Noninvasive and Online Detection of Thrombus in Extracorporeal Circulation Based on Electrical Impedance Tomography

被引:0
|
作者
Zhang, Jia-Hui [1 ]
Gao, Gui-Feng [1 ,2 ]
Wu, Yang [1 ]
Zhou, Tong [3 ]
Liu, Kai [1 ]
Yao, Jia-Feng [1 ]
机构
[1] Nanjing Univ Aeronaut & Astronaut, Coll Mech & Elect Engn, Nanjing 210016, Peoples R China
[2] Amoul, Shenzhen 518100, Peoples R China
[3] SooChow Univ, Affiliated Hosp 2, Suzhou 215004, Peoples R China
基金
中国国家自然科学基金;
关键词
electrical impedance tomography; thrombus online detection; extracorporeal circulation; joint simulation; RECONSTRUCTION; SYSTEM; BLOOD;
D O I
10.16476/j.pibb.2022.0255
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
Objective The online detection of thrombus and visualization are of great importance in the extracorporeal circulatory system. In this paper, a noninvasive and online detection method of thrombus in extracorporeal circulation based on electrical impedance tomography (EIT) is proposed. Methods Firstly, this paper investigates the effect of sensor size on the imaging performance through joint simulation. Secondly, a 16 -copper electrode EIT sensor with a diameter of 20 mm is designed based on the simulation results, a cyclic flow experimental platform is built, and static and cyclic flow experiments are designed. Instead of thrombus, pig blood clots of 3-6 mm in size are placed in fresh pig blood samples and image using the Tikhonov regularization algorithm. The 3 mm and 5 mm clots are placed in the circulation system, and the images of the size and position of the clots in the sensor section are reconstructed by the EIT system and compared with the results taken by the high-speed camera. Results The simulation results show that the image correlation coefficients (IC) are greater than 0.06 for the sensor diameters of 20 mm and 30 mm when the target-to-sensor area ratio (AR) is not less than 0.01, and the imaging effect is well. The static imaging results show that the relative size coverage ratio is less than or equal to 0.1. The circulating flow experiments show that the normalized relative conductivity value changes of 80 and 200 are detected as the clots pass through the sensor, and the results show that the method is capable of detecting clots in the circulatory system. Conclusion The method has the advantages of real-time and noninvasive, and is expected to be applied to the detection of thrombus in vitro.
引用
收藏
页码:1480 / 1488
页数:9
相关论文
共 28 条
  • [11] Time Sequence Learning for Electrical Impedance Tomography Using Bayesian Spatiotemporal Priors
    Liu, Shengheng
    Cao, Ruisong
    Huang, Yongming
    Ouypornkochagorn, Taweechai
    Ji, Jiabin
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2020, 69 (09) : 6045 - 6057
  • [12] Thrombelastography/thromboelastometry
    Luddington, RJ
    [J]. CLINICAL AND LABORATORY HAEMATOLOGY, 2005, 27 (02): : 81 - 90
  • [13] Development of Optical Sensing System for Noninvasive and Dynamic Monitoring of Thrombogenic Process
    Oshima, Shiori
    Sankai, Yoshiyuki
    [J]. ASAIO JOURNAL, 2010, 56 (05) : 460 - 467
  • [14] Pop G., 2003, Applied Rheology, V13, P305
  • [15] HEMATOLOGICAL AND PLASMA BIOCHEMICAL VALUES FOR HEALTHY YUCATAN MICROPIGS
    RISPAT, G
    SLAOUI, M
    WEBER, D
    SALEMINK, P
    BERTHOUX, C
    SHRIVASTAVA, R
    [J]. LABORATORY ANIMALS, 1993, 27 (04) : 368 - 373
  • [16] Application of electrical resistance tomography for thrombus visualization in blood
    Sapkota, Achyut
    Fuse, Takuto
    Seki, Masatoshi
    Maruyama, Osamu
    Sugawara, Michiko
    Takei, Masahiro
    [J]. FLOW MEASUREMENT AND INSTRUMENTATION, 2015, 46 : 334 - 340
  • [17] Shi J H., 2019, CHINA METROLOGY, P4
  • [18] Tikhonov regularization and prior information in electrical impedance tomography
    Vauhkonen, M
    Vadasz, D
    Karjalainen, PA
    Somersalo, E
    Kaipio, JP
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (02) : 285 - 293
  • [19] Imbalances in regional lung ventilation - A validation study on electrical impedance tomography
    Victorino, JA
    Borges, JB
    Okamoto, VN
    Matos, GFJ
    Tucci, MR
    Caramez, MPR
    Tanaka, H
    Sipmann, FS
    Santos, DCB
    Barbas, CSV
    Carvalho, CRR
    Amato, MBP
    [J]. AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2004, 169 (07) : 791 - 800
  • [20] Image Reconstruction for Electrical Impedance Tomography Using Radial Basis Function Neural Network Based on Hybrid Particle Swarm Optimization Algorithm
    Wang, Huan
    Liu, Kai
    Wu, Yang
    Wang, Song
    Zhang, Zheng
    Li, Fang
    Yao, Jiafeng
    [J]. IEEE SENSORS JOURNAL, 2021, 21 (02) : 1926 - 1934