Synopsis of JBS recommendations for magnetic resonance imaging in patients with cardiac implantable electronic devices

被引:0
|
作者
Zghaib, Tarek [1 ]
Nazarian, Saman [1 ]
机构
[1] Univ Penn, Perelman Sch Med, Philadelphia, PA 19104 USA
基金
美国国家卫生研究院;
关键词
CONSENSUS STATEMENT;
D O I
10.1136/heartjnl-2022-321480
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Written consent is not required with MR conditional CIEDs. Prior to scan acquisition, the device is programmed to MR Mode which uses OVO/ODO or VOO/DOO pacing depending on the presence or absence of a stable underlying rhythm, respectively. This reduces the risks associated with over-sensing electromagnetic interference leading to asystole or inappropriate tracking of electromagnetic interference leading to unwarranted pacing at the upper track rate and/or the risk of malignant arrhythmia due to € R-on-T'. Antitachycardia therapies are turned off in defibrillators to prevent inappropriate therapy attempts and voltage depletion. Required monitoring modalities, qualified personnel and resuscitation equipment are listed in table 1. In most cases, scanning in Normal Operating Mode is recommended with standard specific absorption rates (SARs). All MR conditional devices allow using 1.5 T scans and most allow 3 T, but checking MR conditional parameters set by the manufacturers is recommended. Metal artefact reduction techniques may be used for thoracic imaging. If malignant arrhythmias are detected, scanning should be halted promptly, advanced cardiac life support delivered and the device reprogrammed to deliver appropriate therapies. After the scan, full device interrogation should be performed and MR Mode disabled, reverting the CIED to its original settings. Any significant change in device parameters should be communicated to the patient and cardiologist. © Author(s) (or their employer(s)) 2023. No commercial re-use. See rights and permissions. Published by BMJ.
引用
收藏
页码:225 / 227
页数:3
相关论文
共 50 条
  • [31] Remote programming of cardiac implantable electronic devices: A novel approach to program cardiac devices for magnetic resonance imaging
    Siddamsetti, Sisir
    Shinn, Alexander
    Gautam, Sandeep
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (05) : 1005 - 1009
  • [32] "Power-on resets" in cardiac implantable electronic devices during magnetic resonance imaging
    Higgins, John V.
    Sheldon, Seth H.
    Watson, Robert E., Jr.
    Dalzell, Connie
    Acker, Nancy
    Cha, Yong-Mei
    Asirvatham, Samuel J.
    Kapa, Suraj
    Felmlee, Joel P.
    Friedman, Paul A.
    HEART RHYTHM, 2015, 12 (03) : 540 - 544
  • [33] Clinical Performance of Magnetic Resonance Imaging Conditional and Nonconditional Cardiac Implantable Electronic Devices
    Shah, Anand D.
    Patel, Adarsh U.
    Knezevic, Andrea
    Hoskins, Michael H.
    Hirsh, David S.
    Merchant, Faisal M.
    El Chami, Mikhael F.
    Delurgio, David B.
    Patel, Anshul M.
    Leon, Angel R.
    Langberg, Jonathan J.
    Lloyd, Michael S.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2017, 40 (05): : 467 - 475
  • [34] RECOMMENDATIONS FOR PERIOPERATIVE MANAGEMENT OF PATIENTS WITH CARDIAC IMPLANTABLE ELECTRONIC DEVICES
    Pavlovic, Nikola
    Manola, Sime
    Vrazic, Hrvoje
    Vucic, Marinko
    Brusich, Sandro
    Radeljic, Vjekoslav
    Zeljkovic, Ivan
    Matasic, Richard
    Anic, Ante
    Benko, Ivica
    Gavranovic, Zeljka
    Glogoski, Milana Zlatic
    ACTA CLINICA CROATICA, 2018, 57 (02) : 383 - 390
  • [35] Safety of magnetic resonance imaging in patients with cardiac implantable electronic devices with generator and lead(s) brand mismatch
    Minaskeian, Nareg
    Hajnal, Sofia P.
    Liu, Michael B.
    Klooster, Lindsay M.
    Devick, Katrina L.
    Schwartz, Linda
    Jokerst, Clinton E.
    Sorajja, Dan
    Scott, Luis R. P.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (03):
  • [36] Magnetic resonance imaging in individuals with cardiovascular implantable electronic devices
    Roguin, Ariel
    Schwitter, Juerg
    Vahlhaus, Christian
    Lombardi, Massimo
    Brugada, Josep
    Vardas, Panos
    Auricchio, Angelo
    Priori, Silvia
    Sommer, Torsten
    EUROPACE, 2008, 10 (03): : 334 - 344
  • [37] Magnetic Resonance Imaging in People With Cardiac Implantable Electronic Devices: A Population Based Cohort Study
    Gillam, Marianne H.
    Inacio, Maria C. S.
    Pratt, Nicole L.
    Shakib, Sepehr
    Roughead, Elizabeth E.
    HEART LUNG AND CIRCULATION, 2018, 27 (06): : 748 - 751
  • [38] Cardiac Magnetic Resonance and Cardiac Implantable Electronic Devices: Are They Truly Still "Enemies"?
    Fogante, Marco
    Volpato, Giovanni
    Pirani, Paolo Esposto
    Cela, Fatjon
    Compagnucci, Paolo
    Valeri, Yari
    Selimi, Adelina
    Alfieri, Michele
    Brugiatelli, Leonardo
    Belleggia, Sara
    Coraducci, Francesca
    Argalia, Giulio
    Casella, Michela
    Dello Russo, Antonio
    Schicchi, Nicolo
    MEDICINA-LITHUANIA, 2024, 60 (04):
  • [39] Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols
    Horwood, Laura
    Attili, Anil
    Luba, Frank
    Ibrahim, El-Sayed H.
    Parmar, Hemant
    Stojanovska, Jadranka
    Gadoth-Goodman, Sharon
    Fette, Carey
    Oral, Hakan
    Bogun, Frank
    EUROPACE, 2017, 19 (05): : 812 - 817
  • [40] Cardiovascular magnetic resonance imaging in patients with cardiac implantable electronic devices: best practice and real-world experience
    Seewoester, Timm
    Loebe, Susanne
    Hilbert, Sebastian
    Bollmann, Andreas
    Sommer, Philipp
    Lindemann, Frank
    Bacevicius, Justinas
    Schoene, Katharina
    Richter, Sergio
    Doering, Michael
    Paetsch, Ingo
    Hindricks, Gerhard
    Jahnke, Cosima
    EUROPACE, 2019, 21 (08): : 1220 - 1228