G
    Schilling, K
    Lüers, GH
    HISTOCHEMISTRY AND CELL BIOLOGY, 2002, 118 (06) : 501 - 506
  • [30] Electroporation of primary neural cultures: a simple method for directed gene transfer in vitro
    Kirsten D. Mertz
    Gunnar Weisheit
    Karl Schilling
    Georg H. Lüers
    Histochemistry and Cell Biology, 2003, 119 : 175 - 175