Accuracy of Crit-Line® monitors in predicting hematocrit and change in blood volume of dogs during hemodialysis

被引:0
|
作者
Vickery, Whitney E. [1 ,3 ]
Steinbach, Sarah M. L. [1 ]
Moore, George E. [2 ]
机构
[1] Purdue Univ, Dept Vet Clin Sci, W Lafayette, IN USA
[2] Purdue Univ, Dept Vet Adm, W Lafayette, IN USA
[3] JBSA, 2899 Schofield Rd, Ft Sam Houston, TX 78234 USA
关键词
acute kidney injury; fluid overload; intradialytic morbid event; packed cell volume; ultrafiltration; AGREEMENT; BLAND;
D O I
10.1111/jvim.16917
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background: Ultrafiltration (UF) is an extracorporeal technique for treating fluid overload and is monitored with noninvasive blood volume monitors.Hypothesis/Objectives: The primary objective was to determine the accuracy of Crit-Line (R) III and IV noninvasive blood volume monitors to estimate canine packed cell volume (PCV) and changes in blood volume during UF. A secondary aim was to determine accuracy of targeted ultrafiltration rates (UFR) compared to calculated delivered UFR.Methods: An ex vivo study with a single Phoenix (R) X36 platform and canine packed red blood cells (pRBC) was performed. Through dilution and UF, clinically applicable PCV values were obtained and compared to Crit-Line (R) hematocrit (Hct). Blood volume was constantly measured and compared to targeted UFR. Systematic and proportional bias were calculated using a Bayes method.Results: Crit-Line (R) III and IV reported Hct was significantly lower than PCV (n = 140, median 26%, range, 8%-50%) when PCV was >25% and >30%, respectively. Crit-Line (R) III and IV calculated change in blood volume (Delta BV%) was significantly different from measured blood volume changes at Delta BV%+/- 20% and >-20%, respectively. Comparing targeted and delivered UFR (mL/h), less than targeted UFR was removed at UFR100 through UFR400 and UFR0 adding volume.Conclusions and Clinical Importance: Crit-Line (R) III and IV monitors provide accurate estimates of canine PCV and UF volume change within specific ranges and are useful in monitoring canine UF and hemodialysis. Veterinary hemodialysis services should evaluate individual machines for UFR inaccuracies, which can meaningfully affect small animals.
引用
收藏
页码:2137 / 2144
页数:8
相关论文
共 23 条
  • [21] Correlation between changes in blood volume during hemodialysis estimated by continuous hematocrit measurement and changes in inferior caval vein diameter.
    Krepel, HP
    vandenDorpel, MA
    Weimar, W
    Zietse, R
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1997, 8 : A1119 - A1119
  • [22] Deterioration of Hepatic Oxygenation Precedes an Onset of Intradialytic Hypotension with Little Change in Blood Volume during Hemodialysis
    Imai, Sojiro
    Ookawara, Susumu
    Ito, Kiyonori
    Kiryu, Satoshi
    Tabei, Kaoru
    Morishita, Yoshiyuki
    BLOOD PURIFICATION, 2018, 45 (04) : 345 - 346
  • [23] Diagnostic and therapeutic tools of in-line measurement of blood volume and oxygen saturation during acute and chronic hemodialysis.
    vanOlden, RW
    deBlok, K
    Groeneveld, JO
    Arisz, L
    Krediet, RT
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1996, 7 (09): : A1395 - A1395