A highly sensitive immunochromatographic assay for lead ions in drinking water based on antibody-oriented probe and silver enhancement

被引:2
|
作者
Li, Yangyang [1 ]
Zhu, Zhengwei [2 ]
Qu, Wenli [1 ]
Yang, Qing [3 ,4 ]
Liu, Yan [3 ,4 ,5 ]
Wang, Qiao [3 ,4 ]
Duan, Shuo [3 ,4 ]
Wu, Jine [3 ,4 ]
Gong, Zhiyong [3 ,4 ]
Xu, Lin [3 ,4 ]
机构
[1] Wuhan Polytech Univ, Coll Food Sci & Engn, Xuefunan Rd 68, Wuhan 430023, Peoples R China
[2] Hubei Prov Inst Food Supervis & Test, Wuhan 430075, Peoples R China
[3] Minist Educ, Key Lab Deep Proc Major Grain & Oil, Wuhan 430023, Peoples R China
[4] Hubei Key Lab Proc & Transformat Agr Prod, Wuhan 430023, Peoples R China
[5] Key Res Inst Humanities & Social Sci Hubei Prov, Food Safety Res Ctr, Wuhan 430023, Peoples R China
基金
中国国家自然科学基金;
关键词
Silver enhancement; Antibody-oriented probe; Immunochromatographic assay; Lead ions; LATERAL FLOW IMMUNOASSAY; STRIP ASSAY;
D O I
10.1007/s00217-023-04351-5
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
Lead ions are a toxic metal that can enter the human body through the diet and negatively impact human health. A wide variety of foods, such as vegetables, meat, aquatic product, and drinking water, can be contaminated with lead ions. In this study, a highly sensitive immunochromatographic assay (ICA) based on antibody-oriented probe and silver enhancement was established for the qualitative and quantitative determination of lead ions in drinking water. The recombinant protein A (PrA) was used in probe preparation to improve antibody bioactivity while silver enhancement method was applied to enhance the signal reporter of ICA. It was determined that the visual limit of detection and the scanning limit of quantification of the ICA based on antibody-oriented probe and silver enhancement for lead ions in drinking water were 0.2 ng/mL and 0.08 ng/mL, respectively. The results obtained from developed ICA strip and ICP-MS had no obvious difference using statistical analysis of t test. Compared to previously reported ICA methods for lead ions, the developed ICA method enabled analyzing lead ions in drinking water with a superior sensitivity and detection range.
引用
收藏
页码:3097 / 3103
页数:7
相关论文
共 23 条
  • [21] Development and Comparative Evaluation of Two Highly Sensitive Immunosensor Platforms for Trace Determination of Copper Ions in Drinking Water Using a Monoclonal Antibody Specific to Copper-EDTA Complex
    Darwish, Ibrahim A.
    Wang, Zongzhi
    Darling, Ryhan J.
    MOLECULES, 2023, 28 (20):
  • [22] Novel ultra-sensitive and highly selective cyanine sensors based on solvent-free microwave synthesis for the detection of trace hypochlorite ions in drinking water
    Hafez, Nermeen S.
    Amer, Wael A.
    Okba, Ehab A.
    Sakr, Mahmoud A. S.
    Alganzory, Hussein H.
    Ebeid, El-Zeiny M.
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2023, 303
  • [23] A highly sensitive, selective and renewable carbon paste electrode based on a unique acyclic diamide ionophore for the potentiometric determination of lead ions in polluted water samples
    Zayed, M. A.
    Mahmoud, Walaa H.
    Abbas, Ashraf A.
    Ali, Aya E.
    Mohamed, Gehad G.
    RSC ADVANCES, 2020, 10 (30) : 17552 - 17560